Cargando…

Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority

Background. In March 2010, ANSM (Agence Nationale de Sécurité du Medicament), the French Medical Regulatory Authority, withdrew Poly Implant Prothèse (PIP) breast implants from the market due to the use of non-medical-grade silicone gel. The aim of this study was to compare the removal rate (and rea...

Descripción completa

Detalles Bibliográficos
Autores principales: Leduey, Alexandre, Mazouni, Chafika, Leymarie, Nicolas, Alkhashnam, Heba, Sarfati, Benjamin, Garbay, Jean-Rémi, Gaudin, Amélie, Kolb, Frédéric, Rimareix, Françoise
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4620385/
https://www.ncbi.nlm.nih.gov/pubmed/26543648
http://dx.doi.org/10.1155/2015/519497
_version_ 1782397286431588352
author Leduey, Alexandre
Mazouni, Chafika
Leymarie, Nicolas
Alkhashnam, Heba
Sarfati, Benjamin
Garbay, Jean-Rémi
Gaudin, Amélie
Kolb, Frédéric
Rimareix, Françoise
author_facet Leduey, Alexandre
Mazouni, Chafika
Leymarie, Nicolas
Alkhashnam, Heba
Sarfati, Benjamin
Garbay, Jean-Rémi
Gaudin, Amélie
Kolb, Frédéric
Rimareix, Françoise
author_sort Leduey, Alexandre
collection PubMed
description Background. In March 2010, ANSM (Agence Nationale de Sécurité du Medicament), the French Medical Regulatory Authority, withdrew Poly Implant Prothèse (PIP) breast implants from the market due to the use of non-medical-grade silicone gel. The aim of this study was to compare the removal rate (and reasons thereof) of breast implants produced by different manufacturers before the ANSM alert. Materials and Methods. From October 2006 to January 2010, 652 women received 944 implants after breast cancer surgery at the Gustave Roussy Comprehensive Cancer Center, Paris (France). The complications and removal rates of the different implant brands used (PIP, Allergan, and Pérouse) were evaluated and compared. Results. PIP implants represented 50.6% of the used implants, Allergan 33.4%, and Pérouse 16%. The main reasons for implant removal were patient dissatisfaction due to aesthetic problems (43.2%), infection (22.2%), and capsular contracture (13.6%). Two years after implantation, 82% of Pérouse implants, 79% of PIP, and 79% of Allergan were still in situ. There was no difference in removal rate among implant brands. Conclusion. Before the ANSM alert concerning the higher rupture rate of PIP breast implants, our implant removal rate did not predict PIP implant failure related to the use of nonapproved silicone gel.
format Online
Article
Text
id pubmed-4620385
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-46203852015-11-05 Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority Leduey, Alexandre Mazouni, Chafika Leymarie, Nicolas Alkhashnam, Heba Sarfati, Benjamin Garbay, Jean-Rémi Gaudin, Amélie Kolb, Frédéric Rimareix, Françoise Int J Breast Cancer Clinical Study Background. In March 2010, ANSM (Agence Nationale de Sécurité du Medicament), the French Medical Regulatory Authority, withdrew Poly Implant Prothèse (PIP) breast implants from the market due to the use of non-medical-grade silicone gel. The aim of this study was to compare the removal rate (and reasons thereof) of breast implants produced by different manufacturers before the ANSM alert. Materials and Methods. From October 2006 to January 2010, 652 women received 944 implants after breast cancer surgery at the Gustave Roussy Comprehensive Cancer Center, Paris (France). The complications and removal rates of the different implant brands used (PIP, Allergan, and Pérouse) were evaluated and compared. Results. PIP implants represented 50.6% of the used implants, Allergan 33.4%, and Pérouse 16%. The main reasons for implant removal were patient dissatisfaction due to aesthetic problems (43.2%), infection (22.2%), and capsular contracture (13.6%). Two years after implantation, 82% of Pérouse implants, 79% of PIP, and 79% of Allergan were still in situ. There was no difference in removal rate among implant brands. Conclusion. Before the ANSM alert concerning the higher rupture rate of PIP breast implants, our implant removal rate did not predict PIP implant failure related to the use of nonapproved silicone gel. Hindawi Publishing Corporation 2015 2015-10-12 /pmc/articles/PMC4620385/ /pubmed/26543648 http://dx.doi.org/10.1155/2015/519497 Text en Copyright © 2015 Alexandre Leduey et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Clinical Study
Leduey, Alexandre
Mazouni, Chafika
Leymarie, Nicolas
Alkhashnam, Heba
Sarfati, Benjamin
Garbay, Jean-Rémi
Gaudin, Amélie
Kolb, Frédéric
Rimareix, Françoise
Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority
title Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority
title_full Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority
title_fullStr Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority
title_full_unstemmed Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority
title_short Comparison of the Explantation Rate of Poly Implant Prothèse, Allergan, and Pérouse Silicone Breast Implants within the First Four Years after Reconstructive Surgery before the Poly Implant Prothèse Alert by the French Regulatory Authority
title_sort comparison of the explantation rate of poly implant prothèse, allergan, and pérouse silicone breast implants within the first four years after reconstructive surgery before the poly implant prothèse alert by the french regulatory authority
topic Clinical Study
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4620385/
https://www.ncbi.nlm.nih.gov/pubmed/26543648
http://dx.doi.org/10.1155/2015/519497
work_keys_str_mv AT ledueyalexandre comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT mazounichafika comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT leymarienicolas comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT alkhashnamheba comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT sarfatibenjamin comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT garbayjeanremi comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT gaudinamelie comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT kolbfrederic comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority
AT rimareixfrancoise comparisonoftheexplantationrateofpolyimplantprotheseallerganandperousesiliconebreastimplantswithinthefirstfouryearsafterreconstructivesurgerybeforethepolyimplantprothesealertbythefrenchregulatoryauthority