Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment
BACKGROUND: Pulse oxygen saturation (SpO(2)) is an important parameter for healthcare, and wearable sensors and systems for SpO(2) monitoring have become increasingly popular. The aim of this paper is to develop a novel SpO(2) monitoring system, which detects photoplethysmographic (PPG) signals at h...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4523957/ https://www.ncbi.nlm.nih.gov/pubmed/26242309 http://dx.doi.org/10.1186/s12938-015-0071-z |
_version_ | 1782384143651307520 |
---|---|
author | Guo, Tao Cao, Zhengtao Zhang, Zhengbo Li, Deyu Yu, Mengsun |
author_facet | Guo, Tao Cao, Zhengtao Zhang, Zhengbo Li, Deyu Yu, Mengsun |
author_sort | Guo, Tao |
collection | PubMed |
description | BACKGROUND: Pulse oxygen saturation (SpO(2)) is an important parameter for healthcare, and wearable sensors and systems for SpO(2) monitoring have become increasingly popular. The aim of this paper is to develop a novel SpO(2) monitoring system, which detects photoplethysmographic (PPG) signals at hypothenar with a reflection-mode sensor embedded into a glove. METHODS: A special photo-detector section was designed with two photodiodes arranged symmetrically to the red and infrared light-emitting diodes (LED) to enhance the signal quality. The reflective sensor was placed in a soft silicon substrate sewn in a glove to fit the surface of the hypothenar. To lower the power consumption, the LED driving current was reduced and energy-efficient electronic components were applied. The performance for PPG signal detection and SpO(2) monitoring was evaluated by human hypoxia experiments. Accelerometer-based adaptive noise cancellation (ANC) methods applying the least mean squares (LMS) and recursive least squares (RLS) algorithms were studied to suppress motion artifact. RESULTS: A total of 20 subjects participated in the hypoxia experiment. The degree of comfort for wearing this system was accepted by them. The PPG signals were detected effectively at SpO(2) levels from about 100–70%. The experiment validated the accuracy of the system was 2.34%, compared to the invasive measurements. Both the LMS and RLS algorithms improved the performance during motion. The total current consumed by the system was only 8 mA. CONCLUSIONS: It is feasible to detect PPG signal and monitor SpO(2) at the location of hypothenar. This novel system can achieve reliable SpO(2) measurements at different SpO(2) levels and on different individuals. The system is light-weighted, easy to wear and power-saving. It has the potential to be a solution for wearable monitoring, although more work should be conducted to improve the motion-resistant performance significantly. |
format | Online Article Text |
id | pubmed-4523957 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-45239572015-08-05 Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment Guo, Tao Cao, Zhengtao Zhang, Zhengbo Li, Deyu Yu, Mengsun Biomed Eng Online Research BACKGROUND: Pulse oxygen saturation (SpO(2)) is an important parameter for healthcare, and wearable sensors and systems for SpO(2) monitoring have become increasingly popular. The aim of this paper is to develop a novel SpO(2) monitoring system, which detects photoplethysmographic (PPG) signals at hypothenar with a reflection-mode sensor embedded into a glove. METHODS: A special photo-detector section was designed with two photodiodes arranged symmetrically to the red and infrared light-emitting diodes (LED) to enhance the signal quality. The reflective sensor was placed in a soft silicon substrate sewn in a glove to fit the surface of the hypothenar. To lower the power consumption, the LED driving current was reduced and energy-efficient electronic components were applied. The performance for PPG signal detection and SpO(2) monitoring was evaluated by human hypoxia experiments. Accelerometer-based adaptive noise cancellation (ANC) methods applying the least mean squares (LMS) and recursive least squares (RLS) algorithms were studied to suppress motion artifact. RESULTS: A total of 20 subjects participated in the hypoxia experiment. The degree of comfort for wearing this system was accepted by them. The PPG signals were detected effectively at SpO(2) levels from about 100–70%. The experiment validated the accuracy of the system was 2.34%, compared to the invasive measurements. Both the LMS and RLS algorithms improved the performance during motion. The total current consumed by the system was only 8 mA. CONCLUSIONS: It is feasible to detect PPG signal and monitor SpO(2) at the location of hypothenar. This novel system can achieve reliable SpO(2) measurements at different SpO(2) levels and on different individuals. The system is light-weighted, easy to wear and power-saving. It has the potential to be a solution for wearable monitoring, although more work should be conducted to improve the motion-resistant performance significantly. BioMed Central 2015-08-05 /pmc/articles/PMC4523957/ /pubmed/26242309 http://dx.doi.org/10.1186/s12938-015-0071-z Text en © Guo et al. 2015 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Guo, Tao Cao, Zhengtao Zhang, Zhengbo Li, Deyu Yu, Mengsun Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment |
title | Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment |
title_full | Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment |
title_fullStr | Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment |
title_full_unstemmed | Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment |
title_short | Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment |
title_sort | reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4523957/ https://www.ncbi.nlm.nih.gov/pubmed/26242309 http://dx.doi.org/10.1186/s12938-015-0071-z |
work_keys_str_mv | AT guotao reflectiveoxygensaturationmonitoringathypothenaranditsvalidationbyhumanhypoxiaexperiment AT caozhengtao reflectiveoxygensaturationmonitoringathypothenaranditsvalidationbyhumanhypoxiaexperiment AT zhangzhengbo reflectiveoxygensaturationmonitoringathypothenaranditsvalidationbyhumanhypoxiaexperiment AT lideyu reflectiveoxygensaturationmonitoringathypothenaranditsvalidationbyhumanhypoxiaexperiment AT yumengsun reflectiveoxygensaturationmonitoringathypothenaranditsvalidationbyhumanhypoxiaexperiment |