Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case

BACKGROUND: Breast cancer is the most common cancer in women worldwide and is the second most common cause of cancer death in women. Electrochemotherapy (ECT) used in early-phase clinical trials for the treatment of primary breast cancer resulted in a not complete tumor necrosis in most cases. The p...

Descripción completa

Detalles Bibliográficos
Autores principales: Denzi, Agnese, Strigari, Lidia, Di Filippo, Franco, Botti, Claudio, Di Filippo, Simona, Perracchio, Letizia, Ronchetti, Mattia, Cadossi, Ruggero, Liberti, Micaela
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4565101/
https://www.ncbi.nlm.nih.gov/pubmed/26355703
http://dx.doi.org/10.1186/1475-925X-14-S3-S1
_version_ 1782389556729872384
author Denzi, Agnese
Strigari, Lidia
Di Filippo, Franco
Botti, Claudio
Di Filippo, Simona
Perracchio, Letizia
Ronchetti, Mattia
Cadossi, Ruggero
Liberti, Micaela
author_facet Denzi, Agnese
Strigari, Lidia
Di Filippo, Franco
Botti, Claudio
Di Filippo, Simona
Perracchio, Letizia
Ronchetti, Mattia
Cadossi, Ruggero
Liberti, Micaela
author_sort Denzi, Agnese
collection PubMed
description BACKGROUND: Breast cancer is the most common cancer in women worldwide and is the second most common cause of cancer death in women. Electrochemotherapy (ECT) used in early-phase clinical trials for the treatment of primary breast cancer resulted in a not complete tumor necrosis in most cases. The present study was undertaken to analyze the feasibility to use ECT to treat patients with histologically proven unifocal ductal breast cancer. In particular, results of ECT treatment in a clinical case are compared with the ones of a simplified 3D dosimetric model. METHODS: This clinical study was conducted with the pulse generator Cliniporator Vitae (IGEA, Carpi, Italy). ECT procedures were performed according to ESOPE standard operating procedures. Five single needle electrodes were used with one positioned in the center of the tumor, and the other four distributed around the nodule. Histological images of the resected tumor are compared with the maps of the electric field obtained with a simplified 3D model in Comsol Multiphysics v 4.3. RESULTS: The results of the clinical case demonstrated a reduced efficacy of the ECT treatment described. The proposed simple numerical model of the breast tumor located in a low conductive tissue suggests that this is due to the reduced electric field induced inside the tumor with such 5 electrodes placement. However, where the electric field is predicted higher than the reversible electroporation threshold (E>400 V/cm), also the histological images confirm the necrosis of the target with a good agreement between the modeled and clinical results. CONCLUSIONS: The results suggest the dependence of the effectiveness of the treatment on the careful placement of the electrodes. A detailed planned procedure for the tumor analysis after the treatment is also needed in order to better correlate the single electrode positions and the histological images. Simulation models could be used to identify better electrodes configuration in planning the experimental protocol for ECT treatment of breast tumors.
format Online
Article
Text
id pubmed-4565101
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-45651012015-09-18 Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case Denzi, Agnese Strigari, Lidia Di Filippo, Franco Botti, Claudio Di Filippo, Simona Perracchio, Letizia Ronchetti, Mattia Cadossi, Ruggero Liberti, Micaela Biomed Eng Online Research BACKGROUND: Breast cancer is the most common cancer in women worldwide and is the second most common cause of cancer death in women. Electrochemotherapy (ECT) used in early-phase clinical trials for the treatment of primary breast cancer resulted in a not complete tumor necrosis in most cases. The present study was undertaken to analyze the feasibility to use ECT to treat patients with histologically proven unifocal ductal breast cancer. In particular, results of ECT treatment in a clinical case are compared with the ones of a simplified 3D dosimetric model. METHODS: This clinical study was conducted with the pulse generator Cliniporator Vitae (IGEA, Carpi, Italy). ECT procedures were performed according to ESOPE standard operating procedures. Five single needle electrodes were used with one positioned in the center of the tumor, and the other four distributed around the nodule. Histological images of the resected tumor are compared with the maps of the electric field obtained with a simplified 3D model in Comsol Multiphysics v 4.3. RESULTS: The results of the clinical case demonstrated a reduced efficacy of the ECT treatment described. The proposed simple numerical model of the breast tumor located in a low conductive tissue suggests that this is due to the reduced electric field induced inside the tumor with such 5 electrodes placement. However, where the electric field is predicted higher than the reversible electroporation threshold (E>400 V/cm), also the histological images confirm the necrosis of the target with a good agreement between the modeled and clinical results. CONCLUSIONS: The results suggest the dependence of the effectiveness of the treatment on the careful placement of the electrodes. A detailed planned procedure for the tumor analysis after the treatment is also needed in order to better correlate the single electrode positions and the histological images. Simulation models could be used to identify better electrodes configuration in planning the experimental protocol for ECT treatment of breast tumors. BioMed Central 2015-08-27 /pmc/articles/PMC4565101/ /pubmed/26355703 http://dx.doi.org/10.1186/1475-925X-14-S3-S1 Text en Copyright © 2015 Denzi et al.; http://creativecommons.org/licenses/by/4.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Denzi, Agnese
Strigari, Lidia
Di Filippo, Franco
Botti, Claudio
Di Filippo, Simona
Perracchio, Letizia
Ronchetti, Mattia
Cadossi, Ruggero
Liberti, Micaela
Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case
title Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case
title_full Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case
title_fullStr Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case
title_full_unstemmed Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case
title_short Modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case
title_sort modeling the positioning of single needle electrodes for the treatment of breast cancer in a clinical case
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4565101/
https://www.ncbi.nlm.nih.gov/pubmed/26355703
http://dx.doi.org/10.1186/1475-925X-14-S3-S1
work_keys_str_mv AT denziagnese modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT strigarilidia modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT difilippofranco modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT botticlaudio modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT difilipposimona modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT perracchioletizia modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT ronchettimattia modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT cadossiruggero modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase
AT libertimicaela modelingthepositioningofsingleneedleelectrodesforthetreatmentofbreastcancerinaclinicalcase