Comparison Between NIST and AF Laser Energy Standards Using High Power Lasers
We report the results of a high-energy laser calorimeter comparison conducted by the National Institute of Standards and Technology (NIST), Boulder, Colorado and the U.S. Air Force Primary Standards laboratory (AFPSL), Heath, Ohio. A laser power meter, used as a transfer standard, was calibrated at...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2007
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4656016/ https://www.ncbi.nlm.nih.gov/pubmed/27110471 http://dx.doi.org/10.6028/jres.112.021 |
Sumario: | We report the results of a high-energy laser calorimeter comparison conducted by the National Institute of Standards and Technology (NIST), Boulder, Colorado and the U.S. Air Force Primary Standards laboratory (AFPSL), Heath, Ohio. A laser power meter, used as a transfer standard, was calibrated at each laboratory, sequentially, and the measurement results were compared. These measurements were performed at a nominal power of 800 W and a wavelength of 10.6 μm using CO(2) lasers. Excellent measurement agreement (1.02 %) was demonstrated, which was well within each of the expanded uncertainties from the two laboratories involved in this comparison. |
---|