Cargando…

Nonthermal ablation of deep brain targets: A simulation study on a large animal model

PURPOSE: Thermal ablation with transcranial MRI-guided focused ultrasound (FUS) is currently limited to central brain targets because of heating and other beam effects caused by the presence of the skull. Recently, it was shown that it is possible to ablate tissues without depositing thermal energy...

Descripción completa

Detalles Bibliográficos
Autores principales: Top, Can Barış, White, P. Jason, McDannold, Nathan J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Association of Physicists in Medicine 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4723413/
https://www.ncbi.nlm.nih.gov/pubmed/26843248
http://dx.doi.org/10.1118/1.4939809
_version_ 1782411507519193088
author Top, Can Barış
White, P. Jason
McDannold, Nathan J.
author_facet Top, Can Barış
White, P. Jason
McDannold, Nathan J.
author_sort Top, Can Barış
collection PubMed
description PURPOSE: Thermal ablation with transcranial MRI-guided focused ultrasound (FUS) is currently limited to central brain targets because of heating and other beam effects caused by the presence of the skull. Recently, it was shown that it is possible to ablate tissues without depositing thermal energy by driving intravenously administered microbubbles to inertial cavitation using low-duty-cycle burst sonications. A recent study demonstrated that this ablation method could ablate tissue volumes near the skull base in nonhuman primates without thermally damaging the nearby bone. However, blood–brain disruption was observed in the prefocal region, and in some cases, this region contained small areas of tissue damage. The objective of this study was to analyze the experimental model with simulations and to interpret the cause of these effects. METHODS: The authors simulated prior experiments where nonthermal ablation was performed in the brain in anesthetized rhesus macaques using a 220 kHz clinical prototype transcranial MRI-guided FUS system. Low-duty-cycle sonications were applied at deep brain targets with the ultrasound contrast agent Definity. For simulations, a 3D pseudospectral finite difference time domain tool was used. The effects of shear mode conversion, focal steering, skull aberrations, nonlinear propagation, and the presence of skull base on the pressure field were investigated using acoustic and elastic wave propagation models. RESULTS: The simulation results were in agreement with the experimental findings in the prefocal region. In the postfocal region, however, side lobes were predicted by the simulations, but no effects were evident in the experiments. The main beam was not affected by the different simulated scenarios except for a shift of about 1 mm in peak position due to skull aberrations. However, the authors observed differences in the volume, amplitude, and distribution of the side lobes. In the experiments, a single element passive cavitation detector was used to measure the inertial cavitation threshold and to determine the pressure amplitude to use for ablation. Simulations of the detector’s acoustic field suggest that its maximum sensitivity was in the lower part of the main beam, which may have led to excessive exposure levels in the experiments that may have contributed to damage in the prefocal area. CONCLUSIONS: Overall, these results suggest that case-specific full wave simulations before the procedure can be useful to predict the focal and the prefocal side lobes and the extent of the resulting bioeffects produced by nonthermal ablation. Such simulations can also be used to optimally position passive cavitation detectors. The disagreement between the simulations and the experiments in the postfocal region may have been due to shielding of the ultrasound field due to microbubble activity in the focal region. Future efforts should include the effects of microbubble activity and vascularization on the pressure field.
format Online
Article
Text
id pubmed-4723413
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher American Association of Physicists in Medicine
record_format MEDLINE/PubMed
spelling pubmed-47234132016-02-08 Nonthermal ablation of deep brain targets: A simulation study on a large animal model Top, Can Barış White, P. Jason McDannold, Nathan J. Med Phys THERAPEUTIC INTERVENTIONS PURPOSE: Thermal ablation with transcranial MRI-guided focused ultrasound (FUS) is currently limited to central brain targets because of heating and other beam effects caused by the presence of the skull. Recently, it was shown that it is possible to ablate tissues without depositing thermal energy by driving intravenously administered microbubbles to inertial cavitation using low-duty-cycle burst sonications. A recent study demonstrated that this ablation method could ablate tissue volumes near the skull base in nonhuman primates without thermally damaging the nearby bone. However, blood–brain disruption was observed in the prefocal region, and in some cases, this region contained small areas of tissue damage. The objective of this study was to analyze the experimental model with simulations and to interpret the cause of these effects. METHODS: The authors simulated prior experiments where nonthermal ablation was performed in the brain in anesthetized rhesus macaques using a 220 kHz clinical prototype transcranial MRI-guided FUS system. Low-duty-cycle sonications were applied at deep brain targets with the ultrasound contrast agent Definity. For simulations, a 3D pseudospectral finite difference time domain tool was used. The effects of shear mode conversion, focal steering, skull aberrations, nonlinear propagation, and the presence of skull base on the pressure field were investigated using acoustic and elastic wave propagation models. RESULTS: The simulation results were in agreement with the experimental findings in the prefocal region. In the postfocal region, however, side lobes were predicted by the simulations, but no effects were evident in the experiments. The main beam was not affected by the different simulated scenarios except for a shift of about 1 mm in peak position due to skull aberrations. However, the authors observed differences in the volume, amplitude, and distribution of the side lobes. In the experiments, a single element passive cavitation detector was used to measure the inertial cavitation threshold and to determine the pressure amplitude to use for ablation. Simulations of the detector’s acoustic field suggest that its maximum sensitivity was in the lower part of the main beam, which may have led to excessive exposure levels in the experiments that may have contributed to damage in the prefocal area. CONCLUSIONS: Overall, these results suggest that case-specific full wave simulations before the procedure can be useful to predict the focal and the prefocal side lobes and the extent of the resulting bioeffects produced by nonthermal ablation. Such simulations can also be used to optimally position passive cavitation detectors. The disagreement between the simulations and the experiments in the postfocal region may have been due to shielding of the ultrasound field due to microbubble activity in the focal region. Future efforts should include the effects of microbubble activity and vascularization on the pressure field. American Association of Physicists in Medicine 2016-02 2016-01-21 /pmc/articles/PMC4723413/ /pubmed/26843248 http://dx.doi.org/10.1118/1.4939809 Text en © 2016 American Association of Physicists in Medicine. 0094-2405/2016/43(2)/870/13/$30.00 All article content, except where otherwise noted, is licensed under a Creative Commons Attribution (CC BY) license ( http://creativecommons.org/licenses/by/4.0/ ).
spellingShingle THERAPEUTIC INTERVENTIONS
Top, Can Barış
White, P. Jason
McDannold, Nathan J.
Nonthermal ablation of deep brain targets: A simulation study on a large animal model
title Nonthermal ablation of deep brain targets: A simulation study on a large animal model
title_full Nonthermal ablation of deep brain targets: A simulation study on a large animal model
title_fullStr Nonthermal ablation of deep brain targets: A simulation study on a large animal model
title_full_unstemmed Nonthermal ablation of deep brain targets: A simulation study on a large animal model
title_short Nonthermal ablation of deep brain targets: A simulation study on a large animal model
title_sort nonthermal ablation of deep brain targets: a simulation study on a large animal model
topic THERAPEUTIC INTERVENTIONS
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4723413/
https://www.ncbi.nlm.nih.gov/pubmed/26843248
http://dx.doi.org/10.1118/1.4939809
work_keys_str_mv AT topcanbarıs nonthermalablationofdeepbraintargetsasimulationstudyonalargeanimalmodel
AT whitepjason nonthermalablationofdeepbraintargetsasimulationstudyonalargeanimalmodel
AT mcdannoldnathanj nonthermalablationofdeepbraintargetsasimulationstudyonalargeanimalmodel