Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators
PURPOSE: In-vivo measurements to determine doses to organs-at-risk can be an essential part of brachytherapy quality assurance (QA). This study compares calculated doses to the rectum with measured dose values as a means of QA in vaginal vault brachytherapy using cylinder applicators. MATERIAL AND M...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4716134/ https://www.ncbi.nlm.nih.gov/pubmed/26816506 http://dx.doi.org/10.5114/jcb.2015.56767 |
_version_ | 1782410506858921984 |
---|---|
author | Oyekunle, Emmanuel Oyeyemi Obed, Rachel Ibhade Akinlade, Bidemi Idayat Ntekim, Atara |
author_facet | Oyekunle, Emmanuel Oyeyemi Obed, Rachel Ibhade Akinlade, Bidemi Idayat Ntekim, Atara |
author_sort | Oyekunle, Emmanuel Oyeyemi |
collection | PubMed |
description | PURPOSE: In-vivo measurements to determine doses to organs-at-risk can be an essential part of brachytherapy quality assurance (QA). This study compares calculated doses to the rectum with measured dose values as a means of QA in vaginal vault brachytherapy using cylinder applicators. MATERIAL AND METHODS: At the Department of Radiotherapy, University College Hospital (UCH), Ibadan, Nigeria, intracavitary brachytherapy (ICBT) was delivered by a GyneSource high-dose-rate (HDR) unit with (60)Co. Standard 2D treatment plans were created with HDR basic 2.6 software for prescription doses 5-7 Gy at points 5 mm away from the posterior surface of vaginal cylinder applicators (20, 25, and 30 mm diameters). The LiF:Mg, Ti thermoluminescent dosimeter rods (1 x 6 mm) were irradiated to a dose of 7 Gy on Theratron (60)Co machine for calibration purpose prior to clinical use. Measurements in each of 34 insertions involving fourteen patients were performed with 5 TLD-100 rods placed along a re-usable rectal marker positioned in the rectum. The dosimeters were read in Harshaw 3500 TLD reader and compared with doses derived from the treatment planning system (TPS) at 1 cm away from the dose prescription points. RESULTS: The mean calculated and measured doses ranged from 2.1-3.8 Gy and 1.2-5.6 Gy with averages of 3.0 ± 0.5 Gy and 3.1 ± 1.1 Gy, respectively, for treatment lengths 2-8 cm along the cylinder-applicators. The mean values correspond to 48.9% and 50.8% of the prescribed doses, respectively. The deviations of the mean in-vivo doses from the TPS values ranged from –1.9 to 2.1 Gy with a p-value of 0.427. CONCLUSIONS: This study was part of efforts to verify rectal dose obtained from the TPS during vaginal vault brachytherapy. There was no significant difference in the dose to the rectum from the two methods of measurements. |
format | Online Article Text |
id | pubmed-4716134 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Termedia Publishing House |
record_format | MEDLINE/PubMed |
spelling | pubmed-47161342016-01-26 Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators Oyekunle, Emmanuel Oyeyemi Obed, Rachel Ibhade Akinlade, Bidemi Idayat Ntekim, Atara J Contemp Brachytherapy Original Paper PURPOSE: In-vivo measurements to determine doses to organs-at-risk can be an essential part of brachytherapy quality assurance (QA). This study compares calculated doses to the rectum with measured dose values as a means of QA in vaginal vault brachytherapy using cylinder applicators. MATERIAL AND METHODS: At the Department of Radiotherapy, University College Hospital (UCH), Ibadan, Nigeria, intracavitary brachytherapy (ICBT) was delivered by a GyneSource high-dose-rate (HDR) unit with (60)Co. Standard 2D treatment plans were created with HDR basic 2.6 software for prescription doses 5-7 Gy at points 5 mm away from the posterior surface of vaginal cylinder applicators (20, 25, and 30 mm diameters). The LiF:Mg, Ti thermoluminescent dosimeter rods (1 x 6 mm) were irradiated to a dose of 7 Gy on Theratron (60)Co machine for calibration purpose prior to clinical use. Measurements in each of 34 insertions involving fourteen patients were performed with 5 TLD-100 rods placed along a re-usable rectal marker positioned in the rectum. The dosimeters were read in Harshaw 3500 TLD reader and compared with doses derived from the treatment planning system (TPS) at 1 cm away from the dose prescription points. RESULTS: The mean calculated and measured doses ranged from 2.1-3.8 Gy and 1.2-5.6 Gy with averages of 3.0 ± 0.5 Gy and 3.1 ± 1.1 Gy, respectively, for treatment lengths 2-8 cm along the cylinder-applicators. The mean values correspond to 48.9% and 50.8% of the prescribed doses, respectively. The deviations of the mean in-vivo doses from the TPS values ranged from –1.9 to 2.1 Gy with a p-value of 0.427. CONCLUSIONS: This study was part of efforts to verify rectal dose obtained from the TPS during vaginal vault brachytherapy. There was no significant difference in the dose to the rectum from the two methods of measurements. Termedia Publishing House 2015-12-30 2015-12 /pmc/articles/PMC4716134/ /pubmed/26816506 http://dx.doi.org/10.5114/jcb.2015.56767 Text en Copyright © 2015 Termedia Sp. z o. o. http://creativecommons.org/licenses/by-nc-sa/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license. |
spellingShingle | Original Paper Oyekunle, Emmanuel Oyeyemi Obed, Rachel Ibhade Akinlade, Bidemi Idayat Ntekim, Atara Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators |
title | Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators |
title_full | Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators |
title_fullStr | Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators |
title_full_unstemmed | Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators |
title_short | Comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators |
title_sort | comparison of doses to the rectum derived from treatment planning system with in-vivo dose values in vaginal vault brachytherapy using cylinder applicators |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4716134/ https://www.ncbi.nlm.nih.gov/pubmed/26816506 http://dx.doi.org/10.5114/jcb.2015.56767 |
work_keys_str_mv | AT oyekunleemmanueloyeyemi comparisonofdosestotherectumderivedfromtreatmentplanningsystemwithinvivodosevaluesinvaginalvaultbrachytherapyusingcylinderapplicators AT obedrachelibhade comparisonofdosestotherectumderivedfromtreatmentplanningsystemwithinvivodosevaluesinvaginalvaultbrachytherapyusingcylinderapplicators AT akinladebidemiidayat comparisonofdosestotherectumderivedfromtreatmentplanningsystemwithinvivodosevaluesinvaginalvaultbrachytherapyusingcylinderapplicators AT ntekimatara comparisonofdosestotherectumderivedfromtreatmentplanningsystemwithinvivodosevaluesinvaginalvaultbrachytherapyusingcylinderapplicators |