The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T
BACKGROUND: Dual-channel transmit technology improves the image quality in cardiovascular magnetic resonance (CMR) at 3 T by reducing the degree of radiofrequency (RF) shading over the heart by using RF shimming. Further improvements in image quality have been shown on a dual-transmit system using h...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4575474/ https://www.ncbi.nlm.nih.gov/pubmed/26385206 http://dx.doi.org/10.1186/s12968-015-0188-z |
_version_ | 1782390784451936256 |
---|---|
author | Brink, Wyger M. van den Brink, Johan S. Webb, Andrew G. |
author_facet | Brink, Wyger M. van den Brink, Johan S. Webb, Andrew G. |
author_sort | Brink, Wyger M. |
collection | PubMed |
description | BACKGROUND: Dual-channel transmit technology improves the image quality in cardiovascular magnetic resonance (CMR) at 3 T by reducing the degree of radiofrequency (RF) shading over the heart by using RF shimming. Further improvements in image quality have been shown on a dual-transmit system using high permittivity pads. The aim of this study is to investigate the transmit field (B(1)(+)) homogeneity and the specific absorption rate (SAR) using high permittivity pads as a function of the complete range of possible RF-shim settings in order to gauge the efficacy and safety of this approach. METHODS: Electromagnetic (EM) simulations were performed in five different body models using a dual-transmit RF coil, with and without high permittivity pads. The RF shimming behaviour in terms of B(1)(+) homogeneity and local SAR were determined as a function of different RF-shim settings. Comparative experimental data were obtained in healthy volunteers (n = 33) on either a standard-bore (60 cm diameter) or wide-bore (70 cm diameter) 3 T CMR system. RESULTS: EM simulations and experimental data showed higher (B(1)(+)) homogeneity and lower SAR for optimized RF-shim settings when using the high permittivity pads. The power distribution between the two channels was also much closer to being equal using the pads. EM simulations showed that for all five body models studied, optimized RF-shim settings corresponded to reduced local SAR using high permittivity pads. However, there are also specific, non-optimal RF-shim settings for which the actual SAR using the pads would be higher (up to ~20 %) than that calculated by the CMR system. CONCLUSIONS: The combination of active (dual transmit) and passive (high permittivity pads) RF shimming shows great promise for increasing image quality for cardiac imaging at 3 T. Optimized RF-shim settings result in increased B(1)(+) homogeneity and reduced SAR with the high permittivity pads: however, there are non-optimal cases in which SAR might be underestimated, and these merit further investigation. |
format | Online Article Text |
id | pubmed-4575474 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-45754742015-09-23 The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T Brink, Wyger M. van den Brink, Johan S. Webb, Andrew G. J Cardiovasc Magn Reson Research BACKGROUND: Dual-channel transmit technology improves the image quality in cardiovascular magnetic resonance (CMR) at 3 T by reducing the degree of radiofrequency (RF) shading over the heart by using RF shimming. Further improvements in image quality have been shown on a dual-transmit system using high permittivity pads. The aim of this study is to investigate the transmit field (B(1)(+)) homogeneity and the specific absorption rate (SAR) using high permittivity pads as a function of the complete range of possible RF-shim settings in order to gauge the efficacy and safety of this approach. METHODS: Electromagnetic (EM) simulations were performed in five different body models using a dual-transmit RF coil, with and without high permittivity pads. The RF shimming behaviour in terms of B(1)(+) homogeneity and local SAR were determined as a function of different RF-shim settings. Comparative experimental data were obtained in healthy volunteers (n = 33) on either a standard-bore (60 cm diameter) or wide-bore (70 cm diameter) 3 T CMR system. RESULTS: EM simulations and experimental data showed higher (B(1)(+)) homogeneity and lower SAR for optimized RF-shim settings when using the high permittivity pads. The power distribution between the two channels was also much closer to being equal using the pads. EM simulations showed that for all five body models studied, optimized RF-shim settings corresponded to reduced local SAR using high permittivity pads. However, there are also specific, non-optimal RF-shim settings for which the actual SAR using the pads would be higher (up to ~20 %) than that calculated by the CMR system. CONCLUSIONS: The combination of active (dual transmit) and passive (high permittivity pads) RF shimming shows great promise for increasing image quality for cardiac imaging at 3 T. Optimized RF-shim settings result in increased B(1)(+) homogeneity and reduced SAR with the high permittivity pads: however, there are non-optimal cases in which SAR might be underestimated, and these merit further investigation. BioMed Central 2015-09-19 /pmc/articles/PMC4575474/ /pubmed/26385206 http://dx.doi.org/10.1186/s12968-015-0188-z Text en © Brink et al. 2015 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Brink, Wyger M. van den Brink, Johan S. Webb, Andrew G. The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T |
title | The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T |
title_full | The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T |
title_fullStr | The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T |
title_full_unstemmed | The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T |
title_short | The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T |
title_sort | effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3t |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4575474/ https://www.ncbi.nlm.nih.gov/pubmed/26385206 http://dx.doi.org/10.1186/s12968-015-0188-z |
work_keys_str_mv | AT brinkwygerm theeffectofhighpermittivitypadsonspecificabsorptionrateinradiofrequencyshimmeddualtransmitcardiovascularmagneticresonanceat3t AT vandenbrinkjohans theeffectofhighpermittivitypadsonspecificabsorptionrateinradiofrequencyshimmeddualtransmitcardiovascularmagneticresonanceat3t AT webbandrewg theeffectofhighpermittivitypadsonspecificabsorptionrateinradiofrequencyshimmeddualtransmitcardiovascularmagneticresonanceat3t AT brinkwygerm effectofhighpermittivitypadsonspecificabsorptionrateinradiofrequencyshimmeddualtransmitcardiovascularmagneticresonanceat3t AT vandenbrinkjohans effectofhighpermittivitypadsonspecificabsorptionrateinradiofrequencyshimmeddualtransmitcardiovascularmagneticresonanceat3t AT webbandrewg effectofhighpermittivitypadsonspecificabsorptionrateinradiofrequencyshimmeddualtransmitcardiovascularmagneticresonanceat3t |