Cargando…
Biomechanical properties of five different currently used implants for open-wedge high tibial osteotomy
BACKGROUND: As several new tibial osteotomy plates recently appeared on the market, the aim of the present study was to compare mechanical static and fatigue strength of three newly designed plates with gold standard plates for the treatment of medial knee joint osteoarthritis. METHODS: Sixteen four...
Autores principales: | Diffo Kaze, Arnaud, Maas, Stefan, Waldmann, Danièle, Zilian, Andreas, Dueck, Klaus, Pape, Dietrich |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4538734/ https://www.ncbi.nlm.nih.gov/pubmed/26914882 http://dx.doi.org/10.1186/s40634-015-0030-4 |
Ejemplares similares
-
Static and fatigue strength of a novel anatomically contoured implant compared to five current open-wedge high tibial osteotomy plates
por: Diffo Kaze, Arnaud, et al.
Publicado: (2017) -
The biomechanical effects of allograft wedges used for large corrections during medial opening wedge high tibial osteotomy
por: Belsey, James, et al.
Publicado: (2019) -
Static and Dynamic Differences in Fixation Stability between a Spacer Plate and a Small Stature Plate Fixator Used for High Tibial Osteotomies: A Biomechanical Bone Composite Study
por: Maas, Stefan, et al.
Publicado: (2013) -
Mechanical strength assessment of a drilled hole in the contralateral cortex at the end of the open wedge for high tibial osteotomy
por: Diffo Kaze, Arnaud, et al.
Publicado: (2017) -
Mechanical strength of a new plate compared to six previously tested opening wedge high tibial osteotomy implants
por: Diffo Kaze, Arnaud, et al.
Publicado: (2019)