Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment

To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging. MATERIALS AND METHODS: A phantom with 8 cells in periphery region and 1 cell in central region were used. Th...

Descripción completa

Detalles Bibliográficos
Autores principales: He, Tian, Qian, Xiaojun, Zhai, Renyou, Yang, Zongtao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4654266/
https://www.ncbi.nlm.nih.gov/pubmed/26196347
http://dx.doi.org/10.1097/RCT.0000000000000287
_version_ 1782402032543465472
author He, Tian
Qian, Xiaojun
Zhai, Renyou
Yang, Zongtao
author_facet He, Tian
Qian, Xiaojun
Zhai, Renyou
Yang, Zongtao
author_sort He, Tian
collection PubMed
description To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging. MATERIALS AND METHODS: A phantom with 8 cells in periphery region and 1 cell in central region were used. The 8 conditioning tubes in the periphery region were filled with 1 of the 3 iodine solutions to simulate different beam hardening conditions: 0 for no beam hardening (NBH), 20 mg/mL for weak beam hardening (WBH) and 50 mg/mL for severe beam hardening (SBH) condition. Test tube filled with 0, 0.1, 0.5, 1, 2, 5, 10, 20, and 50 mg/mL iodine solution was placed in the central cell alternately. The phantom was scanned with conventional CT mode with 80, 100, 120, and 140 kVp and dual energy spectral CT mode. For spectral CT, 11 monochromatic image sets from 40 to 140 keV with interval of 10 keV were reconstructed. The CT number shift caused by beam hardening was evaluated by measuring the CT number difference (ΔCT) with and without beam hardening, with the following formulas: ΔCT(WBH) = |CT(WBH) − CT(NBH)| and ΔCT(SBH) = |CT(SBH) − CT(NBH)|. Data were compared with 1-way analysis of variance. RESULTS: Under both WBH and SBH conditions, the CT number shifts in all monochromatic image sets were less than those for polychromatic images (all P < 0.001). Under WBH condition, the maximum CT number shift was less than 6 Hounsfield units for monochromatic spectral CT images of all energy levels; under SBH condition, only monochromatic images at 70 keV and 80 keV had CT number shift less than 6 HU. CONCLUSION: Dual energy spectral CT imaging provided more accurate CT number measurement than conventional CT under various beam hardening conditions. The optimal keV level for monochromatic spectral CT images with the most accurate CT number measurement depends on the severities of beam hardening condition.
format Online
Article
Text
id pubmed-4654266
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-46542662015-11-30 Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment He, Tian Qian, Xiaojun Zhai, Renyou Yang, Zongtao J Comput Assist Tomogr Imaging Physics To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging. MATERIALS AND METHODS: A phantom with 8 cells in periphery region and 1 cell in central region were used. The 8 conditioning tubes in the periphery region were filled with 1 of the 3 iodine solutions to simulate different beam hardening conditions: 0 for no beam hardening (NBH), 20 mg/mL for weak beam hardening (WBH) and 50 mg/mL for severe beam hardening (SBH) condition. Test tube filled with 0, 0.1, 0.5, 1, 2, 5, 10, 20, and 50 mg/mL iodine solution was placed in the central cell alternately. The phantom was scanned with conventional CT mode with 80, 100, 120, and 140 kVp and dual energy spectral CT mode. For spectral CT, 11 monochromatic image sets from 40 to 140 keV with interval of 10 keV were reconstructed. The CT number shift caused by beam hardening was evaluated by measuring the CT number difference (ΔCT) with and without beam hardening, with the following formulas: ΔCT(WBH) = |CT(WBH) − CT(NBH)| and ΔCT(SBH) = |CT(SBH) − CT(NBH)|. Data were compared with 1-way analysis of variance. RESULTS: Under both WBH and SBH conditions, the CT number shifts in all monochromatic image sets were less than those for polychromatic images (all P < 0.001). Under WBH condition, the maximum CT number shift was less than 6 Hounsfield units for monochromatic spectral CT images of all energy levels; under SBH condition, only monochromatic images at 70 keV and 80 keV had CT number shift less than 6 HU. CONCLUSION: Dual energy spectral CT imaging provided more accurate CT number measurement than conventional CT under various beam hardening conditions. The optimal keV level for monochromatic spectral CT images with the most accurate CT number measurement depends on the severities of beam hardening condition. Lippincott Williams & Wilkins 2015-11 2015-11-26 /pmc/articles/PMC4654266/ /pubmed/26196347 http://dx.doi.org/10.1097/RCT.0000000000000287 Text en Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share thework provided it is properly cited. The work cannot be changed in any way or used commercially.
spellingShingle Imaging Physics
He, Tian
Qian, Xiaojun
Zhai, Renyou
Yang, Zongtao
Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
title Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
title_full Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
title_fullStr Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
title_full_unstemmed Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
title_short Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
title_sort computed tomography number measurement consistency under different beam hardening conditions: comparison between dual-energy spectral computed tomography and conventional computed tomography imaging in phantom experiment
topic Imaging Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4654266/
https://www.ncbi.nlm.nih.gov/pubmed/26196347
http://dx.doi.org/10.1097/RCT.0000000000000287
work_keys_str_mv AT hetian computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment
AT qianxiaojun computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment
AT zhairenyou computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment
AT yangzongtao computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment