Investigation of LiF, Mg and Ti (TLD-100) Reproducibility

LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the...

Descripción completa

Detalles Bibliográficos
Autores principales: Sadeghi, M., Sina, S., Faghihi, R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Shiraz University of Medical Sciences 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4681467/
https://www.ncbi.nlm.nih.gov/pubmed/26688801
_version_ 1782405753997361152
author Sadeghi, M.
Sina, S.
Faghihi, R.
author_facet Sadeghi, M.
Sina, S.
Faghihi, R.
author_sort Sadeghi, M.
collection PubMed
description LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses of 81 mGy (two times), 162mGy and 40.5mGy. Results of this study indicate that the average reproducibility of ECC calculation for 40 TLDs is 1.5%, while these values for all chips do not exceed 5%.
format Online
Article
Text
id pubmed-4681467
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Shiraz University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-46814672015-12-18 Investigation of LiF, Mg and Ti (TLD-100) Reproducibility Sadeghi, M. Sina, S. Faghihi, R. J Biomed Phys Eng Technical Note LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses of 81 mGy (two times), 162mGy and 40.5mGy. Results of this study indicate that the average reproducibility of ECC calculation for 40 TLDs is 1.5%, while these values for all chips do not exceed 5%. Shiraz University of Medical Sciences 2015-12-01 /pmc/articles/PMC4681467/ /pubmed/26688801 Text en © 2015: Journal of Biomedical Physics and Engineering This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 3.0 Unported License (http://creativecommons.org/licenses/by/3.0/), which permits unrestricted use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Note
Sadeghi, M.
Sina, S.
Faghihi, R.
Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
title Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
title_full Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
title_fullStr Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
title_full_unstemmed Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
title_short Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
title_sort investigation of lif, mg and ti (tld-100) reproducibility
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4681467/
https://www.ncbi.nlm.nih.gov/pubmed/26688801
work_keys_str_mv AT sadeghim investigationoflifmgandtitld100reproducibility
AT sinas investigationoflifmgandtitld100reproducibility
AT faghihir investigationoflifmgandtitld100reproducibility