Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment
To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging. MATERIALS AND METHODS: A phantom with 8 cells in periphery region and 1 cell in central region were used. Th...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4654266/ https://www.ncbi.nlm.nih.gov/pubmed/26196347 http://dx.doi.org/10.1097/RCT.0000000000000287 |
_version_ | 1782402032543465472 |
---|---|
author | He, Tian Qian, Xiaojun Zhai, Renyou Yang, Zongtao |
author_facet | He, Tian Qian, Xiaojun Zhai, Renyou Yang, Zongtao |
author_sort | He, Tian |
collection | PubMed |
description | To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging. MATERIALS AND METHODS: A phantom with 8 cells in periphery region and 1 cell in central region were used. The 8 conditioning tubes in the periphery region were filled with 1 of the 3 iodine solutions to simulate different beam hardening conditions: 0 for no beam hardening (NBH), 20 mg/mL for weak beam hardening (WBH) and 50 mg/mL for severe beam hardening (SBH) condition. Test tube filled with 0, 0.1, 0.5, 1, 2, 5, 10, 20, and 50 mg/mL iodine solution was placed in the central cell alternately. The phantom was scanned with conventional CT mode with 80, 100, 120, and 140 kVp and dual energy spectral CT mode. For spectral CT, 11 monochromatic image sets from 40 to 140 keV with interval of 10 keV were reconstructed. The CT number shift caused by beam hardening was evaluated by measuring the CT number difference (ΔCT) with and without beam hardening, with the following formulas: ΔCT(WBH) = |CT(WBH) − CT(NBH)| and ΔCT(SBH) = |CT(SBH) − CT(NBH)|. Data were compared with 1-way analysis of variance. RESULTS: Under both WBH and SBH conditions, the CT number shifts in all monochromatic image sets were less than those for polychromatic images (all P < 0.001). Under WBH condition, the maximum CT number shift was less than 6 Hounsfield units for monochromatic spectral CT images of all energy levels; under SBH condition, only monochromatic images at 70 keV and 80 keV had CT number shift less than 6 HU. CONCLUSION: Dual energy spectral CT imaging provided more accurate CT number measurement than conventional CT under various beam hardening conditions. The optimal keV level for monochromatic spectral CT images with the most accurate CT number measurement depends on the severities of beam hardening condition. |
format | Online Article Text |
id | pubmed-4654266 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Lippincott Williams & Wilkins |
record_format | MEDLINE/PubMed |
spelling | pubmed-46542662015-11-30 Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment He, Tian Qian, Xiaojun Zhai, Renyou Yang, Zongtao J Comput Assist Tomogr Imaging Physics To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging. MATERIALS AND METHODS: A phantom with 8 cells in periphery region and 1 cell in central region were used. The 8 conditioning tubes in the periphery region were filled with 1 of the 3 iodine solutions to simulate different beam hardening conditions: 0 for no beam hardening (NBH), 20 mg/mL for weak beam hardening (WBH) and 50 mg/mL for severe beam hardening (SBH) condition. Test tube filled with 0, 0.1, 0.5, 1, 2, 5, 10, 20, and 50 mg/mL iodine solution was placed in the central cell alternately. The phantom was scanned with conventional CT mode with 80, 100, 120, and 140 kVp and dual energy spectral CT mode. For spectral CT, 11 monochromatic image sets from 40 to 140 keV with interval of 10 keV were reconstructed. The CT number shift caused by beam hardening was evaluated by measuring the CT number difference (ΔCT) with and without beam hardening, with the following formulas: ΔCT(WBH) = |CT(WBH) − CT(NBH)| and ΔCT(SBH) = |CT(SBH) − CT(NBH)|. Data were compared with 1-way analysis of variance. RESULTS: Under both WBH and SBH conditions, the CT number shifts in all monochromatic image sets were less than those for polychromatic images (all P < 0.001). Under WBH condition, the maximum CT number shift was less than 6 Hounsfield units for monochromatic spectral CT images of all energy levels; under SBH condition, only monochromatic images at 70 keV and 80 keV had CT number shift less than 6 HU. CONCLUSION: Dual energy spectral CT imaging provided more accurate CT number measurement than conventional CT under various beam hardening conditions. The optimal keV level for monochromatic spectral CT images with the most accurate CT number measurement depends on the severities of beam hardening condition. Lippincott Williams & Wilkins 2015-11 2015-11-26 /pmc/articles/PMC4654266/ /pubmed/26196347 http://dx.doi.org/10.1097/RCT.0000000000000287 Text en Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share thework provided it is properly cited. The work cannot be changed in any way or used commercially. |
spellingShingle | Imaging Physics He, Tian Qian, Xiaojun Zhai, Renyou Yang, Zongtao Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment |
title | Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment |
title_full | Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment |
title_fullStr | Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment |
title_full_unstemmed | Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment |
title_short | Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment |
title_sort | computed tomography number measurement consistency under different beam hardening conditions: comparison between dual-energy spectral computed tomography and conventional computed tomography imaging in phantom experiment |
topic | Imaging Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4654266/ https://www.ncbi.nlm.nih.gov/pubmed/26196347 http://dx.doi.org/10.1097/RCT.0000000000000287 |
work_keys_str_mv | AT hetian computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment AT qianxiaojun computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment AT zhairenyou computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment AT yangzongtao computedtomographynumbermeasurementconsistencyunderdifferentbeamhardeningconditionscomparisonbetweendualenergyspectralcomputedtomographyandconventionalcomputedtomographyimaginginphantomexperiment |