Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation

This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist. Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (stand...

Descripción completa

Detalles Bibliográficos
Autores principales: Cao, Xiaoqin, Song, Chunhua, Guo, Leilei, Yang, Junxia, Deng, Songyuan, Xu, Yajuan, Chen, Xiaolin, Sapa, Wolde Bekalo, Wang, Kaijuan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer Health 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4635802/
https://www.ncbi.nlm.nih.gov/pubmed/26376388
http://dx.doi.org/10.1097/MD.0000000000001475
_version_ 1782399559402520576
author Cao, Xiaoqin
Song, Chunhua
Guo, Leilei
Yang, Junxia
Deng, Songyuan
Xu, Yajuan
Chen, Xiaolin
Sapa, Wolde Bekalo
Wang, Kaijuan
author_facet Cao, Xiaoqin
Song, Chunhua
Guo, Leilei
Yang, Junxia
Deng, Songyuan
Xu, Yajuan
Chen, Xiaolin
Sapa, Wolde Bekalo
Wang, Kaijuan
author_sort Cao, Xiaoqin
collection PubMed
description This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist. Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (standard device). Two percent of subjects enrolled in an epidemiological investigation were systematically sampled in this study. Intraclass correlation coefficient (ICC) was used to evaluate measurement reliability, paired t-test was used to evaluate trueness, and linear regression was used for calibration. The Association for the Advancement of Medical Instrumentation (AAMI) standards and British Hypertension Society (BHS) protocols were used for validation quality assessment. Both mercury sphygmomanometer (standard device) and oscillometric BP monitor (test device) displayed high reliability. A significant difference in systolic blood pressure (SBP) was observed between devices. SBP calibration was achieved by using an effective linear regression model (B = 0.803 and constant = 19.592, P < 0.001). The calibrated model was corroborated by verification samples (P = 0.120) and was found to pass AAMI standards and BHS protocol requirements. Calibrated SBP measurements from the Omron HBP-1300 device were valid. Use of a combination of statistical methods, such as ICC for reliability assessment as well as paired t-test for trueness evaluation can be used to validate data from the oscillometric BP monitors.
format Online
Article
Text
id pubmed-4635802
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Wolters Kluwer Health
record_format MEDLINE/PubMed
spelling pubmed-46358022015-11-30 Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation Cao, Xiaoqin Song, Chunhua Guo, Leilei Yang, Junxia Deng, Songyuan Xu, Yajuan Chen, Xiaolin Sapa, Wolde Bekalo Wang, Kaijuan Medicine (Baltimore) 6600 This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist. Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (standard device). Two percent of subjects enrolled in an epidemiological investigation were systematically sampled in this study. Intraclass correlation coefficient (ICC) was used to evaluate measurement reliability, paired t-test was used to evaluate trueness, and linear regression was used for calibration. The Association for the Advancement of Medical Instrumentation (AAMI) standards and British Hypertension Society (BHS) protocols were used for validation quality assessment. Both mercury sphygmomanometer (standard device) and oscillometric BP monitor (test device) displayed high reliability. A significant difference in systolic blood pressure (SBP) was observed between devices. SBP calibration was achieved by using an effective linear regression model (B = 0.803 and constant = 19.592, P < 0.001). The calibrated model was corroborated by verification samples (P = 0.120) and was found to pass AAMI standards and BHS protocol requirements. Calibrated SBP measurements from the Omron HBP-1300 device were valid. Use of a combination of statistical methods, such as ICC for reliability assessment as well as paired t-test for trueness evaluation can be used to validate data from the oscillometric BP monitors. Wolters Kluwer Health 2015-09-18 /pmc/articles/PMC4635802/ /pubmed/26376388 http://dx.doi.org/10.1097/MD.0000000000001475 Text en Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved. http://creativecommons.org/licenses/by-nc/4.0 This is an open access article distributed under the Creative Commons Attribution-NonCommercial License, where it is permissible to download, share and reproduce the work in any medium, provided it is properly cited. The work cannot be used commercially. http://creativecommons.org/licenses/by-nc/4.0
spellingShingle 6600
Cao, Xiaoqin
Song, Chunhua
Guo, Leilei
Yang, Junxia
Deng, Songyuan
Xu, Yajuan
Chen, Xiaolin
Sapa, Wolde Bekalo
Wang, Kaijuan
Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
title Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
title_full Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
title_fullStr Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
title_full_unstemmed Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
title_short Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation
title_sort quality control and validation of oscillometric blood pressure measurements taken during an epidemiological investigation
topic 6600
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4635802/
https://www.ncbi.nlm.nih.gov/pubmed/26376388
http://dx.doi.org/10.1097/MD.0000000000001475
work_keys_str_mv AT caoxiaoqin qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT songchunhua qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT guoleilei qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT yangjunxia qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT dengsongyuan qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT xuyajuan qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT chenxiaolin qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT sapawoldebekalo qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation
AT wangkaijuan qualitycontrolandvalidationofoscillometricbloodpressuremeasurementstakenduringanepidemiologicalinvestigation