Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor

The Nucleus(®) 5 System Sound Processor (CP810, Cochlear™, Macquarie University, NSW, Australia) contains two omnidirectional microphones. They can be configured as a fixed directional microphone combination (called Zoom) or as an adaptive beamformer (called Beam), which adjusts the directivity cont...

Descripción completa

Detalles Bibliográficos
Autores principales: Dillier, Norbert, Lai, Wai Kong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: PAGEPress Publications, Pavia, Italy 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4698600/
https://www.ncbi.nlm.nih.gov/pubmed/26779327
http://dx.doi.org/10.4081/audiores.2015.132
_version_ 1782408055235805184
author Dillier, Norbert
Lai, Wai Kong
author_facet Dillier, Norbert
Lai, Wai Kong
author_sort Dillier, Norbert
collection PubMed
description The Nucleus(®) 5 System Sound Processor (CP810, Cochlear™, Macquarie University, NSW, Australia) contains two omnidirectional microphones. They can be configured as a fixed directional microphone combination (called Zoom) or as an adaptive beamformer (called Beam), which adjusts the directivity continuously to maximally reduce the interfering noise. Initial evaluation studies with the CP810 had compared performance and usability of the new processor in comparison with the Freedom™ Sound Processor (Cochlear™) for speech in quiet and noise for a subset of the processing options. This study compares the two processing options suggested to be used in noisy environments, Zoom and Beam, for various sound field conditions using a standardized speech in noise matrix test (Oldenburg sentences test). Nine German-speaking subjects who previously had been using the Freedom speech processor and subsequently were upgraded to the CP810 device participated in this series of additional evaluation tests. The speech reception threshold (SRT for 50% speech intelligibility in noise) was determined using sentences presented via loudspeaker at 65 dB SPL in front of the listener and noise presented either via the same loudspeaker (S(0)N(0)) or at 90 degrees at either the ear with the sound processor (S(0)N(CI+)) or the opposite unaided ear (S(0)N(CI-)). The fourth noise condition consisted of three uncorrelated noise sources placed at 90, 180 and 270 degrees. The noise level was adjusted through an adaptive procedure to yield a signal to noise ratio where 50% of the words in the sentences were correctly understood. In spatially separated speech and noise conditions both Zoom and Beam could improve the SRT significantly. For single noise sources, either ipsilateral or contralateral to the cochlear implant sound processor, average improvements with Beam of 12.9 and 7.9 dB in SRT were found. The average SRT of –8 dB for Beam in the diffuse noise condition (uncorrelated noise from both sides and back) is truly remarkable and comparable to the performance of normal hearing listeners in the same test environment. The static directivity (Zoom) option in the diffuse noise condition still provides a significant benefit of 5.9 dB in comparison with the standard omnidirectional microphone setting. These results indicate that CI recipients may improve their speech recognition in noisy environments significantly using these directional microphone-processing options.
format Online
Article
Text
id pubmed-4698600
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher PAGEPress Publications, Pavia, Italy
record_format MEDLINE/PubMed
spelling pubmed-46986002016-01-15 Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor Dillier, Norbert Lai, Wai Kong Audiol Res Article The Nucleus(®) 5 System Sound Processor (CP810, Cochlear™, Macquarie University, NSW, Australia) contains two omnidirectional microphones. They can be configured as a fixed directional microphone combination (called Zoom) or as an adaptive beamformer (called Beam), which adjusts the directivity continuously to maximally reduce the interfering noise. Initial evaluation studies with the CP810 had compared performance and usability of the new processor in comparison with the Freedom™ Sound Processor (Cochlear™) for speech in quiet and noise for a subset of the processing options. This study compares the two processing options suggested to be used in noisy environments, Zoom and Beam, for various sound field conditions using a standardized speech in noise matrix test (Oldenburg sentences test). Nine German-speaking subjects who previously had been using the Freedom speech processor and subsequently were upgraded to the CP810 device participated in this series of additional evaluation tests. The speech reception threshold (SRT for 50% speech intelligibility in noise) was determined using sentences presented via loudspeaker at 65 dB SPL in front of the listener and noise presented either via the same loudspeaker (S(0)N(0)) or at 90 degrees at either the ear with the sound processor (S(0)N(CI+)) or the opposite unaided ear (S(0)N(CI-)). The fourth noise condition consisted of three uncorrelated noise sources placed at 90, 180 and 270 degrees. The noise level was adjusted through an adaptive procedure to yield a signal to noise ratio where 50% of the words in the sentences were correctly understood. In spatially separated speech and noise conditions both Zoom and Beam could improve the SRT significantly. For single noise sources, either ipsilateral or contralateral to the cochlear implant sound processor, average improvements with Beam of 12.9 and 7.9 dB in SRT were found. The average SRT of –8 dB for Beam in the diffuse noise condition (uncorrelated noise from both sides and back) is truly remarkable and comparable to the performance of normal hearing listeners in the same test environment. The static directivity (Zoom) option in the diffuse noise condition still provides a significant benefit of 5.9 dB in comparison with the standard omnidirectional microphone setting. These results indicate that CI recipients may improve their speech recognition in noisy environments significantly using these directional microphone-processing options. PAGEPress Publications, Pavia, Italy 2015-10-07 /pmc/articles/PMC4698600/ /pubmed/26779327 http://dx.doi.org/10.4081/audiores.2015.132 Text en ©Copyright N. Dillier and W.K. Lai http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Article
Dillier, Norbert
Lai, Wai Kong
Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor
title Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor
title_full Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor
title_fullStr Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor
title_full_unstemmed Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor
title_short Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 CP810 Sound Processor
title_sort speech intelligibility in various noise conditions with the nucleus® 5 cp810 sound processor
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4698600/
https://www.ncbi.nlm.nih.gov/pubmed/26779327
http://dx.doi.org/10.4081/audiores.2015.132
work_keys_str_mv AT dilliernorbert speechintelligibilityinvariousnoiseconditionswiththenucleus5cp810soundprocessor
AT laiwaikong speechintelligibilityinvariousnoiseconditionswiththenucleus5cp810soundprocessor