Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model

BACKGROUND: Recently, a lot of energy devices in the surgical field, especially in the liver surgery, have been developed, and a fine tip LigaSure™, Dolphin Tip Sealer/Divider (DT-SD) also has been used frequently to dissect liver parenchyma as well as ultrasonically activated device (USAD). However...

Descripción completa

Detalles Bibliográficos
Autores principales: Toyama, Yoichi, Yoshida, Seiya, Saito, Ryota, Iwase, Ryota, Haruki, Koichiro, Okui, Norimitsu, Shimada, Jun-ichi, Kitamura, Hiroaki, Matsumoto, Michinori, Yanaga, Katsuhiko
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4543461/
https://www.ncbi.nlm.nih.gov/pubmed/26289073
http://dx.doi.org/10.1186/s13104-015-1316-4
_version_ 1782386595926638592
author Toyama, Yoichi
Yoshida, Seiya
Saito, Ryota
Iwase, Ryota
Haruki, Koichiro
Okui, Norimitsu
Shimada, Jun-ichi
Kitamura, Hiroaki
Matsumoto, Michinori
Yanaga, Katsuhiko
author_facet Toyama, Yoichi
Yoshida, Seiya
Saito, Ryota
Iwase, Ryota
Haruki, Koichiro
Okui, Norimitsu
Shimada, Jun-ichi
Kitamura, Hiroaki
Matsumoto, Michinori
Yanaga, Katsuhiko
author_sort Toyama, Yoichi
collection PubMed
description BACKGROUND: Recently, a lot of energy devices in the surgical field, especially in the liver surgery, have been developed, and a fine tip LigaSure™, Dolphin Tip Sealer/Divider (DT-SD) also has been used frequently to dissect liver parenchyma as well as ultrasonically activated device (USAD). However, the utility of this instrument for liver dissection (LD) is still unknown. Moreover, to reduce bleeding during LD, a half-grip technique (HGT) was contrived. We herein report an experimental study in swine model to evaluate the feasibility and effectiveness of HGT using DT-SD for LD. METHODS: The swine model experiment was carried out under general anesthesia by veterinarians. LD was performed repeatedly by DT-SD with the HGT (Group A, n = 6), or the conventional clamp-crush technique (CCT) (Group B, n = 6), and by variable mode USAD (Group C, n = 6). The dissection length and depth (cm) as well as bleeding volume (g) were measured carefully, and the dissection area (cm(2)) and speed (cm(2)/min) were calculated precisely. Histological examinations of the dissection surfaces were also executed. Mann–Whitney’s U test was used for Statistical analyses with variance at a significance level of 0.05. RESULTS: Among the three groups, the three averages of dissection lengths were unexpectedly equalized to 8.3 cm. The dissection area (cm(2)) was 9.9 ± 5.1 in Group A, 9.8 ± 4.7 in Group B, and 9.9 ± 4.5 in Group C. The mean blood loss during LD was 10.6 ± 14.8 g in Group A, 41.4 ± 39.2 g in Group B, and 34.3 ± 39.2 g in Group C. For Group A, the bleeding rate was the least, 0.9 ± 1.0 g/cm(2), and the average depth of coagulation was the thickest, 1.47 ± 0.29 mm, among the three groups (p < 0.05). The dissection speed in Group A (1.3 ± 0.3 cm(2)/min) was slower, than that in Group C (p < 0.05). CONCLUSIONS: This report indicates firstly that the HGT using DT-SD bring the least blood loss when compared with CCT or USAD. Although the HGT is feasible and useful for LD, to popularize the HGT, further clinical studies will be needed.
format Online
Article
Text
id pubmed-4543461
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-45434612015-08-22 Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model Toyama, Yoichi Yoshida, Seiya Saito, Ryota Iwase, Ryota Haruki, Koichiro Okui, Norimitsu Shimada, Jun-ichi Kitamura, Hiroaki Matsumoto, Michinori Yanaga, Katsuhiko BMC Res Notes Research Article BACKGROUND: Recently, a lot of energy devices in the surgical field, especially in the liver surgery, have been developed, and a fine tip LigaSure™, Dolphin Tip Sealer/Divider (DT-SD) also has been used frequently to dissect liver parenchyma as well as ultrasonically activated device (USAD). However, the utility of this instrument for liver dissection (LD) is still unknown. Moreover, to reduce bleeding during LD, a half-grip technique (HGT) was contrived. We herein report an experimental study in swine model to evaluate the feasibility and effectiveness of HGT using DT-SD for LD. METHODS: The swine model experiment was carried out under general anesthesia by veterinarians. LD was performed repeatedly by DT-SD with the HGT (Group A, n = 6), or the conventional clamp-crush technique (CCT) (Group B, n = 6), and by variable mode USAD (Group C, n = 6). The dissection length and depth (cm) as well as bleeding volume (g) were measured carefully, and the dissection area (cm(2)) and speed (cm(2)/min) were calculated precisely. Histological examinations of the dissection surfaces were also executed. Mann–Whitney’s U test was used for Statistical analyses with variance at a significance level of 0.05. RESULTS: Among the three groups, the three averages of dissection lengths were unexpectedly equalized to 8.3 cm. The dissection area (cm(2)) was 9.9 ± 5.1 in Group A, 9.8 ± 4.7 in Group B, and 9.9 ± 4.5 in Group C. The mean blood loss during LD was 10.6 ± 14.8 g in Group A, 41.4 ± 39.2 g in Group B, and 34.3 ± 39.2 g in Group C. For Group A, the bleeding rate was the least, 0.9 ± 1.0 g/cm(2), and the average depth of coagulation was the thickest, 1.47 ± 0.29 mm, among the three groups (p < 0.05). The dissection speed in Group A (1.3 ± 0.3 cm(2)/min) was slower, than that in Group C (p < 0.05). CONCLUSIONS: This report indicates firstly that the HGT using DT-SD bring the least blood loss when compared with CCT or USAD. Although the HGT is feasible and useful for LD, to popularize the HGT, further clinical studies will be needed. BioMed Central 2015-08-20 /pmc/articles/PMC4543461/ /pubmed/26289073 http://dx.doi.org/10.1186/s13104-015-1316-4 Text en © Toyama et al. 2015 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Toyama, Yoichi
Yoshida, Seiya
Saito, Ryota
Iwase, Ryota
Haruki, Koichiro
Okui, Norimitsu
Shimada, Jun-ichi
Kitamura, Hiroaki
Matsumoto, Michinori
Yanaga, Katsuhiko
Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model
title Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model
title_full Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model
title_fullStr Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model
title_full_unstemmed Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model
title_short Efficacy of a half-grip technique using a fine tip LigaSure™, Dolphin Tip Sealer/Divider, on liver dissection in swine model
title_sort efficacy of a half-grip technique using a fine tip ligasure™, dolphin tip sealer/divider, on liver dissection in swine model
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4543461/
https://www.ncbi.nlm.nih.gov/pubmed/26289073
http://dx.doi.org/10.1186/s13104-015-1316-4
work_keys_str_mv AT toyamayoichi efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT yoshidaseiya efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT saitoryota efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT iwaseryota efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT harukikoichiro efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT okuinorimitsu efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT shimadajunichi efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT kitamurahiroaki efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT matsumotomichinori efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel
AT yanagakatsuhiko efficacyofahalfgriptechniqueusingafinetipligasuredolphintipsealerdivideronliverdissectioninswinemodel