Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams

Purpose: To evaluate the dosimetric characteristics of a two-dimensional (2D) diode array detector irradiated with passively scattered proton beams. Materials and Methods: A diode array detector, MapCHECK (Model 1175, Sun Nuclear, Melbourne, FL, USA) was characterized in passive-scattered proton bea...

Descripción completa

Detalles Bibliográficos
Autores principales: Liengsawangwong, Praimakorn, Sahoo, Nanayan, Ding, Xiaoning, Lii, MingFwu, Gillin, Michale T., Zhu, Xiaorong Ronald
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4586777/
https://www.ncbi.nlm.nih.gov/pubmed/26264024
http://dx.doi.org/10.3390/cancers7030844
_version_ 1782392431310798848
author Liengsawangwong, Praimakorn
Sahoo, Nanayan
Ding, Xiaoning
Lii, MingFwu
Gillin, Michale T.
Zhu, Xiaorong Ronald
author_facet Liengsawangwong, Praimakorn
Sahoo, Nanayan
Ding, Xiaoning
Lii, MingFwu
Gillin, Michale T.
Zhu, Xiaorong Ronald
author_sort Liengsawangwong, Praimakorn
collection PubMed
description Purpose: To evaluate the dosimetric characteristics of a two-dimensional (2D) diode array detector irradiated with passively scattered proton beams. Materials and Methods: A diode array detector, MapCHECK (Model 1175, Sun Nuclear, Melbourne, FL, USA) was characterized in passive-scattered proton beams. The relative sensitivity of the diodes and absolute dose calibration were determined using a 250 MeV beam. The pristine Bragg curves (PBCs) measured by MapCHECK diodes were compared with those of an ion chamber using a range shift method. The water-equivalent thickness (WET) of the diode array detector’s intrinsic buildup also was determined. The inverse square dependence, linearity, and other proton dosimetric quantities measured by MapCHECK were also compared with those of the ion chambers. The change in the absolute dose response of the MapCHECK as a function of accumulated radiation dose was used as an indicator of radiation damage to the diodes. 2D dose distribution with and without the compensator were measured and compared with the treatment planning system (TPS) calculations. Results: The WET of the MapCHECK diode’s buildup was determined to be 1.7 cm. The MapCHECK-measured PBC were virtually identical to those measured by a parallel-plate ion chamber for 160, 180, and 250 MeV proton beams. The inverse square results of the MapCHECK were within ±0.4% of the ion chamber results. The linearity of MapCHECK results was within 1% of those from the ion chamber as measured in the range between 10 and 300 MU. All other dosimetric quantities were within 1.3% of the ion chamber results. The 2D dose distributions for non-clinical fields without compensator and the patient treatment fields with the compensator were consistent with the TPS results. The absolute dose response of the MapCHECK was changed by 7.4% after an accumulated dose increased by 170 Gy. Conclusions: The MapCHECK is a convenient and useful tool for 2D dose distribution measurements using passively scattered proton beams. Variations in MapCHECK’s dose response with increasing levels of total accumulated radiation dose should be carefully monitored.
format Online
Article
Text
id pubmed-4586777
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-45867772015-10-06 Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams Liengsawangwong, Praimakorn Sahoo, Nanayan Ding, Xiaoning Lii, MingFwu Gillin, Michale T. Zhu, Xiaorong Ronald Cancers (Basel) Article Purpose: To evaluate the dosimetric characteristics of a two-dimensional (2D) diode array detector irradiated with passively scattered proton beams. Materials and Methods: A diode array detector, MapCHECK (Model 1175, Sun Nuclear, Melbourne, FL, USA) was characterized in passive-scattered proton beams. The relative sensitivity of the diodes and absolute dose calibration were determined using a 250 MeV beam. The pristine Bragg curves (PBCs) measured by MapCHECK diodes were compared with those of an ion chamber using a range shift method. The water-equivalent thickness (WET) of the diode array detector’s intrinsic buildup also was determined. The inverse square dependence, linearity, and other proton dosimetric quantities measured by MapCHECK were also compared with those of the ion chambers. The change in the absolute dose response of the MapCHECK as a function of accumulated radiation dose was used as an indicator of radiation damage to the diodes. 2D dose distribution with and without the compensator were measured and compared with the treatment planning system (TPS) calculations. Results: The WET of the MapCHECK diode’s buildup was determined to be 1.7 cm. The MapCHECK-measured PBC were virtually identical to those measured by a parallel-plate ion chamber for 160, 180, and 250 MeV proton beams. The inverse square results of the MapCHECK were within ±0.4% of the ion chamber results. The linearity of MapCHECK results was within 1% of those from the ion chamber as measured in the range between 10 and 300 MU. All other dosimetric quantities were within 1.3% of the ion chamber results. The 2D dose distributions for non-clinical fields without compensator and the patient treatment fields with the compensator were consistent with the TPS results. The absolute dose response of the MapCHECK was changed by 7.4% after an accumulated dose increased by 170 Gy. Conclusions: The MapCHECK is a convenient and useful tool for 2D dose distribution measurements using passively scattered proton beams. Variations in MapCHECK’s dose response with increasing levels of total accumulated radiation dose should be carefully monitored. MDPI 2015-07-30 /pmc/articles/PMC4586777/ /pubmed/26264024 http://dx.doi.org/10.3390/cancers7030844 Text en © 2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Liengsawangwong, Praimakorn
Sahoo, Nanayan
Ding, Xiaoning
Lii, MingFwu
Gillin, Michale T.
Zhu, Xiaorong Ronald
Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams
title Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams
title_full Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams
title_fullStr Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams
title_full_unstemmed Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams
title_short Dosimetric Characteristics of a Two-Dimensional Diode Array Detector Irradiated with Passively Scattered Proton Beams
title_sort dosimetric characteristics of a two-dimensional diode array detector irradiated with passively scattered proton beams
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4586777/
https://www.ncbi.nlm.nih.gov/pubmed/26264024
http://dx.doi.org/10.3390/cancers7030844
work_keys_str_mv AT liengsawangwongpraimakorn dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams
AT sahoonanayan dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams
AT dingxiaoning dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams
AT liimingfwu dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams
AT gillinmichalet dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams
AT zhuxiaorongronald dosimetriccharacteristicsofatwodimensionaldiodearraydetectorirradiatedwithpassivelyscatteredprotonbeams