Cargando…

Comparison of sequentially measured Aloka echo-tracking one-point pulse wave velocity with SphygmoCor carotid–femoral pulse wave velocity

OBJECTIVES: Recently, echo-tracking-derived measures of arterial stiffness have been introduced in clinical practice for the assessment of one-point pulse wave velocity. The purpose of this study was to find a relation between carotid–femoral pulse wave velocity and one-point carotid pulse wave velo...

Descripción completa

Detalles Bibliográficos
Autores principales: Vriz, Olga, Driussi, Caterina, La Carrubba, Salvatore, Di Bello, Vitantonio, Zito, Concetta, Carerj, Scipione, Antonini-Canterin, Francesco
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4687782/
https://www.ncbi.nlm.nih.gov/pubmed/26770685
http://dx.doi.org/10.1177/2050312113507563
Descripción
Sumario:OBJECTIVES: Recently, echo-tracking-derived measures of arterial stiffness have been introduced in clinical practice for the assessment of one-point pulse wave velocity. The purpose of this study was to find a relation between carotid–femoral pulse wave velocity and one-point carotid pulse wave velocity, and to find a value of one-point carotid pulse wave velocity that predicts carotid–femoral pulse wave velocity higher than 12 m/s. METHODS: A total of 160 consecutive subjects (112 male/48 female, mean age = 51.5 ± 14.1 years; 96 healthy, 44 hypertensives, 13 with aortic valve disease, and 7 with left ventricular dysfunction) were studied. Carotid–femoral pulse wave velocity was measured with the SphygmoCor system and one-point carotid pulse wave velocity with high-definition echo-tracking system (ProSound Alpha10; Aloka, Tokyo, Japan). RESULTS: Both carotid–femoral pulse wave velocity and one-point carotid pulse wave velocity correlated significantly with each other (r = 0.539, p < 0.001) and with age (one-point carotid pulse wave velocity r = 0.618, carotid–femoral pulse wave velocity r = 0.617, p < 0.0001 for both). Median value of carotid–femoral pulse wave velocity (7.2 m/s, 95% confidence interval = 6.2–8.9) was systematically higher than that of one-point carotid pulse wave velocity (5.8 m/s, 95% confidence interval = 5–6.6). The area under the receiver operating characteristic curve was 0.85, identifying the cutoff for one-point pulse wave velocity of 6.65 m/s as the best predictor of carotid–femoral pulse wave velocity more than 12 m/s (sensitivity = 0.818, specificity = 0.819). CONCLUSIONS: One-point carotid pulse wave velocity correlates with carotid–femoral pulse wave velocity, and the cutoff of 6.65 m/s was the best predictor of carotid–femoral pulse wave velocity over 12 m/s.