Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert

PURPOSE: To provide a practical solution that can be adopted in clinical routine to fulfill the AAPM-ESTRO recommendations regarding quality assurance of seeds used in prostate permanent brachytherapy. The aim is to design a new insert for the well-ionization chamber SourceCheck(4π) 33005 (PTW, Germ...

Descripción completa

Detalles Bibliográficos
Autores principales: Candela-Juan, Cristian, Ballester, Facundo, Perez-Calatayud, Jose, Vijande, Javier
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4716135/
https://www.ncbi.nlm.nih.gov/pubmed/26816507
http://dx.doi.org/10.5114/jcb.2015.56768
_version_ 1782410507084365824
author Candela-Juan, Cristian
Ballester, Facundo
Perez-Calatayud, Jose
Vijande, Javier
author_facet Candela-Juan, Cristian
Ballester, Facundo
Perez-Calatayud, Jose
Vijande, Javier
author_sort Candela-Juan, Cristian
collection PubMed
description PURPOSE: To provide a practical solution that can be adopted in clinical routine to fulfill the AAPM-ESTRO recommendations regarding quality assurance of seeds used in prostate permanent brachytherapy. The aim is to design a new insert for the well-ionization chamber SourceCheck(4π) 33005 (PTW, Germany) that allows evaluating the mean air-kerma strength of up to ten (125)I seeds with one single measurement instead of measuring each seed individually. MATERIAL AND METHODS: The material required is: a) the SourceCheck(4π) 33005 well-ionization chamber provided with a PTW insert to measure the air-kerma strength S (K) of one single seed at a time; b) a newly designed insert that accommodates ten seeds in one column, which allows measuring the mean S (K) of the ten seeds in one single measurement; and c) a container with ten seeds from the same batch and class of the seeds used for the patient implant, and a set of nine non-radioactive seeds. The new insert is characterized by determining its calibration coefficient, used to convert the reading of the well-chamber when ten seeds are measured to their mean S (K). The proposed method is validated by comparing the mean S (K) of the ten seeds obtained from the new insert with the individual measurement of S (K) of each seed, evaluated with the PTW insert. RESULTS: The ratio between the calibration coefficient of the new insert and the calibration coefficient of the PTW insert for the SourceCheck(4π) 33005 is 1.135 ± 0.007 (k = 1). The mean S (K) of a set of ten seeds evaluated with this new system is in agreement with the mean value obtained from measuring independently the S (K) of each seed. CONCLUSIONS: The new insert and procedure allow evaluating the mean S (K) of ten seeds prior to the implant in a single measurement. The method is faster and more efficient from radiation protection point of view than measuring the individual S (K) of each seed.
format Online
Article
Text
id pubmed-4716135
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-47161352016-01-26 Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert Candela-Juan, Cristian Ballester, Facundo Perez-Calatayud, Jose Vijande, Javier J Contemp Brachytherapy Original Paper PURPOSE: To provide a practical solution that can be adopted in clinical routine to fulfill the AAPM-ESTRO recommendations regarding quality assurance of seeds used in prostate permanent brachytherapy. The aim is to design a new insert for the well-ionization chamber SourceCheck(4π) 33005 (PTW, Germany) that allows evaluating the mean air-kerma strength of up to ten (125)I seeds with one single measurement instead of measuring each seed individually. MATERIAL AND METHODS: The material required is: a) the SourceCheck(4π) 33005 well-ionization chamber provided with a PTW insert to measure the air-kerma strength S (K) of one single seed at a time; b) a newly designed insert that accommodates ten seeds in one column, which allows measuring the mean S (K) of the ten seeds in one single measurement; and c) a container with ten seeds from the same batch and class of the seeds used for the patient implant, and a set of nine non-radioactive seeds. The new insert is characterized by determining its calibration coefficient, used to convert the reading of the well-chamber when ten seeds are measured to their mean S (K). The proposed method is validated by comparing the mean S (K) of the ten seeds obtained from the new insert with the individual measurement of S (K) of each seed, evaluated with the PTW insert. RESULTS: The ratio between the calibration coefficient of the new insert and the calibration coefficient of the PTW insert for the SourceCheck(4π) 33005 is 1.135 ± 0.007 (k = 1). The mean S (K) of a set of ten seeds evaluated with this new system is in agreement with the mean value obtained from measuring independently the S (K) of each seed. CONCLUSIONS: The new insert and procedure allow evaluating the mean S (K) of ten seeds prior to the implant in a single measurement. The method is faster and more efficient from radiation protection point of view than measuring the individual S (K) of each seed. Termedia Publishing House 2015-12-30 2015-12 /pmc/articles/PMC4716135/ /pubmed/26816507 http://dx.doi.org/10.5114/jcb.2015.56768 Text en Copyright © 2015 Termedia Sp. z o. o. http://creativecommons.org/licenses/by-nc-sa/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
spellingShingle Original Paper
Candela-Juan, Cristian
Ballester, Facundo
Perez-Calatayud, Jose
Vijande, Javier
Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert
title Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert
title_full Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert
title_fullStr Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert
title_full_unstemmed Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert
title_short Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert
title_sort assaying multiple (125)i seeds with the well-ionization chamber sourcecheck(4π) 33005 and a new insert
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4716135/
https://www.ncbi.nlm.nih.gov/pubmed/26816507
http://dx.doi.org/10.5114/jcb.2015.56768
work_keys_str_mv AT candelajuancristian assayingmultiple125iseedswiththewellionizationchambersourcecheck4p33005andanewinsert
AT ballesterfacundo assayingmultiple125iseedswiththewellionizationchambersourcecheck4p33005andanewinsert
AT perezcalatayudjose assayingmultiple125iseedswiththewellionizationchambersourcecheck4p33005andanewinsert
AT vijandejavier assayingmultiple125iseedswiththewellionizationchambersourcecheck4p33005andanewinsert