Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission

The present work validates the Geant4 Application for Tomographic Emission Monte Carlo software for the simulation of a 6 MV photon beam given by Elekta Synergy Platform medical linear accelerator treatment head. The simulation includes the major components of the linear accelerator (LINAC) with mul...

Descripción completa

Detalles Bibliográficos
Autores principales: Didi, Samir, Moussa, Abdelilah, Yahya, Tayalati, Mustafa, Zerfaoui
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4594382/
https://www.ncbi.nlm.nih.gov/pubmed/26500399
http://dx.doi.org/10.4103/0971-6203.165077
_version_ 1782393443036692480
author Didi, Samir
Moussa, Abdelilah
Yahya, Tayalati
Mustafa, Zerfaoui
author_facet Didi, Samir
Moussa, Abdelilah
Yahya, Tayalati
Mustafa, Zerfaoui
author_sort Didi, Samir
collection PubMed
description The present work validates the Geant4 Application for Tomographic Emission Monte Carlo software for the simulation of a 6 MV photon beam given by Elekta Synergy Platform medical linear accelerator treatment head. The simulation includes the major components of the linear accelerator (LINAC) with multi-leaf collimator and a homogeneous water phantom. Calculations were performed for the photon beam with several treatment field sizes ranging from 5 cm × 5 cm to 30 cm × 30 cm at 100 cm distance from the source. The simulation was successfully validated by comparison with experimental distributions. Good agreement between simulations and measurements was observed, with dose differences of about 0.02% and 2.5% for depth doses and lateral dose profiles, respectively. This agreement was also emphasized by the Kolmogorov–Smirnov goodness-of-fit test and by the gamma-index comparisons where more than 99% of the points for all simulations fulfill the quality assurance criteria of 2 mm/2%.
format Online
Article
Text
id pubmed-4594382
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-45943822015-10-23 Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission Didi, Samir Moussa, Abdelilah Yahya, Tayalati Mustafa, Zerfaoui J Med Phys Original Article The present work validates the Geant4 Application for Tomographic Emission Monte Carlo software for the simulation of a 6 MV photon beam given by Elekta Synergy Platform medical linear accelerator treatment head. The simulation includes the major components of the linear accelerator (LINAC) with multi-leaf collimator and a homogeneous water phantom. Calculations were performed for the photon beam with several treatment field sizes ranging from 5 cm × 5 cm to 30 cm × 30 cm at 100 cm distance from the source. The simulation was successfully validated by comparison with experimental distributions. Good agreement between simulations and measurements was observed, with dose differences of about 0.02% and 2.5% for depth doses and lateral dose profiles, respectively. This agreement was also emphasized by the Kolmogorov–Smirnov goodness-of-fit test and by the gamma-index comparisons where more than 99% of the points for all simulations fulfill the quality assurance criteria of 2 mm/2%. Medknow Publications & Media Pvt Ltd 2015 /pmc/articles/PMC4594382/ /pubmed/26500399 http://dx.doi.org/10.4103/0971-6203.165077 Text en Copyright: © 2015 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Didi, Samir
Moussa, Abdelilah
Yahya, Tayalati
Mustafa, Zerfaoui
Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission
title Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission
title_full Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission
title_fullStr Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission
title_full_unstemmed Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission
title_short Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission
title_sort simulation of the 6 mv elekta synergy platform linac photon beam using geant4 application for tomographic emission
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4594382/
https://www.ncbi.nlm.nih.gov/pubmed/26500399
http://dx.doi.org/10.4103/0971-6203.165077
work_keys_str_mv AT didisamir simulationofthe6mvelektasynergyplatformlinacphotonbeamusinggeant4applicationfortomographicemission
AT moussaabdelilah simulationofthe6mvelektasynergyplatformlinacphotonbeamusinggeant4applicationfortomographicemission
AT yahyatayalati simulationofthe6mvelektasynergyplatformlinacphotonbeamusinggeant4applicationfortomographicemission
AT mustafazerfaoui simulationofthe6mvelektasynergyplatformlinacphotonbeamusinggeant4applicationfortomographicemission