Cargando…
Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Shiraz University of Medical Sciences
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4681467/ https://www.ncbi.nlm.nih.gov/pubmed/26688801 |
_version_ | 1782405753997361152 |
---|---|
author | Sadeghi, M. Sina, S. Faghihi, R. |
author_facet | Sadeghi, M. Sina, S. Faghihi, R. |
author_sort | Sadeghi, M. |
collection | PubMed |
description | LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses of 81 mGy (two times), 162mGy and 40.5mGy. Results of this study indicate that the average reproducibility of ECC calculation for 40 TLDs is 1.5%, while these values for all chips do not exceed 5%. |
format | Online Article Text |
id | pubmed-4681467 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Shiraz University of Medical Sciences |
record_format | MEDLINE/PubMed |
spelling | pubmed-46814672015-12-18 Investigation of LiF, Mg and Ti (TLD-100) Reproducibility Sadeghi, M. Sina, S. Faghihi, R. J Biomed Phys Eng Technical Note LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses of 81 mGy (two times), 162mGy and 40.5mGy. Results of this study indicate that the average reproducibility of ECC calculation for 40 TLDs is 1.5%, while these values for all chips do not exceed 5%. Shiraz University of Medical Sciences 2015-12-01 /pmc/articles/PMC4681467/ /pubmed/26688801 Text en © 2015: Journal of Biomedical Physics and Engineering This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 3.0 Unported License (http://creativecommons.org/licenses/by/3.0/), which permits unrestricted use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Note Sadeghi, M. Sina, S. Faghihi, R. Investigation of LiF, Mg and Ti (TLD-100) Reproducibility |
title | Investigation of LiF, Mg and Ti (TLD-100) Reproducibility |
title_full | Investigation of LiF, Mg and Ti (TLD-100) Reproducibility |
title_fullStr | Investigation of LiF, Mg and Ti (TLD-100) Reproducibility |
title_full_unstemmed | Investigation of LiF, Mg and Ti (TLD-100) Reproducibility |
title_short | Investigation of LiF, Mg and Ti (TLD-100) Reproducibility |
title_sort | investigation of lif, mg and ti (tld-100) reproducibility |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4681467/ https://www.ncbi.nlm.nih.gov/pubmed/26688801 |
work_keys_str_mv | AT sadeghim investigationoflifmgandtitld100reproducibility AT sinas investigationoflifmgandtitld100reproducibility AT faghihir investigationoflifmgandtitld100reproducibility |