Cargando…

Novel strategy of endoscopic submucosal dissection using an insulation-tipped knife for early gastric cancer: near-side approach method

Background and study aims: Endoscopic submucosal dissection (ESD) using insulation-tipped knives (IT knives) to treat gastric lesions located on the greater curvature of the gastric body remains technically challenging because of the associated bleeding, control of which can be difficult and time co...

Descripción completa

Detalles Bibliográficos
Autores principales: Mori, Genki, Nonaka, Satoru, Oda, Ichiro, Abe, Seiichiro, Suzuki, Haruhisa, Yoshinaga, Shigetaka, Nakajima, Takeshi, Saito, Yutaka
Formato: Online Artículo Texto
Lenguaje:English
Publicado: © Georg Thieme Verlag KG 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4612249/
https://www.ncbi.nlm.nih.gov/pubmed/26528496
http://dx.doi.org/10.1055/s-0034-1392567
Descripción
Sumario:Background and study aims: Endoscopic submucosal dissection (ESD) using insulation-tipped knives (IT knives) to treat gastric lesions located on the greater curvature of the gastric body remains technically challenging because of the associated bleeding, control of which can be difficult and time consuming. To eliminate these difficulties, we developed a novel strategy which we have called the “near-side approach method” and assessed its utility. Patients and methods: We reviewed patients who underwent ESD for solitary early gastric cancer located on the greater curvature of the gastric body from January 2003 to September 2014. The technical results of ESD were compared between the group treated with the novel near-side approach method and the group treated with the conventional method. Results: This study included 238 patients with 238 lesions, 118 of which were removed using the near-side approach method and 120 of which were removed using the conventional method. The median procedure time was 92 minutes for the near-side approach method and 120 minutes for the conventional method. The procedure time was significantly shorter in the near-side approach method arm. Although, the procedure time required by an experienced endoscopist was not significantly different between the two groups (100 vs. 110 minutes), the near-side approach group showed significantly shorter procedure time for a less-experienced endoscopist (90 vs. 120 minutes). Conclusions: The near-side approach method appears to require less time to complete gastric ESD than the conventional method using IT knives for technically challenging lesions located on the greater curvature of the gastric body, especially if the procedure is performed by less-experienced endoscopists.