Cargando…

Feasibility of Ultra-Thin Fiber-Optic Dosimeters for Radiotherapy Dosimetry

In this study, prototype ultra-thin fiber-optic dosimeters were fabricated using organic scintillators, wavelength shifting fibers, and plastic optical fibers. The sensor probes of the ultra-thin fiber-optic dosimeters consisted of very thin organic scintillators with thicknesses of 100, 150 and 200...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Bongsoo, Kwon, Guwon, Shin, Sang Hun, Kim, Jaeseok, Yoo, Wook Jae, Ji, Young Hoon, Jang, Kyoung Won
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4701318/
https://www.ncbi.nlm.nih.gov/pubmed/26593917
http://dx.doi.org/10.3390/s151129003
Descripción
Sumario:In this study, prototype ultra-thin fiber-optic dosimeters were fabricated using organic scintillators, wavelength shifting fibers, and plastic optical fibers. The sensor probes of the ultra-thin fiber-optic dosimeters consisted of very thin organic scintillators with thicknesses of 100, 150 and 200 μm. These types of sensors cannot only be used to measure skin or surface doses but also provide depth dose measurements with high spatial resolution. With the ultra-thin fiber-optic dosimeters, surface doses for gamma rays generated from a Co-60 therapy machine were measured. Additionally, percentage depth doses in the build-up regions were obtained by using the ultra-thin fiber-optic dosimeters, and the results were compared with those of external beam therapy films and a conventional fiber-optic dosimeter.