Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer

The measurement of free immunoglobulin light chains is typically performed on serum; however, the use of alternative specimen types has potential benefits. Using the Freelite™ kappa and lambda free light chains assay on a Roche Diagnostics cobas 8000 c502 analyzer, we compared three specimen types (...

Descripción completa

Detalles Bibliográficos
Autores principales: Nelson, Louis S., Steussy, Bryan, Morris, Cory S., Krasowski, Matthew D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4671985/
https://www.ncbi.nlm.nih.gov/pubmed/26682113
http://dx.doi.org/10.1186/s40064-015-1546-x
_version_ 1782404478733910016
author Nelson, Louis S.
Steussy, Bryan
Morris, Cory S.
Krasowski, Matthew D.
author_facet Nelson, Louis S.
Steussy, Bryan
Morris, Cory S.
Krasowski, Matthew D.
author_sort Nelson, Louis S.
collection PubMed
description The measurement of free immunoglobulin light chains is typically performed on serum; however, the use of alternative specimen types has potential benefits. Using the Freelite™ kappa and lambda free light chains assay on a Roche Diagnostics cobas 8000 c502 analyzer, we compared three specimen types (serum, EDTA-plasma and lithium heparin plasma separator gel-plasma) on 100 patients. Using Deming regression and eliminating outliers (limiting data to light chain concentrations below 400 mg/L), the three specimen types showed comparable results for kappa light chain concentration, lambda light chain concentration, and kappa/lambda ratio with slopes close to 1.0 and y-intercepts close to zero. EDTA-plasma showed slightly more positive bias relative to serum than lithium heparin. Analysis using EDTA-plasma and lithium heparin plasma showed comparable linearity, precision, and temperature stability. A single sample showing hook effect (not in the comparison set) gave comparable results using either plasma specimen type. For the Freelite™ kappa and lambda free light chains assay, both EDTA-plasma or lithium heparin-plasma can serve as acceptable substitutes for serum, at least for the Roche cobas 8000 analyzer.
format Online
Article
Text
id pubmed-4671985
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-46719852015-12-17 Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer Nelson, Louis S. Steussy, Bryan Morris, Cory S. Krasowski, Matthew D. Springerplus Research The measurement of free immunoglobulin light chains is typically performed on serum; however, the use of alternative specimen types has potential benefits. Using the Freelite™ kappa and lambda free light chains assay on a Roche Diagnostics cobas 8000 c502 analyzer, we compared three specimen types (serum, EDTA-plasma and lithium heparin plasma separator gel-plasma) on 100 patients. Using Deming regression and eliminating outliers (limiting data to light chain concentrations below 400 mg/L), the three specimen types showed comparable results for kappa light chain concentration, lambda light chain concentration, and kappa/lambda ratio with slopes close to 1.0 and y-intercepts close to zero. EDTA-plasma showed slightly more positive bias relative to serum than lithium heparin. Analysis using EDTA-plasma and lithium heparin plasma showed comparable linearity, precision, and temperature stability. A single sample showing hook effect (not in the comparison set) gave comparable results using either plasma specimen type. For the Freelite™ kappa and lambda free light chains assay, both EDTA-plasma or lithium heparin-plasma can serve as acceptable substitutes for serum, at least for the Roche cobas 8000 analyzer. Springer International Publishing 2015-12-08 /pmc/articles/PMC4671985/ /pubmed/26682113 http://dx.doi.org/10.1186/s40064-015-1546-x Text en © Nelson et al. 2015 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Research
Nelson, Louis S.
Steussy, Bryan
Morris, Cory S.
Krasowski, Matthew D.
Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer
title Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer
title_full Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer
title_fullStr Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer
title_full_unstemmed Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer
title_short Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer
title_sort effect of specimen type on free immunoglobulin light chains analysis on the roche diagnostics cobas 8000 analyzer
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4671985/
https://www.ncbi.nlm.nih.gov/pubmed/26682113
http://dx.doi.org/10.1186/s40064-015-1546-x
work_keys_str_mv AT nelsonlouiss effectofspecimentypeonfreeimmunoglobulinlightchainsanalysisontherochediagnosticscobas8000analyzer
AT steussybryan effectofspecimentypeonfreeimmunoglobulinlightchainsanalysisontherochediagnosticscobas8000analyzer
AT morriscorys effectofspecimentypeonfreeimmunoglobulinlightchainsanalysisontherochediagnosticscobas8000analyzer
AT krasowskimatthewd effectofspecimentypeonfreeimmunoglobulinlightchainsanalysisontherochediagnosticscobas8000analyzer