Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults
BACKGROUND: The use of evaluation tools such as the manovacuometer and respirometer is frequent and disinfection is usually limited to the external surfaces, which is insufficient and raises concerns because of the potential spread of infectious diseases. Hydrophobic heat and moisture exchangers (HM...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4705747/ https://www.ncbi.nlm.nih.gov/pubmed/26753093 http://dx.doi.org/10.1186/s40248-015-0037-9 |
_version_ | 1782409073984012288 |
---|---|
author | Lucato, Jeanette Janaina Jaber Nogueira da Cunha, Thiago Marraccini Rocha, Sara Solange Oliveira Costa Palmieri de Carvalho, Fernanda Maria Botega, Daniele Cristina Torquato, Jamili Anbar Gimenes, Ana Cristina Righetti, Renato Fraga |
author_facet | Lucato, Jeanette Janaina Jaber Nogueira da Cunha, Thiago Marraccini Rocha, Sara Solange Oliveira Costa Palmieri de Carvalho, Fernanda Maria Botega, Daniele Cristina Torquato, Jamili Anbar Gimenes, Ana Cristina Righetti, Renato Fraga |
author_sort | Lucato, Jeanette Janaina Jaber |
collection | PubMed |
description | BACKGROUND: The use of evaluation tools such as the manovacuometer and respirometer is frequent and disinfection is usually limited to the external surfaces, which is insufficient and raises concerns because of the potential spread of infectious diseases. Hydrophobic heat and moisture exchangers (HME) are used in mechanical ventilation and have microbiological filters, which can possibly reduce contamination, increasing the safety of related procedures. It is unknown, however, if the addition of an exchanger affects the measurements obtained. Aim of this study was to verify if the use of an HME interferes in maximal inspiratory and expiratory pressures assessed using the manovacuometer and vital capacity evaluated using the respirometer in healthy adults. METHODS: A controlled transversal trial was carried out. Twenty healthy young adults were included in the study. Vital capacity by respirometer and, maximal inspiratory pressure (MIP) and maximal expiratory pressure (MEP) were assessed with and without the use of HME. RESULTS: No significant difference was found between the values pre and post HME use in vital capacity measurements: (3878.8 ± 202.2 mL vs. 3925.5 ± 206.0 mL, p = 0.116) and the respiratory muscle strength measurements: MIP (−99.0 ± 8.9 vs −95.5 ± 9.0 cm H(2)O, p = 0.149) and MEP (92.5 ± 7.5 vs 92.5 ± 7.7 cm H(2)O, p = 1.0) respectively. CONCLUSION: We conclude that the use of HME does not modify the lung volumes or respiratory muscle strength, and can be used in order to reduce the occurrence of pulmonary infection. |
format | Online Article Text |
id | pubmed-4705747 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-47057472016-01-09 Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults Lucato, Jeanette Janaina Jaber Nogueira da Cunha, Thiago Marraccini Rocha, Sara Solange Oliveira Costa Palmieri de Carvalho, Fernanda Maria Botega, Daniele Cristina Torquato, Jamili Anbar Gimenes, Ana Cristina Righetti, Renato Fraga Multidiscip Respir Med Original Research Article BACKGROUND: The use of evaluation tools such as the manovacuometer and respirometer is frequent and disinfection is usually limited to the external surfaces, which is insufficient and raises concerns because of the potential spread of infectious diseases. Hydrophobic heat and moisture exchangers (HME) are used in mechanical ventilation and have microbiological filters, which can possibly reduce contamination, increasing the safety of related procedures. It is unknown, however, if the addition of an exchanger affects the measurements obtained. Aim of this study was to verify if the use of an HME interferes in maximal inspiratory and expiratory pressures assessed using the manovacuometer and vital capacity evaluated using the respirometer in healthy adults. METHODS: A controlled transversal trial was carried out. Twenty healthy young adults were included in the study. Vital capacity by respirometer and, maximal inspiratory pressure (MIP) and maximal expiratory pressure (MEP) were assessed with and without the use of HME. RESULTS: No significant difference was found between the values pre and post HME use in vital capacity measurements: (3878.8 ± 202.2 mL vs. 3925.5 ± 206.0 mL, p = 0.116) and the respiratory muscle strength measurements: MIP (−99.0 ± 8.9 vs −95.5 ± 9.0 cm H(2)O, p = 0.149) and MEP (92.5 ± 7.5 vs 92.5 ± 7.7 cm H(2)O, p = 1.0) respectively. CONCLUSION: We conclude that the use of HME does not modify the lung volumes or respiratory muscle strength, and can be used in order to reduce the occurrence of pulmonary infection. BioMed Central 2015-12-19 /pmc/articles/PMC4705747/ /pubmed/26753093 http://dx.doi.org/10.1186/s40248-015-0037-9 Text en © Lucato et al. 2015 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Original Research Article Lucato, Jeanette Janaina Jaber Nogueira da Cunha, Thiago Marraccini Rocha, Sara Solange Oliveira Costa Palmieri de Carvalho, Fernanda Maria Botega, Daniele Cristina Torquato, Jamili Anbar Gimenes, Ana Cristina Righetti, Renato Fraga Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults |
title | Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults |
title_full | Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults |
title_fullStr | Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults |
title_full_unstemmed | Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults |
title_short | Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults |
title_sort | influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults |
topic | Original Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4705747/ https://www.ncbi.nlm.nih.gov/pubmed/26753093 http://dx.doi.org/10.1186/s40248-015-0037-9 |
work_keys_str_mv | AT lucatojeanettejanainajaber influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults AT nogueiradacunhathiagomarraccini influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults AT rochasarasolangeoliveiracosta influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults AT palmieridecarvalhofernandamaria influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults AT botegadanielecristina influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults AT torquatojamilianbar influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults AT gimenesanacristina influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults AT righettirenatofraga influenceofheatandmoistureexchangeruseonmeasurementsperformedwithmanovacuometerandrespirometerinhealthyadults |