Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging

Deep Brain Stimulation (DBS) is increasingly used to treat a variety of brain diseases by sending electrical impulses to deep brain nuclei through long, electrically conductive leads. Magnetic resonance imaging (MRI) of patients pre- and post-implantation is desirable to target and position the impl...

Descripción completa

Detalles Bibliográficos
Autores principales: McElcheran, Clare E., Yang, Benson, Anderson, Kevan J. T., Golenstani-Rad, Laleh, Graham, Simon J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4523176/
https://www.ncbi.nlm.nih.gov/pubmed/26237218
http://dx.doi.org/10.1371/journal.pone.0134379
_version_ 1782384043465113600
author McElcheran, Clare E.
Yang, Benson
Anderson, Kevan J. T.
Golenstani-Rad, Laleh
Graham, Simon J.
author_facet McElcheran, Clare E.
Yang, Benson
Anderson, Kevan J. T.
Golenstani-Rad, Laleh
Graham, Simon J.
author_sort McElcheran, Clare E.
collection PubMed
description Deep Brain Stimulation (DBS) is increasingly used to treat a variety of brain diseases by sending electrical impulses to deep brain nuclei through long, electrically conductive leads. Magnetic resonance imaging (MRI) of patients pre- and post-implantation is desirable to target and position the implant, to evaluate possible side-effects and to examine DBS patients who have other health conditions. Although MRI is the preferred modality for pre-operative planning, MRI post-implantation is limited due to the risk of high local power deposition, and therefore tissue heating, at the tip of the lead. The localized power deposition arises from currents induced in the leads caused by coupling with the radiofrequency (RF) transmission field during imaging. In the present work, parallel RF transmission (pTx) is used to tailor the RF electric field to suppress coupling effects. Electromagnetic simulations were performed for three pTx coil configurations with 2, 4, and 8-elements, respectively. Optimal input voltages to minimize coupling, while maintaining RF magnetic field homogeneity, were determined for all configurations using a Nelder-Mead optimization algorithm. Resulting electric and magnetic fields were compared to that of a 16-rung birdcage coil. Experimental validation was performed with a custom-built 4-element pTx coil. In simulation, 95-99% reduction of the electric field at the tip of the lead was observed between the various pTx coil configurations and the birdcage coil. Maximal reduction in E-field was obtained with the 8-element pTx coil. Magnetic field homogeneity was comparable to the birdcage coil for the 4- and 8-element pTx configurations. In experiment, a temperature increase of 2±0.15°C was observed at the tip of the wire using the birdcage coil, whereas negligible increase (0.2±0.15°C) was observed with the optimized pTx system. Although further research is required, these initial results suggest that the concept of optimizing pTx to reduce DBS heating effects holds considerable promise.
format Online
Article
Text
id pubmed-4523176
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-45231762015-08-06 Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging McElcheran, Clare E. Yang, Benson Anderson, Kevan J. T. Golenstani-Rad, Laleh Graham, Simon J. PLoS One Research Article Deep Brain Stimulation (DBS) is increasingly used to treat a variety of brain diseases by sending electrical impulses to deep brain nuclei through long, electrically conductive leads. Magnetic resonance imaging (MRI) of patients pre- and post-implantation is desirable to target and position the implant, to evaluate possible side-effects and to examine DBS patients who have other health conditions. Although MRI is the preferred modality for pre-operative planning, MRI post-implantation is limited due to the risk of high local power deposition, and therefore tissue heating, at the tip of the lead. The localized power deposition arises from currents induced in the leads caused by coupling with the radiofrequency (RF) transmission field during imaging. In the present work, parallel RF transmission (pTx) is used to tailor the RF electric field to suppress coupling effects. Electromagnetic simulations were performed for three pTx coil configurations with 2, 4, and 8-elements, respectively. Optimal input voltages to minimize coupling, while maintaining RF magnetic field homogeneity, were determined for all configurations using a Nelder-Mead optimization algorithm. Resulting electric and magnetic fields were compared to that of a 16-rung birdcage coil. Experimental validation was performed with a custom-built 4-element pTx coil. In simulation, 95-99% reduction of the electric field at the tip of the lead was observed between the various pTx coil configurations and the birdcage coil. Maximal reduction in E-field was obtained with the 8-element pTx coil. Magnetic field homogeneity was comparable to the birdcage coil for the 4- and 8-element pTx configurations. In experiment, a temperature increase of 2±0.15°C was observed at the tip of the wire using the birdcage coil, whereas negligible increase (0.2±0.15°C) was observed with the optimized pTx system. Although further research is required, these initial results suggest that the concept of optimizing pTx to reduce DBS heating effects holds considerable promise. Public Library of Science 2015-08-03 /pmc/articles/PMC4523176/ /pubmed/26237218 http://dx.doi.org/10.1371/journal.pone.0134379 Text en © 2015 McElcheran et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
McElcheran, Clare E.
Yang, Benson
Anderson, Kevan J. T.
Golenstani-Rad, Laleh
Graham, Simon J.
Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging
title Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging
title_full Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging
title_fullStr Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging
title_full_unstemmed Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging
title_short Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging
title_sort investigation of parallel radiofrequency transmission for the reduction of heating in long conductive leads in 3 tesla magnetic resonance imaging
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4523176/
https://www.ncbi.nlm.nih.gov/pubmed/26237218
http://dx.doi.org/10.1371/journal.pone.0134379
work_keys_str_mv AT mcelcheranclaree investigationofparallelradiofrequencytransmissionforthereductionofheatinginlongconductiveleadsin3teslamagneticresonanceimaging
AT yangbenson investigationofparallelradiofrequencytransmissionforthereductionofheatinginlongconductiveleadsin3teslamagneticresonanceimaging
AT andersonkevanjt investigationofparallelradiofrequencytransmissionforthereductionofheatinginlongconductiveleadsin3teslamagneticresonanceimaging
AT golenstaniradlaleh investigationofparallelradiofrequencytransmissionforthereductionofheatinginlongconductiveleadsin3teslamagneticresonanceimaging
AT grahamsimonj investigationofparallelradiofrequencytransmissionforthereductionofheatinginlongconductiveleadsin3teslamagneticresonanceimaging