Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry
Determination of trace amounts of bisphenol A (BPA) may cause problems mainly related to the presence of BPA in solvents (even in LC-MS grade), laboratory vessels, and plastic equipment used for sample preparation. Variable and sometimes significant amounts of BPA present in the background cause pro...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4709387/ https://www.ncbi.nlm.nih.gov/pubmed/26635019 http://dx.doi.org/10.1007/s00216-015-9215-z |
_version_ | 1782409631737315328 |
---|---|
author | Wilczewska, Kamila Namieśnik, Jacek Wasik, Andrzej |
author_facet | Wilczewska, Kamila Namieśnik, Jacek Wasik, Andrzej |
author_sort | Wilczewska, Kamila |
collection | PubMed |
description | Determination of trace amounts of bisphenol A (BPA) may cause problems mainly related to the presence of BPA in solvents (even in LC-MS grade), laboratory vessels, and plastic equipment used for sample preparation. Variable and sometimes significant amounts of BPA present in the background cause problems in obtaining good repeatability of measurements at the ultra-trace levels. Such observations (i.e., poor repeatability of results) were made during development of the LC-MS/MS method for determination of BPA in human serum samples. The method included gradient separation of the sample’s constituents. The BPA peak was present in the chromatograms not only when procedural blanks were injected but also when void injections were made. One of the possible ways to eliminate background contamination is to change the source of solvents, use a different water purification system, and introduce rigorous equipment cleaning procedures. However, despite the use of these recommended guidelines, the peak of BPA was still present in the system blank. It was observed that the intensity of the BPA peak was proportional to the time of column conditioning. It was concluded that BPA, present in the components of the mobile phase, is being enriched in the front of the separation column during its conditioning (i.e., when mobile phase elution strength was low). This paper describes effects of gradient and isocratic elution conditions on LC-MS/MS system blank. The problem of spurious BPA peak, originating from the mobile phase, was solved by replacing gradient with isocratic elution mode. The use of isocratic elution conditions with the mobile phase of relatively high elution strength (50 % of acetonitrile) allowed elimination of the peak of BPA coming from the mobile phase and significantly improved the precision of determination of BPA at low concentration levels. |
format | Online Article Text |
id | pubmed-4709387 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-47093872016-01-19 Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry Wilczewska, Kamila Namieśnik, Jacek Wasik, Andrzej Anal Bioanal Chem Note Determination of trace amounts of bisphenol A (BPA) may cause problems mainly related to the presence of BPA in solvents (even in LC-MS grade), laboratory vessels, and plastic equipment used for sample preparation. Variable and sometimes significant amounts of BPA present in the background cause problems in obtaining good repeatability of measurements at the ultra-trace levels. Such observations (i.e., poor repeatability of results) were made during development of the LC-MS/MS method for determination of BPA in human serum samples. The method included gradient separation of the sample’s constituents. The BPA peak was present in the chromatograms not only when procedural blanks were injected but also when void injections were made. One of the possible ways to eliminate background contamination is to change the source of solvents, use a different water purification system, and introduce rigorous equipment cleaning procedures. However, despite the use of these recommended guidelines, the peak of BPA was still present in the system blank. It was observed that the intensity of the BPA peak was proportional to the time of column conditioning. It was concluded that BPA, present in the components of the mobile phase, is being enriched in the front of the separation column during its conditioning (i.e., when mobile phase elution strength was low). This paper describes effects of gradient and isocratic elution conditions on LC-MS/MS system blank. The problem of spurious BPA peak, originating from the mobile phase, was solved by replacing gradient with isocratic elution mode. The use of isocratic elution conditions with the mobile phase of relatively high elution strength (50 % of acetonitrile) allowed elimination of the peak of BPA coming from the mobile phase and significantly improved the precision of determination of BPA at low concentration levels. Springer Berlin Heidelberg 2015-12-03 2016 /pmc/articles/PMC4709387/ /pubmed/26635019 http://dx.doi.org/10.1007/s00216-015-9215-z Text en © The Author(s) 2015 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Note Wilczewska, Kamila Namieśnik, Jacek Wasik, Andrzej Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry |
title | Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry |
title_full | Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry |
title_fullStr | Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry |
title_full_unstemmed | Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry |
title_short | Troubleshooting of the determination of bisphenol A at ultra-trace levels by liquid chromatography and tandem mass spectrometry |
title_sort | troubleshooting of the determination of bisphenol a at ultra-trace levels by liquid chromatography and tandem mass spectrometry |
topic | Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4709387/ https://www.ncbi.nlm.nih.gov/pubmed/26635019 http://dx.doi.org/10.1007/s00216-015-9215-z |
work_keys_str_mv | AT wilczewskakamila troubleshootingofthedeterminationofbisphenolaatultratracelevelsbyliquidchromatographyandtandemmassspectrometry AT namiesnikjacek troubleshootingofthedeterminationofbisphenolaatultratracelevelsbyliquidchromatographyandtandemmassspectrometry AT wasikandrzej troubleshootingofthedeterminationofbisphenolaatultratracelevelsbyliquidchromatographyandtandemmassspectrometry |