Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams
Purpose: To evaluate the dosimetric characteristics of a two-dimensional (2D) diode array detector irradiated with passively scattered proton beams. Materials and Methods: A diode array detector, MapCHECK (Model 1175, Sun Nuclear, Melbourne, FL, USA) was characterized in passive-scattered proton bea...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4586777/ https://www.ncbi.nlm.nih.gov/pubmed/26264024 http://dx.doi.org/10.3390/cancers7030844 |
_version_ | 1782392431310798848 |
---|---|
author | Liengsawangwong, Praimakorn Sahoo, Nanayan Ding, Xiaoning Lii, MingFwu Gillin, Michale T. Zhu, Xiaorong Ronald |
author_facet | Liengsawangwong, Praimakorn Sahoo, Nanayan Ding, Xiaoning Lii, MingFwu Gillin, Michale T. Zhu, Xiaorong Ronald |
author_sort | Liengsawangwong, Praimakorn |
collection | PubMed |
description | Purpose: To evaluate the dosimetric characteristics of a two-dimensional (2D) diode array detector irradiated with passively scattered proton beams. Materials and Methods: A diode array detector, MapCHECK (Model 1175, Sun Nuclear, Melbourne, FL, USA) was characterized in passive-scattered proton beams. The relative sensitivity of the diodes and absolute dose calibration were determined using a 250 MeV beam. The pristine Bragg curves (PBCs) measured by MapCHECK diodes were compared with those of an ion chamber using a range shift method. The water-equivalent thickness (WET) of the diode array detector’s intrinsic buildup also was determined. The inverse square dependence, linearity, and other proton dosimetric quantities measured by MapCHECK were also compared with those of the ion chambers. The change in the absolute dose response of the MapCHECK as a function of accumulated radiation dose was used as an indicator of radiation damage to the diodes. 2D dose distribution with and without the compensator were measured and compared with the treatment planning system (TPS) calculations. Results: The WET of the MapCHECK diode’s buildup was determined to be 1.7 cm. The MapCHECK-measured PBC were virtually identical to those measured by a parallel-plate ion chamber for 160, 180, and 250 MeV proton beams. The inverse square results of the MapCHECK were within ±0.4% of the ion chamber results. The linearity of MapCHECK results was within 1% of those from the ion chamber as measured in the range between 10 and 300 MU. All other dosimetric quantities were within 1.3% of the ion chamber results. The 2D dose distributions for non-clinical fields without compensator and the patient treatment fields with the compensator were consistent with the TPS results. The absolute dose response of the MapCHECK was changed by 7.4% after an accumulated dose increased by 170 Gy. Conclusions: The MapCHECK is a convenient and useful tool for 2D dose distribution measurements using passively scattered proton beams. Variations in MapCHECK’s dose response with increasing levels of total accumulated radiation dose should be carefully monitored. |
format | Online Article Text |
id | pubmed-4586777 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-45867772015-10-06 Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams Liengsawangwong, Praimakorn Sahoo, Nanayan Ding, Xiaoning Lii, MingFwu Gillin, Michale T. Zhu, Xiaorong Ronald Cancers (Basel) Article Purpose: To evaluate the dosimetric characteristics of a two-dimensional (2D) diode array detector irradiated with passively scattered proton beams. Materials and Methods: A diode array detector, MapCHECK (Model 1175, Sun Nuclear, Melbourne, FL, USA) was characterized in passive-scattered proton beams. The relative sensitivity of the diodes and absolute dose calibration were determined using a 250 MeV beam. The pristine Bragg curves (PBCs) measured by MapCHECK diodes were compared with those of an ion chamber using a range shift method. The water-equivalent thickness (WET) of the diode array detector’s intrinsic buildup also was determined. The inverse square dependence, linearity, and other proton dosimetric quantities measured by MapCHECK were also compared with those of the ion chambers. The change in the absolute dose response of the MapCHECK as a function of accumulated radiation dose was used as an indicator of radiation damage to the diodes. 2D dose distribution with and without the compensator were measured and compared with the treatment planning system (TPS) calculations. Results: The WET of the MapCHECK diode’s buildup was determined to be 1.7 cm. The MapCHECK-measured PBC were virtually identical to those measured by a parallel-plate ion chamber for 160, 180, and 250 MeV proton beams. The inverse square results of the MapCHECK were within ±0.4% of the ion chamber results. The linearity of MapCHECK results was within 1% of those from the ion chamber as measured in the range between 10 and 300 MU. All other dosimetric quantities were within 1.3% of the ion chamber results. The 2D dose distributions for non-clinical fields without compensator and the patient treatment fields with the compensator were consistent with the TPS results. The absolute dose response of the MapCHECK was changed by 7.4% after an accumulated dose increased by 170 Gy. Conclusions: The MapCHECK is a convenient and useful tool for 2D dose distribution measurements using passively scattered proton beams. Variations in MapCHECK’s dose response with increasing levels of total accumulated radiation dose should be carefully monitored. MDPI 2015-07-30 /pmc/articles/PMC4586777/ /pubmed/26264024 http://dx.doi.org/10.3390/cancers7030844 Text en © 2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Liengsawangwong, Praimakorn Sahoo, Nanayan Ding, Xiaoning Lii, MingFwu Gillin, Michale T. Zhu, Xiaorong Ronald Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams |
title | Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams |
title_full | Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams |
title_fullStr | Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams |
title_full_unstemmed | Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams |
title_short | Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams |
title_sort | dosimetric characteristics of a two-dimensional diode array detector irradiated with passively scattered proton beams |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4586777/ https://www.ncbi.nlm.nih.gov/pubmed/26264024 http://dx.doi.org/10.3390/cancers7030844 |
work_keys_str_mv | AT liengsawangwongpraimakorn dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams AT sahoonanayan dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams AT dingxiaoning dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams AT liimingfwu dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams AT gillinmichalet dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams AT zhuxiaorongronald dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams |