Validation and reliability testing of a new, fully integrated gait analysis insole

BACKGROUND: A new tool (OpenGo, Moticon GmbH) was introduced to continuously measure kinetic and temporospatial gait parameters independently through an insole over up to 4 weeks. The goal of this study was to investigate the validity and reliability of this new insole system in a group of healthy i...

Descripción completa

Detalles Bibliográficos
Autores principales: Braun, Benedikt Johannes, Veith, Nils Thomas, Hell, Rebecca, Döbele, Stefan, Roland, Michael, Rollmann, Mika, Holstein, Jörg, Pohlemann, Tim
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4578601/
https://www.ncbi.nlm.nih.gov/pubmed/26396594
http://dx.doi.org/10.1186/s13047-015-0111-8
Descripción
Sumario:BACKGROUND: A new tool (OpenGo, Moticon GmbH) was introduced to continuously measure kinetic and temporospatial gait parameters independently through an insole over up to 4 weeks. The goal of this study was to investigate the validity and reliability of this new insole system in a group of healthy individuals. METHODS: Gait data were collected from 12 healthy individuals on a treadmill at two different speeds. In total, six trials of three minutes each were performed by every participant. Validation was performed with the FDM-S System (Zebris). Complete sensor data were used for a within test reliability analysis of over 10000 steps. Intraclass correlation was calculated for different gait parameters and analysis of variance performed. RESULTS: Intraclass correlation for the validation was >0.796 for temporospatial and kinetic gait parameters. No statistical difference was seen between the insole and force plate measurements (difference between means: 36.3 ± 27.19 N; p = 0.19 and 0.027 ± 0.028 s; p = 0.36). Intraclass correlation for the reliability was >0.994 for all parameters measured. CONCLUSION: The system is feasible for clinical trials that require step by step as well as grouped analysis of gait over a long period of time. Comparable validity and reliability to a stationary analysis tool has been shown.