Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI

BACKGROUND: Hybrid positron emission tomography/magnetic resonance imaging (PET/MRI) shows high potential for patients with oropharyngeal cancer. Dental implants can cause substantial artifacts in the oral cavity impairing diagnostic accuracy. Therefore, we evaluated new MRI sequences with multi-acq...

Descripción completa

Detalles Bibliográficos
Autores principales: Gunzinger, Jeanne M, Delso, Gaspar, Boss, Andreas, Porto, Miguel, Davison, Helen, von Schulthess, Gustav K, Huellner, Martin, Stolzmann, Paul, Veit-Haibach, Patrick, Burger, Irene A
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4545455/
https://www.ncbi.nlm.nih.gov/pubmed/26501460
http://dx.doi.org/10.1186/s40658-014-0102-z
_version_ 1782386752699236352
author Gunzinger, Jeanne M
Delso, Gaspar
Boss, Andreas
Porto, Miguel
Davison, Helen
von Schulthess, Gustav K
Huellner, Martin
Stolzmann, Paul
Veit-Haibach, Patrick
Burger, Irene A
author_facet Gunzinger, Jeanne M
Delso, Gaspar
Boss, Andreas
Porto, Miguel
Davison, Helen
von Schulthess, Gustav K
Huellner, Martin
Stolzmann, Paul
Veit-Haibach, Patrick
Burger, Irene A
author_sort Gunzinger, Jeanne M
collection PubMed
description BACKGROUND: Hybrid positron emission tomography/magnetic resonance imaging (PET/MRI) shows high potential for patients with oropharyngeal cancer. Dental implants can cause substantial artifacts in the oral cavity impairing diagnostic accuracy. Therefore, we evaluated new MRI sequences with multi-acquisition variable-resonance image combination (MAVRIC SL) in comparison to conventional high-bandwidth techniques and in a second step showed the effect of artifact size on MRI-based attenuation correction (AC) with a simulation study. METHODS: Twenty-five patients with dental implants prospectively underwent a trimodality PET/CT/MRI examination after informed consent was obtained under the approval of the local ethics committee. A conventional 3D gradient-echo sequence (LAVA-Flex) commonly used for MRI-based AC of PET (acquisition time of 14 s), a T1w fast spin-echo sequence with high bandwidth (acquisition time of 3.2 min), as well as MAVRIC SL sequence without and with increased phase acceleration (MAVRIC, acquisition time of 6 min; MAVRIC-fast, acquisition time of 3.5 min) were applied. The absolute and relative reduction of the signal void artifact was calculated for each implant and tested for statistical significance using the Wilcoxon signed-rank test. The effect of artifact size on PET AC was simulated in one case with a large tumor in the oral cavity. The relative difference of the maximum standardized uptake value (SUV(max)) in the tumor was calculated for increasing artifact sizes centered over the second molar. RESULTS: The absolute reduction of signal void from LAVA-Flex sequences to the T1-weighted fast spin-echo (FSE) sequences was 416 mm(2) (range 4 to 2,010 mm(2)) to MAVRIC 481 mm(2) (range 12 to 2,288 mm(2)) and to MAVRIC-fast 486 mm(2) (range 39 to 2,209 mm(2)). The relative reduction in signal void was significantly improved for both MAVRIC and MAVRIC-fast compared to T1 FSE (−75%/−78% vs. −62%, p < 0.001 for both). The relative error for SUV(max) was negligible for artifacts of 0.5-cm diameter (−0.1%), but substantial for artifacts of 5.2-cm diameter (−33%). CONCLUSIONS: MAVRIC-fast could become useful for artifact reduction in PET/MR for patients with dental implants. This might improve diagnostic accuracy especially for patients with tumors in the oropharynx and substantially improve accuracy of PET quantification.
format Online
Article
Text
id pubmed-4545455
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-45454552015-08-26 Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI Gunzinger, Jeanne M Delso, Gaspar Boss, Andreas Porto, Miguel Davison, Helen von Schulthess, Gustav K Huellner, Martin Stolzmann, Paul Veit-Haibach, Patrick Burger, Irene A EJNMMI Phys Original Research BACKGROUND: Hybrid positron emission tomography/magnetic resonance imaging (PET/MRI) shows high potential for patients with oropharyngeal cancer. Dental implants can cause substantial artifacts in the oral cavity impairing diagnostic accuracy. Therefore, we evaluated new MRI sequences with multi-acquisition variable-resonance image combination (MAVRIC SL) in comparison to conventional high-bandwidth techniques and in a second step showed the effect of artifact size on MRI-based attenuation correction (AC) with a simulation study. METHODS: Twenty-five patients with dental implants prospectively underwent a trimodality PET/CT/MRI examination after informed consent was obtained under the approval of the local ethics committee. A conventional 3D gradient-echo sequence (LAVA-Flex) commonly used for MRI-based AC of PET (acquisition time of 14 s), a T1w fast spin-echo sequence with high bandwidth (acquisition time of 3.2 min), as well as MAVRIC SL sequence without and with increased phase acceleration (MAVRIC, acquisition time of 6 min; MAVRIC-fast, acquisition time of 3.5 min) were applied. The absolute and relative reduction of the signal void artifact was calculated for each implant and tested for statistical significance using the Wilcoxon signed-rank test. The effect of artifact size on PET AC was simulated in one case with a large tumor in the oral cavity. The relative difference of the maximum standardized uptake value (SUV(max)) in the tumor was calculated for increasing artifact sizes centered over the second molar. RESULTS: The absolute reduction of signal void from LAVA-Flex sequences to the T1-weighted fast spin-echo (FSE) sequences was 416 mm(2) (range 4 to 2,010 mm(2)) to MAVRIC 481 mm(2) (range 12 to 2,288 mm(2)) and to MAVRIC-fast 486 mm(2) (range 39 to 2,209 mm(2)). The relative reduction in signal void was significantly improved for both MAVRIC and MAVRIC-fast compared to T1 FSE (−75%/−78% vs. −62%, p < 0.001 for both). The relative error for SUV(max) was negligible for artifacts of 0.5-cm diameter (−0.1%), but substantial for artifacts of 5.2-cm diameter (−33%). CONCLUSIONS: MAVRIC-fast could become useful for artifact reduction in PET/MR for patients with dental implants. This might improve diagnostic accuracy especially for patients with tumors in the oropharynx and substantially improve accuracy of PET quantification. Springer International Publishing 2014-12-20 /pmc/articles/PMC4545455/ /pubmed/26501460 http://dx.doi.org/10.1186/s40658-014-0102-z Text en © Gunzinger et al; licensee Springer. 2014 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited.
spellingShingle Original Research
Gunzinger, Jeanne M
Delso, Gaspar
Boss, Andreas
Porto, Miguel
Davison, Helen
von Schulthess, Gustav K
Huellner, Martin
Stolzmann, Paul
Veit-Haibach, Patrick
Burger, Irene A
Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI
title Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI
title_full Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI
title_fullStr Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI
title_full_unstemmed Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI
title_short Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI
title_sort metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid pet/mri
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4545455/
https://www.ncbi.nlm.nih.gov/pubmed/26501460
http://dx.doi.org/10.1186/s40658-014-0102-z
work_keys_str_mv AT gunzingerjeannem metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT delsogaspar metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT bossandreas metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT portomiguel metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT davisonhelen metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT vonschulthessgustavk metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT huellnermartin metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT stolzmannpaul metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT veithaibachpatrick metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri
AT burgerirenea metalartifactreductioninpatientswithdentalimplantsusingmultispectralthreedimensionaldataacquisitionforhybridpetmri