Cargando…
Effect of cyclic loading and retightening on reverse torque value in external and internal implants
PURPOSE: The aim of this study was to evaluate the effect of cyclic loading and screw retightening on reverse torque value (RTV) in external and internal type implants. MATERIALS AND METHODS: Cement-retained abutments were connected with 30 Ncm torque to external and internal type implants. Experime...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Academy of Prosthodontics
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4551784/ https://www.ncbi.nlm.nih.gov/pubmed/26330975 http://dx.doi.org/10.4047/jap.2015.7.4.288 |
_version_ | 1782387617148436480 |
---|---|
author | Cho, Woong-Rae Huh, Yoon-Hyuk Park, Chan-Jin Cho, Lee-Ra |
author_facet | Cho, Woong-Rae Huh, Yoon-Hyuk Park, Chan-Jin Cho, Lee-Ra |
author_sort | Cho, Woong-Rae |
collection | PubMed |
description | PURPOSE: The aim of this study was to evaluate the effect of cyclic loading and screw retightening on reverse torque value (RTV) in external and internal type implants. MATERIALS AND METHODS: Cement-retained abutments were connected with 30 Ncm torque to external and internal type implants. Experimental groups were classified according to implant connection type and retightening/loading protocol. In groups with no retightening, RTV was evaluated after cyclic loading for 100,000 cycles. In groups with retightening, RTV was measured after 3, 10, 100 cycles as well as every 20,000 cycles until 100,000 cycles of loading. RESULTS: Every group showed decreased RTV after cyclic loading. Before and after cyclic loading, external type implants had significantly higher RTVs than internal type implants. In external type implants, retightening did not affect the decrease in RTV. In contrast, retightening 5 times and retightening after 10 cycles of dynamic loading was effective for maintaining RTV in internal type implants. CONCLUSION: Retightening of screws is more effective in internal type implants than external type implants. Retightening of screws is recommended in the early stage of functional loading. |
format | Online Article Text |
id | pubmed-4551784 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | The Korean Academy of Prosthodontics |
record_format | MEDLINE/PubMed |
spelling | pubmed-45517842015-09-01 Effect of cyclic loading and retightening on reverse torque value in external and internal implants Cho, Woong-Rae Huh, Yoon-Hyuk Park, Chan-Jin Cho, Lee-Ra J Adv Prosthodont Original Article PURPOSE: The aim of this study was to evaluate the effect of cyclic loading and screw retightening on reverse torque value (RTV) in external and internal type implants. MATERIALS AND METHODS: Cement-retained abutments were connected with 30 Ncm torque to external and internal type implants. Experimental groups were classified according to implant connection type and retightening/loading protocol. In groups with no retightening, RTV was evaluated after cyclic loading for 100,000 cycles. In groups with retightening, RTV was measured after 3, 10, 100 cycles as well as every 20,000 cycles until 100,000 cycles of loading. RESULTS: Every group showed decreased RTV after cyclic loading. Before and after cyclic loading, external type implants had significantly higher RTVs than internal type implants. In external type implants, retightening did not affect the decrease in RTV. In contrast, retightening 5 times and retightening after 10 cycles of dynamic loading was effective for maintaining RTV in internal type implants. CONCLUSION: Retightening of screws is more effective in internal type implants than external type implants. Retightening of screws is recommended in the early stage of functional loading. The Korean Academy of Prosthodontics 2015-08 2015-08-18 /pmc/articles/PMC4551784/ /pubmed/26330975 http://dx.doi.org/10.4047/jap.2015.7.4.288 Text en © 2015 The Korean Academy of Prosthodontics http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Cho, Woong-Rae Huh, Yoon-Hyuk Park, Chan-Jin Cho, Lee-Ra Effect of cyclic loading and retightening on reverse torque value in external and internal implants |
title | Effect of cyclic loading and retightening on reverse torque value in external and internal implants |
title_full | Effect of cyclic loading and retightening on reverse torque value in external and internal implants |
title_fullStr | Effect of cyclic loading and retightening on reverse torque value in external and internal implants |
title_full_unstemmed | Effect of cyclic loading and retightening on reverse torque value in external and internal implants |
title_short | Effect of cyclic loading and retightening on reverse torque value in external and internal implants |
title_sort | effect of cyclic loading and retightening on reverse torque value in external and internal implants |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4551784/ https://www.ncbi.nlm.nih.gov/pubmed/26330975 http://dx.doi.org/10.4047/jap.2015.7.4.288 |
work_keys_str_mv | AT chowoongrae effectofcyclicloadingandretighteningonreversetorquevalueinexternalandinternalimplants AT huhyoonhyuk effectofcyclicloadingandretighteningonreversetorquevalueinexternalandinternalimplants AT parkchanjin effectofcyclicloadingandretighteningonreversetorquevalueinexternalandinternalimplants AT choleera effectofcyclicloadingandretighteningonreversetorquevalueinexternalandinternalimplants |