Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts

Cardiac MRI may benefit from increased polarization at high magnetic field strength of 3 Tesla but is challenged by increased field inhomogeneity. Initial human studies have shown that the radiofrequency (RF) excitation field (B(1) (+)) used for signal excitation in the heart is both inhomogeneous a...

Descripción completa

Detalles Bibliográficos
Autores principales: Schär, Michael, Ding, Haiyan, Herzka, Daniel A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4593605/
https://www.ncbi.nlm.nih.gov/pubmed/26436658
http://dx.doi.org/10.1371/journal.pone.0139859
_version_ 1782393348950065152
author Schär, Michael
Ding, Haiyan
Herzka, Daniel A.
author_facet Schär, Michael
Ding, Haiyan
Herzka, Daniel A.
author_sort Schär, Michael
collection PubMed
description Cardiac MRI may benefit from increased polarization at high magnetic field strength of 3 Tesla but is challenged by increased field inhomogeneity. Initial human studies have shown that the radiofrequency (RF) excitation field (B(1) (+)) used for signal excitation in the heart is both inhomogeneous and significantly lower than desired, potentially leading to image artifacts and biased quantitative measures. Recently, multi-channel transmit systems have been introduced allowing localized patient specific RF shimming based on acquired calibration B(1) (+) maps. Some prior human studies have shown lower than desired mean flip angles in the hearts of large patients even after RF shimming. Here, 100 cardiac B(1) (+) map pairs before and after RF shimming were acquired in 55 swine. The mean flip angle and the coefficient of variation (CV) of the flip angle in the heart were determined before and after RF shimming. Mean flip angle, CV, and RF shim values (power ratio and phase difference between the two transmit channels) were tested for correlation with cross sectional body area and the Right-Left/Anterior-Posterior ratio. RF shimming significantly increased the mean flip angle in swine heart from 74.4±6.7% (mean ± standard deviation) to 94.7±4.8% of the desired flip angle and significantly reduced CV from 0.11±0.03 to 0.07±0.02 (p<<1e-10 for both). These results compare well with several previous human studies, except that the mean flip angle in the human heart only improved to 89% with RF shimming, possibly because the RF shimming routine does not consider safety constraints in very large patients. Additionally, mean flip angle decreased and CV increased with larger cross sectional body area, however, the RF shimming parameters did not correlate with cross sectional body area. RF shim power ratio correlated weakly with Right-Left/Anterior-Posterior ratio but phase difference did not, further substantiating the need for subject specific cardiac RF shimming.
format Online
Article
Text
id pubmed-4593605
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-45936052015-10-14 Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts Schär, Michael Ding, Haiyan Herzka, Daniel A. PLoS One Research Article Cardiac MRI may benefit from increased polarization at high magnetic field strength of 3 Tesla but is challenged by increased field inhomogeneity. Initial human studies have shown that the radiofrequency (RF) excitation field (B(1) (+)) used for signal excitation in the heart is both inhomogeneous and significantly lower than desired, potentially leading to image artifacts and biased quantitative measures. Recently, multi-channel transmit systems have been introduced allowing localized patient specific RF shimming based on acquired calibration B(1) (+) maps. Some prior human studies have shown lower than desired mean flip angles in the hearts of large patients even after RF shimming. Here, 100 cardiac B(1) (+) map pairs before and after RF shimming were acquired in 55 swine. The mean flip angle and the coefficient of variation (CV) of the flip angle in the heart were determined before and after RF shimming. Mean flip angle, CV, and RF shim values (power ratio and phase difference between the two transmit channels) were tested for correlation with cross sectional body area and the Right-Left/Anterior-Posterior ratio. RF shimming significantly increased the mean flip angle in swine heart from 74.4±6.7% (mean ± standard deviation) to 94.7±4.8% of the desired flip angle and significantly reduced CV from 0.11±0.03 to 0.07±0.02 (p<<1e-10 for both). These results compare well with several previous human studies, except that the mean flip angle in the human heart only improved to 89% with RF shimming, possibly because the RF shimming routine does not consider safety constraints in very large patients. Additionally, mean flip angle decreased and CV increased with larger cross sectional body area, however, the RF shimming parameters did not correlate with cross sectional body area. RF shim power ratio correlated weakly with Right-Left/Anterior-Posterior ratio but phase difference did not, further substantiating the need for subject specific cardiac RF shimming. Public Library of Science 2015-10-05 /pmc/articles/PMC4593605/ /pubmed/26436658 http://dx.doi.org/10.1371/journal.pone.0139859 Text en © 2015 Schär et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Schär, Michael
Ding, Haiyan
Herzka, Daniel A.
Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts
title Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts
title_full Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts
title_fullStr Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts
title_full_unstemmed Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts
title_short Improvement in B(1) (+) Homogeneity and Average Flip Angle Using Dual-Source Parallel RF Excitation for Cardiac MRI in Swine Hearts
title_sort improvement in b(1) (+) homogeneity and average flip angle using dual-source parallel rf excitation for cardiac mri in swine hearts
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4593605/
https://www.ncbi.nlm.nih.gov/pubmed/26436658
http://dx.doi.org/10.1371/journal.pone.0139859
work_keys_str_mv AT scharmichael improvementinb1homogeneityandaverageflipangleusingdualsourceparallelrfexcitationforcardiacmriinswinehearts
AT dinghaiyan improvementinb1homogeneityandaverageflipangleusingdualsourceparallelrfexcitationforcardiacmriinswinehearts
AT herzkadaniela improvementinb1homogeneityandaverageflipangleusingdualsourceparallelrfexcitationforcardiacmriinswinehearts