Comparison of corneal endothelial cell measurements by two non-contact specular microscopes

BACKGROUND: Measurement of corneal endothelial cell density is important both for clinical diagnosis as well as clinical studies. Since endothelial cell loss is considered irreversible in humans, even small changes in endothelial cell density are relevant. Therefore it is important to know whether d...

Descripción completa

Detalles Bibliográficos
Autores principales: Gasser, Laura, Reinhard, Thomas, Böhringer, Daniel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4517400/
https://www.ncbi.nlm.nih.gov/pubmed/26219961
http://dx.doi.org/10.1186/s12886-015-0068-1
_version_ 1782383183555198976
author Gasser, Laura
Reinhard, Thomas
Böhringer, Daniel
author_facet Gasser, Laura
Reinhard, Thomas
Böhringer, Daniel
author_sort Gasser, Laura
collection PubMed
description BACKGROUND: Measurement of corneal endothelial cell density is important both for clinical diagnosis as well as clinical studies. Since endothelial cell loss is considered irreversible in humans, even small changes in endothelial cell density are relevant. Therefore it is important to know whether different instruments for endothelial cell density measurements give the same results and can thus be used interchangeably. In this study we compare corneal endothelial cell density and morphometry measurements from two widely used non-contact specular microscopes, the Topcon SP3000P and Konan Noncon Robo SP8000. METHODS: Endothelial cell measurements were performed with both the Topcon SP3000P and Konan Noncon Robo SP8000 on 34 eyes of 18 consecutive patients of our cornea clinics with poor image quality being the only exclusion criterion. Images were obtained using the auto-focussing method and manual cell selection. Endothelial cell density (ECD), hexagonal cell ratio (HEX) and coefficient of value (CV) of the endothelial cell layer were calculated by the instruments’ built-in software. RESULTS: ECD values calculated by the Konan were systematically higher than Topcon values: in 94 % of eyes Konan gave a higher value than Topcon, leading to a mean difference in ECD between the instruments of 187 cells/mm(2) (P < 0.001 in paired Wilcoxon test). HEX showed a broad range of values and differed greatly with only weak correlation between the two instruments. CV values for Konan mostly exceeded Topcon values, and only showed a weak correlation between the two instruments as well. CONCLUSIONS: Values for ECD between the Konan and the Topcon do correlate well, but the ECDs calculated by the Konan are systematically higher than Topcon values. Both HEX and CV vary greatly and do not correlate sufficiently. Thus we recommend not to use the Konan and the Topcon instrument interchangeably.
format Online
Article
Text
id pubmed-4517400
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-45174002015-07-29 Comparison of corneal endothelial cell measurements by two non-contact specular microscopes Gasser, Laura Reinhard, Thomas Böhringer, Daniel BMC Ophthalmol Research Article BACKGROUND: Measurement of corneal endothelial cell density is important both for clinical diagnosis as well as clinical studies. Since endothelial cell loss is considered irreversible in humans, even small changes in endothelial cell density are relevant. Therefore it is important to know whether different instruments for endothelial cell density measurements give the same results and can thus be used interchangeably. In this study we compare corneal endothelial cell density and morphometry measurements from two widely used non-contact specular microscopes, the Topcon SP3000P and Konan Noncon Robo SP8000. METHODS: Endothelial cell measurements were performed with both the Topcon SP3000P and Konan Noncon Robo SP8000 on 34 eyes of 18 consecutive patients of our cornea clinics with poor image quality being the only exclusion criterion. Images were obtained using the auto-focussing method and manual cell selection. Endothelial cell density (ECD), hexagonal cell ratio (HEX) and coefficient of value (CV) of the endothelial cell layer were calculated by the instruments’ built-in software. RESULTS: ECD values calculated by the Konan were systematically higher than Topcon values: in 94 % of eyes Konan gave a higher value than Topcon, leading to a mean difference in ECD between the instruments of 187 cells/mm(2) (P < 0.001 in paired Wilcoxon test). HEX showed a broad range of values and differed greatly with only weak correlation between the two instruments. CV values for Konan mostly exceeded Topcon values, and only showed a weak correlation between the two instruments as well. CONCLUSIONS: Values for ECD between the Konan and the Topcon do correlate well, but the ECDs calculated by the Konan are systematically higher than Topcon values. Both HEX and CV vary greatly and do not correlate sufficiently. Thus we recommend not to use the Konan and the Topcon instrument interchangeably. BioMed Central 2015-07-29 /pmc/articles/PMC4517400/ /pubmed/26219961 http://dx.doi.org/10.1186/s12886-015-0068-1 Text en © Gasser et al. 2015 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Gasser, Laura
Reinhard, Thomas
Böhringer, Daniel
Comparison of corneal endothelial cell measurements by two non-contact specular microscopes
title Comparison of corneal endothelial cell measurements by two non-contact specular microscopes
title_full Comparison of corneal endothelial cell measurements by two non-contact specular microscopes
title_fullStr Comparison of corneal endothelial cell measurements by two non-contact specular microscopes
title_full_unstemmed Comparison of corneal endothelial cell measurements by two non-contact specular microscopes
title_short Comparison of corneal endothelial cell measurements by two non-contact specular microscopes
title_sort comparison of corneal endothelial cell measurements by two non-contact specular microscopes
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4517400/
https://www.ncbi.nlm.nih.gov/pubmed/26219961
http://dx.doi.org/10.1186/s12886-015-0068-1
work_keys_str_mv AT gasserlaura comparisonofcornealendothelialcellmeasurementsbytwononcontactspecularmicroscopes
AT reinhardthomas comparisonofcornealendothelialcellmeasurementsbytwononcontactspecularmicroscopes
AT bohringerdaniel comparisonofcornealendothelialcellmeasurementsbytwononcontactspecularmicroscopes