Cargando…
In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance
A quartz crystal microbalance (QCM) was utilized to measure the water content in ethanol. For the improvement of measurement sensitivity, the QCM was modified by applying zeolite particles on the surface with poly(methyl methacrylate) (PMMA) binder. The measurement performance was examined with etha...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4634433/ https://www.ncbi.nlm.nih.gov/pubmed/26516859 http://dx.doi.org/10.3390/s151027273 |
_version_ | 1782399354895597568 |
---|---|
author | Kim, Byoung Chul Yamamoto, Takuji Kim, Young Han |
author_facet | Kim, Byoung Chul Yamamoto, Takuji Kim, Young Han |
author_sort | Kim, Byoung Chul |
collection | PubMed |
description | A quartz crystal microbalance (QCM) was utilized to measure the water content in ethanol. For the improvement of measurement sensitivity, the QCM was modified by applying zeolite particles on the surface with poly(methyl methacrylate) (PMMA) binder. The measurement performance was examined with ethanol of 1% to 5% water content in circulation. The experimental results showed that the frequency drop of the QCM was related with the water content though there was some deviation. The sensitivity of the zeolite-coated QCM was sufficient to be implemented in water content determination, and a higher ratio of silicon to aluminum in the molecular structure of the zeolite gave better performance. The coated surface was inspected by microscopy to show the distribution of zeolite particles and PMMA spread. |
format | Online Article Text |
id | pubmed-4634433 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-46344332015-11-23 In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance Kim, Byoung Chul Yamamoto, Takuji Kim, Young Han Sensors (Basel) Article A quartz crystal microbalance (QCM) was utilized to measure the water content in ethanol. For the improvement of measurement sensitivity, the QCM was modified by applying zeolite particles on the surface with poly(methyl methacrylate) (PMMA) binder. The measurement performance was examined with ethanol of 1% to 5% water content in circulation. The experimental results showed that the frequency drop of the QCM was related with the water content though there was some deviation. The sensitivity of the zeolite-coated QCM was sufficient to be implemented in water content determination, and a higher ratio of silicon to aluminum in the molecular structure of the zeolite gave better performance. The coated surface was inspected by microscopy to show the distribution of zeolite particles and PMMA spread. MDPI 2015-10-27 /pmc/articles/PMC4634433/ /pubmed/26516859 http://dx.doi.org/10.3390/s151027273 Text en © 2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Kim, Byoung Chul Yamamoto, Takuji Kim, Young Han In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance |
title | In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance |
title_full | In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance |
title_fullStr | In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance |
title_full_unstemmed | In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance |
title_short | In-Line Measurement of Water Contents in Ethanol Using a Zeolite-Coated Quartz Crystal Microbalance |
title_sort | in-line measurement of water contents in ethanol using a zeolite-coated quartz crystal microbalance |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4634433/ https://www.ncbi.nlm.nih.gov/pubmed/26516859 http://dx.doi.org/10.3390/s151027273 |
work_keys_str_mv | AT kimbyoungchul inlinemeasurementofwatercontentsinethanolusingazeolitecoatedquartzcrystalmicrobalance AT yamamototakuji inlinemeasurementofwatercontentsinethanolusingazeolitecoatedquartzcrystalmicrobalance AT kimyounghan inlinemeasurementofwatercontentsinethanolusingazeolitecoatedquartzcrystalmicrobalance |