Positive contrast spiral imaging for visualization of commercial nitinol guidewires with reduced heating

BACKGROUND: CMR-guidance has the potential to improve tissue visualization during cardiovascular catheterization procedures and to reduce ionizing radiation exposure, but a lack of commercially available CMR guidewires limits widespread adoption. Standard metallic guidewires are considered to be uns...

Descripción completa

Detalles Bibliográficos
Autores principales: Campbell-Washburn, Adrienne E., Rogers, Toby, Basar, Burcu, Sonmez, Merdim, Kocaturk, Ozgur, Lederman, Robert J., Hansen, Michael S., Faranesh, Anthony Z.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4688983/
https://www.ncbi.nlm.nih.gov/pubmed/26695490
http://dx.doi.org/10.1186/s12968-015-0219-9
Descripción
Sumario:BACKGROUND: CMR-guidance has the potential to improve tissue visualization during cardiovascular catheterization procedures and to reduce ionizing radiation exposure, but a lack of commercially available CMR guidewires limits widespread adoption. Standard metallic guidewires are considered to be unsafe in CMR due to risks of RF-induced heating. Here, we propose the use of RF-efficient gradient echo (GRE) spiral imaging for reduced guidewire heating (low flip angle, long readout), in combination with positive contrast for guidewire visualization. METHODS: A GRE spiral sequence with 8 interleaves was used for imaging. Positive contrast was achieved using through-slice dephasing such that the guidewire appeared bright and the background signal suppressed. Positive contrast images were interleaved with anatomical images, and real-time image processing was used to produce a color overlay of the guidewire on the anatomy. Temperature was measured with a fiber-optic probe attached to the guidewire in an acrylic gel phantom and in vivo. RESULTS: Left heart catheterization was performed on swine using the real-time color overlay for procedural guidance with a frame rate of 6.25 frames/second. Using our standard Cartesian real-time imaging (flip angle 60°), temperature increases up to 50 °C (phantom) and 4 °C (in vivo) were observed. In comparison, spiral GRE images (8 interleaves, flip angle 10°) generated negligible heating measuring 0.37 °C (phantom) and 0.06 °C (in vivo). CONCLUSIONS: The ability to use commercial metallic guidewires safely during CMR-guided catheterization could potentially expedite clinical translation of these methods. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s12968-015-0219-9) contains supplementary material, which is available to authorized users.