Direct Identification and Antimicrobial Susceptibility Testing of Bacteria From Positive Blood Culture Bottles by Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry and the Vitek 2 System
BACKGROUND: We evaluated the reliability and accuracy of the combined use of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) bacterial identification and Vitek 2 antimicrobial susceptibility testing (AST) for bacteria from positive blood culture bottles. M...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society for Laboratory Medicine
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4713844/ https://www.ncbi.nlm.nih.gov/pubmed/26709258 http://dx.doi.org/10.3343/alm.2016.36.2.117 |
Sumario: | BACKGROUND: We evaluated the reliability and accuracy of the combined use of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) bacterial identification and Vitek 2 antimicrobial susceptibility testing (AST) for bacteria from positive blood culture bottles. METHODS: Direct identification and AST were performed in parallel to the standard methods in monomicrobial positive blood culture bottles. In total, 254 isolates grown on aerobic and/or anaerobic bottles were identified with MALDI-TOF Vitek MS (bioMérieux, France), and 1,978 microorganism/antimicrobial agent combinations were assessed. For isolates from anaerobic bottles, an aliquot of the culture broth was centrifuged, washed, and filtered through a nylon mesh. For isolates from aerobic/pediatric bottles, a lysis step using 9.26% ammonium chloride solution and 2% saponin solution was included. RESULTS: The overall correct identification rate was 81.8% (208/254) and that for gram-positive/gram-negative isolates was 73.9%/92.6%, respectively, and it was 81.8%, 87.6%, and 57.9% for isolates from aerobic, anaerobic, and pediatric bottles, respectively. Identification was not possible in 45 cases, and most of these isolates were streptococci (N=14) and coagulase-negative staphylococci (N=11). Misidentification occurred only in one case. Compared with standard methods, direct AST showed 97.9% (1,936/1,978) agreement with very major error of 0.25%, major error of 0.05%, and minor error of 1.8%. CONCLUSIONS: This simple and cost-effective sample preparation method gives reliable results for the direct identification and AST of bacteria. For the identification of streptococci and coagulase-negative staphylococci, the method should be further improved. |
---|