2D cine DENSE with low encoding frequencies accurately quantifies cardiac mechanics with improved image characteristics

BACKGROUND: Displacement Encoding with Stimulated Echoes (DENSE) encodes displacement into the phase of the magnetic resonance signal. The encoding frequency (k(e)) maps the measured phase to tissue displacement while the strength of the encoding gradients affects image quality. 2D cine DENSE studie...

Descripción completa

Detalles Bibliográficos
Autores principales: Wehner, Gregory J., Grabau, Jonathan D., Suever, Jonathan D., Haggerty, Christopher M., Jing, Linyuan, Powell, David K., Hamlet, Sean M., Vandsburger, Moriel H., Zhong, Xiaodong, Fornwalt, Brandon K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4634910/
https://www.ncbi.nlm.nih.gov/pubmed/26538111
http://dx.doi.org/10.1186/s12968-015-0196-z
Descripción
Sumario:BACKGROUND: Displacement Encoding with Stimulated Echoes (DENSE) encodes displacement into the phase of the magnetic resonance signal. The encoding frequency (k(e)) maps the measured phase to tissue displacement while the strength of the encoding gradients affects image quality. 2D cine DENSE studies have used a k(e) of 0.10 cycles/mm, which is high enough to remove an artifact-generating echo from k-space, provide high sensitivity to tissue displacements, and dephase the blood pool. However, through-plane dephasing can remove the unwanted echo and dephase the blood pool without relying on high k(e). Additionally, the high sensitivity comes with the costs of increased phase wrapping and intra-voxel dephasing. We hypothesized that k(e) below 0.10 cycles/mm can be used to improve image characteristics and provide accurate measures of cardiac mechanics. METHODS: Spiral cine DENSE images were obtained for 10 healthy subjects and 10 patients with a history of heart disease on a 3 T Siemens Trio. A mid-ventricular short-axis image was acquired with different k(e): 0.02, 0.04, 0.06, 0.08, and 0.10 cycles/mm. Peak twist, circumferential strain, and radial strain were compared between acquisitions employing different k(e) using Bland-Altman analyses and coefficients of variation. The percentage of wrapped pixels in the phase images at end-systole was calculated for each k(e). The dephasing of the blood signal and signal to noise ratio (SNR) were also calculated and compared. RESULTS: Negligible differences were seen in strains and twist for all k(e) between 0.04 and 0.10 cycles/mm. These differences were of the same magnitude as inter-test differences. Specifically, the acquisitions with 0.04 cycles/mm accurately quantified cardiac mechanics and had zero phase wrapping. Compared to 0.10 cycles/mm, the acquisitions with 0.04 cycles/mm had 9 % greater SNR and negligible differences in blood pool dephasing. CONCLUSIONS: For 2D cine DENSE with through-plane dephasing, the encoding frequency can be lowered to 0.04 cycles/mm without compromising the quantification of twist or strain. The amount of wrapping can be reduced with this lower value to greatly simplify the input to unwrapping algorithms. The strain and twist results from studies using different encoding frequencies can be directly compared.