Cargando…

Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults

BACKGROUD: To determine whether the professional Omron HBP-1300 blood pressure (BP) monitor meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards in Chinese children and adults. METHOD: According to the AAMI protocol, simultaneous auscultatory measuremen...

Descripción completa

Detalles Bibliográficos
Autores principales: Meng, Linghui, Zhao, Di, Pan, Yan, Ding, Wenqing, Wei, Qing, Li, Hua, Gao, Pingjin, Mi, Jie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4711064/
https://www.ncbi.nlm.nih.gov/pubmed/26758197
http://dx.doi.org/10.1186/s12872-015-0177-z
_version_ 1782409911656775680
author Meng, Linghui
Zhao, Di
Pan, Yan
Ding, Wenqing
Wei, Qing
Li, Hua
Gao, Pingjin
Mi, Jie
author_facet Meng, Linghui
Zhao, Di
Pan, Yan
Ding, Wenqing
Wei, Qing
Li, Hua
Gao, Pingjin
Mi, Jie
author_sort Meng, Linghui
collection PubMed
description BACKGROUD: To determine whether the professional Omron HBP-1300 blood pressure (BP) monitor meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards in Chinese children and adults. METHOD: According to the AAMI protocol, simultaneous auscultatory measurements by two observers using a mercury manometer were obtained in participants using the Omron HBP-1300. Triple measurements were obtained after a minimum 5-min rest with a 1-min interval between adjacent measurements. RESULTS: A total of 85 participants submitted to 255 doctor-measured BP and 255 successful professional monitor readings. The initial auscultation systolic BP was <100 mmHg in 25 participants (29.4 %), 100–160 mmHg in 53 participants (62.4 %), and >160 mmHg in seven participants (8.2 %). All of the simultaneous measurements agreed to within ±10 mmHg, while 95 % agreed to within ±4 mmHg for both systolic and diastolic BP, and the consistency between two observers was satisfactory. The difference between the devices was -1.3 ± 3.6 mmHg for systolic BP and 0.7 ± 3.8 mmHg for diastolic BP and by AAMI method 1, which met this guideline. The average difference between two devices by AAMI method 2 was 1.4 ± 3.2 mmHg for systolic BP and 1.0 ± 3.9 mmHg for diastolic BP, which met this guideline. CONCLUSION: The professional BP monitor Omron HBP-1300 is desirable for measuring the BP for Chinese children and adults. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s12872-015-0177-z) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-4711064
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-47110642016-01-14 Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults Meng, Linghui Zhao, Di Pan, Yan Ding, Wenqing Wei, Qing Li, Hua Gao, Pingjin Mi, Jie BMC Cardiovasc Disord Research Article BACKGROUD: To determine whether the professional Omron HBP-1300 blood pressure (BP) monitor meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards in Chinese children and adults. METHOD: According to the AAMI protocol, simultaneous auscultatory measurements by two observers using a mercury manometer were obtained in participants using the Omron HBP-1300. Triple measurements were obtained after a minimum 5-min rest with a 1-min interval between adjacent measurements. RESULTS: A total of 85 participants submitted to 255 doctor-measured BP and 255 successful professional monitor readings. The initial auscultation systolic BP was <100 mmHg in 25 participants (29.4 %), 100–160 mmHg in 53 participants (62.4 %), and >160 mmHg in seven participants (8.2 %). All of the simultaneous measurements agreed to within ±10 mmHg, while 95 % agreed to within ±4 mmHg for both systolic and diastolic BP, and the consistency between two observers was satisfactory. The difference between the devices was -1.3 ± 3.6 mmHg for systolic BP and 0.7 ± 3.8 mmHg for diastolic BP and by AAMI method 1, which met this guideline. The average difference between two devices by AAMI method 2 was 1.4 ± 3.2 mmHg for systolic BP and 1.0 ± 3.9 mmHg for diastolic BP, which met this guideline. CONCLUSION: The professional BP monitor Omron HBP-1300 is desirable for measuring the BP for Chinese children and adults. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s12872-015-0177-z) contains supplementary material, which is available to authorized users. BioMed Central 2016-01-13 /pmc/articles/PMC4711064/ /pubmed/26758197 http://dx.doi.org/10.1186/s12872-015-0177-z Text en © Meng et al. 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Meng, Linghui
Zhao, Di
Pan, Yan
Ding, Wenqing
Wei, Qing
Li, Hua
Gao, Pingjin
Mi, Jie
Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults
title Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults
title_full Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults
title_fullStr Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults
title_full_unstemmed Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults
title_short Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults
title_sort validation of omron hbp-1300 professional blood pressure monitor based on auscultation in children and adults
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4711064/
https://www.ncbi.nlm.nih.gov/pubmed/26758197
http://dx.doi.org/10.1186/s12872-015-0177-z
work_keys_str_mv AT menglinghui validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults
AT zhaodi validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults
AT panyan validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults
AT dingwenqing validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults
AT weiqing validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults
AT lihua validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults
AT gaopingjin validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults
AT mijie validationofomronhbp1300professionalbloodpressuremonitorbasedonauscultationinchildrenandadults