A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry
INTRODUCTION: Total 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status. In this study, we compared two automated immunoassay methods, the Abbott Architect 25-OH Vitamin D assay and the Roche Cobas Vitamin D total assay, with the liquid chromatography-tandem mass spectro...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4622190/ https://www.ncbi.nlm.nih.gov/pubmed/26526462 http://dx.doi.org/10.11613/BM.2015.044 |
_version_ | 1782397558099804160 |
---|---|
author | Kocak, Fatma Emel Ozturk, Bahadir Isiklar, Ozben Ozden Genc, Ozlem Unlu, Ali Altuntas, Irfan |
author_facet | Kocak, Fatma Emel Ozturk, Bahadir Isiklar, Ozben Ozden Genc, Ozlem Unlu, Ali Altuntas, Irfan |
author_sort | Kocak, Fatma Emel |
collection | PubMed |
description | INTRODUCTION: Total 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status. In this study, we compared two automated immunoassay methods, the Abbott Architect 25-OH Vitamin D assay and the Roche Cobas Vitamin D total assay, with the liquid chromatography-tandem mass spectrometry (LC-MS/MS). MATERIALS AND METHODS: One hundred venous blood samples were randomly selected from routine vitamin D tests. Two of the serum aliquots were analyzed at the Abbott Architect i2000 and the Roche Cobas 6000’s module e601 in our laboratory within the same day. The other serum aliquots were analyzed at the LC-MS/MS in different laboratory. Passing-Bablok regression analysis and Bland-Altman plot were used to compare methods. Inter-rater agreement was analyzed using kappa (κ) analysis. RESULTS: The Roche assay showed acceptable agreement with the LC-MS/MS based on Passing-Bablok analysis (intercept: -5.23 nmol/L, 95% CI: -8.73 to 0.19; slope: 0.97, 95% CI: 0.77 to 1.15). The Abbott assay showed proportional (slope: 0.77, 95% CI: 0.67 to 0.85) and constant differences (intercept: 17.08 nmol/L; 95% CI: 12.98 to 21.39). A mean bias of 15.1% was observed for the Abbott and a mean bias of -14.1% was observed for the Roche based on the Bland-Altman plots. We found strong to nearly perfect agreement in vitamin D status between the immunoassays and LC-MS/MS. (κ: 0.83 for Abbott, κ: 0.93 for Roche) using kappa analysis. CONCLUSION: Both immunoassays demonstrated acceptable performance, but the Roche Cobas assay demonstrated better performance than the Abbott Architect in the studied samples. |
format | Online Article Text |
id | pubmed-4622190 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-46221902015-10-30 A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry Kocak, Fatma Emel Ozturk, Bahadir Isiklar, Ozben Ozden Genc, Ozlem Unlu, Ali Altuntas, Irfan Biochem Med (Zagreb) Research Article INTRODUCTION: Total 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status. In this study, we compared two automated immunoassay methods, the Abbott Architect 25-OH Vitamin D assay and the Roche Cobas Vitamin D total assay, with the liquid chromatography-tandem mass spectrometry (LC-MS/MS). MATERIALS AND METHODS: One hundred venous blood samples were randomly selected from routine vitamin D tests. Two of the serum aliquots were analyzed at the Abbott Architect i2000 and the Roche Cobas 6000’s module e601 in our laboratory within the same day. The other serum aliquots were analyzed at the LC-MS/MS in different laboratory. Passing-Bablok regression analysis and Bland-Altman plot were used to compare methods. Inter-rater agreement was analyzed using kappa (κ) analysis. RESULTS: The Roche assay showed acceptable agreement with the LC-MS/MS based on Passing-Bablok analysis (intercept: -5.23 nmol/L, 95% CI: -8.73 to 0.19; slope: 0.97, 95% CI: 0.77 to 1.15). The Abbott assay showed proportional (slope: 0.77, 95% CI: 0.67 to 0.85) and constant differences (intercept: 17.08 nmol/L; 95% CI: 12.98 to 21.39). A mean bias of 15.1% was observed for the Abbott and a mean bias of -14.1% was observed for the Roche based on the Bland-Altman plots. We found strong to nearly perfect agreement in vitamin D status between the immunoassays and LC-MS/MS. (κ: 0.83 for Abbott, κ: 0.93 for Roche) using kappa analysis. CONCLUSION: Both immunoassays demonstrated acceptable performance, but the Roche Cobas assay demonstrated better performance than the Abbott Architect in the studied samples. Croatian Society of Medical Biochemistry and Laboratory Medicine 2015-10-15 /pmc/articles/PMC4622190/ /pubmed/26526462 http://dx.doi.org/10.11613/BM.2015.044 Text en |
spellingShingle | Research Article Kocak, Fatma Emel Ozturk, Bahadir Isiklar, Ozben Ozden Genc, Ozlem Unlu, Ali Altuntas, Irfan A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry |
title | A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry |
title_full | A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry |
title_fullStr | A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry |
title_full_unstemmed | A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry |
title_short | A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry |
title_sort | comparison between two different automated total 25-hydroxyvitamin d immunoassay methods using liquid chromatography-tandem mass spectrometry |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4622190/ https://www.ncbi.nlm.nih.gov/pubmed/26526462 http://dx.doi.org/10.11613/BM.2015.044 |
work_keys_str_mv | AT kocakfatmaemel acomparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT ozturkbahadir acomparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT isiklarozbenozden acomparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT gencozlem acomparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT unluali acomparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT altuntasirfan acomparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT kocakfatmaemel comparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT ozturkbahadir comparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT isiklarozbenozden comparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT gencozlem comparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT unluali comparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry AT altuntasirfan comparisonbetweentwodifferentautomatedtotal25hydroxyvitamindimmunoassaymethodsusingliquidchromatographytandemmassspectrometry |