Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity

BACKGROUND: The transversus abdominis (TrA), which is considered to be involved in controlling spinal stability, is covered by two muscles (i.e. the external and internal oblique muscles) as well as subcutaneous fat. Therefore, there were doubts about whether it was possible to perform highly reliab...

Descripción completa

Detalles Bibliográficos
Autores principales: Hirayama, Kuniaki, Akagi, Ryota, Takahashi, Hideyuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4568934/
https://www.ncbi.nlm.nih.gov/pubmed/26413315
http://dx.doi.org/10.1177/2058460115603420
_version_ 1782389972433633280
author Hirayama, Kuniaki
Akagi, Ryota
Takahashi, Hideyuki
author_facet Hirayama, Kuniaki
Akagi, Ryota
Takahashi, Hideyuki
author_sort Hirayama, Kuniaki
collection PubMed
description BACKGROUND: The transversus abdominis (TrA), which is considered to be involved in controlling spinal stability, is covered by two muscles (i.e. the external and internal oblique muscles) as well as subcutaneous fat. Therefore, there were doubts about whether it was possible to perform highly reliable measurements of muscle elasticity. PURPOSE: To investigate the reliability of ultrasound elastography for the quantification of elasticity of the TrA. MATERIAL AND METHODS: A skilled and an unskilled operator of ultrasound elastography measured TrA elasticity in 10 healthy men (age, 24 ± 4 years; height, 172.0 ± 5.2 cm; weight, 72.3 ± 12.0 kg) in the supine position. The tests were repeated six times; of the six measured values, a group of four measurements showing the lowest coefficient of variation (CV) was adopted and the mean values were used for further analysis. This procedure was repeated twice on each participant on two different days. RESULTS: The intra-class correlation coefficients (ICCs) between days for skilled and unskilled operators were 0.86 (P < 0.00) and 0.59 (P = 0.02), respectively, and the CVs were 8.7% and 13.8%, respectively. The ICCs between operators and CVs were in the range of 0.56–0.57 (P = 0.02–0.03) and 13.5–15.5%, respectively. No systematic error was found for any comparison. CONCLUSION: The reliabilities of the skilled and unskilled operators were high and moderate, respectively.
format Online
Article
Text
id pubmed-4568934
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-45689342015-09-25 Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity Hirayama, Kuniaki Akagi, Ryota Takahashi, Hideyuki Acta Radiol Open Original Article BACKGROUND: The transversus abdominis (TrA), which is considered to be involved in controlling spinal stability, is covered by two muscles (i.e. the external and internal oblique muscles) as well as subcutaneous fat. Therefore, there were doubts about whether it was possible to perform highly reliable measurements of muscle elasticity. PURPOSE: To investigate the reliability of ultrasound elastography for the quantification of elasticity of the TrA. MATERIAL AND METHODS: A skilled and an unskilled operator of ultrasound elastography measured TrA elasticity in 10 healthy men (age, 24 ± 4 years; height, 172.0 ± 5.2 cm; weight, 72.3 ± 12.0 kg) in the supine position. The tests were repeated six times; of the six measured values, a group of four measurements showing the lowest coefficient of variation (CV) was adopted and the mean values were used for further analysis. This procedure was repeated twice on each participant on two different days. RESULTS: The intra-class correlation coefficients (ICCs) between days for skilled and unskilled operators were 0.86 (P < 0.00) and 0.59 (P = 0.02), respectively, and the CVs were 8.7% and 13.8%, respectively. The ICCs between operators and CVs were in the range of 0.56–0.57 (P = 0.02–0.03) and 13.5–15.5%, respectively. No systematic error was found for any comparison. CONCLUSION: The reliabilities of the skilled and unskilled operators were high and moderate, respectively. SAGE Publications 2015-09-08 /pmc/articles/PMC4568934/ /pubmed/26413315 http://dx.doi.org/10.1177/2058460115603420 Text en © The Foundation Acta Radiologica 2015 http://creativecommons.org/licenses/by-nc/3.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 3.0 License (http://www.creativecommons.org/licenses/by-nc/3.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access page(https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Article
Hirayama, Kuniaki
Akagi, Ryota
Takahashi, Hideyuki
Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity
title Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity
title_full Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity
title_fullStr Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity
title_full_unstemmed Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity
title_short Reliability of ultrasound elastography for the quantification of transversus abdominis elasticity
title_sort reliability of ultrasound elastography for the quantification of transversus abdominis elasticity
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4568934/
https://www.ncbi.nlm.nih.gov/pubmed/26413315
http://dx.doi.org/10.1177/2058460115603420
work_keys_str_mv AT hirayamakuniaki reliabilityofultrasoundelastographyforthequantificationoftransversusabdominiselasticity
AT akagiryota reliabilityofultrasoundelastographyforthequantificationoftransversusabdominiselasticity
AT takahashihideyuki reliabilityofultrasoundelastographyforthequantificationoftransversusabdominiselasticity