Cargando…
T2-based temperature monitoring in abdominal fat during MR-guided focused ultrasound treatment of patients with uterine fibroids
BACKGROUND: Near-field heating is a potential problem in focused ultrasound treatments, as it can result in thermal injury to skin, subcutaneous fat, and other tissues. Our goals were to determine if T2-based temperature mapping could be used reliably to measure near-field heating in adipose tissue...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4567827/ https://www.ncbi.nlm.nih.gov/pubmed/26366288 http://dx.doi.org/10.1186/s40349-015-0036-5 |
Sumario: | BACKGROUND: Near-field heating is a potential problem in focused ultrasound treatments, as it can result in thermal injury to skin, subcutaneous fat, and other tissues. Our goals were to determine if T2-based temperature mapping could be used reliably to measure near-field heating in adipose tissue and whether it is practical to perform such mapping during focused ultrasound treatments. METHODS: We investigated the dependence of T2 on temperature in ex vivo adipose tissue at 3T using a double-echo fast spin echo (FSE) sequence. We implemented and evaluated the T2-based temperature mapping technique in the adipose tissue of two healthy volunteers. Finally, we applied the technique during magnetic resonance-guided focused ultrasound (MRgFUS) treatments to measure near-field heating in eight patients with uterine fibroids. RESULTS: Calibration experiments in porcine adipose tissue determined a temperature coefficient of 6.16 ms/°C during heating and 5.37 ms/°C during cooling. The volunteer experiments demonstrated a strong correlation between the skin temperature and T2-based temperature measurements in the fat layer. During the treatments of patients with uterine fibroids, we observed a measurable change in the T2 of fat tissue within the path of the ultrasound beam and a temperature increase of up to 15 °C with sustained heating of more than 10 °C. CONCLUSIONS: Our results demonstrate the feasibility and importance of monitoring near-field heating in fatty tissues. The implementation of near-field monitoring between sonications can shorten treatments by reducing the cooling time. It can help improve safety by avoiding excessive heating in the near field. |
---|