Improving Depiction of Temporal Bone Anatomy With Low-Radiation Dose CT by an Integrated Circuit Detector in Pediatric Patients: A Preliminary Study

The purpose of this study was to determine the performance of low-dose computed tomography (CT) scanning with integrated circuit (IC) detector in defining fine structures of temporal bone in children by comparing with the conventional detector. The study was performed with the approval of our instit...

Descripción completa

Detalles Bibliográficos
Autores principales: He, Jingzhen, Zu, Yuliang, Wang, Qing, Ma, Xiangxing
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer Health 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4603126/
https://www.ncbi.nlm.nih.gov/pubmed/25526489
http://dx.doi.org/10.1097/MD.0000000000000325
Descripción
Sumario:The purpose of this study was to determine the performance of low-dose computed tomography (CT) scanning with integrated circuit (IC) detector in defining fine structures of temporal bone in children by comparing with the conventional detector. The study was performed with the approval of our institutional review board and the patients’ anonymity was maintained. A total of 86 children <3 years of age underwent imaging of temporal bone with low-dose CT (80 kV/150 mAs) equipped with either IC detector or conventional discrete circuit (DC) detector. The image noise was measured for quantitative analysis. Thirty-five structures of temporal bone were further assessed and rated by 2 radiologists for qualitative analysis. κ Statistics were performed to determine the agreement reached between the 2 radiologists on each image. Mann–Whitney U test was used to determine the difference in image quality between the 2 detector systems. Objective analysis showed that the image noise was significantly lower (P < 0.001) with the IC detector than with the DC detector. The κ values for qualitative assessment of the 35 fine anatomical structures revealed high interobserver agreement. The delineation for 30 of the 35 landmarks (86%) with the IC detector was superior to that with the conventional DC detector (P < 0.05) although there were no differences in the delineation of the remaining 5 structures (P > 0.05). The low-dose CT images acquired with the IC detector provide better depiction of fine osseous structures of temporal bone than that with the conventional DC detector.