Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity. MATERIAL/METHODS: Subjects were 50 patient volun...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
International Scientific Literature, Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4725617/ https://www.ncbi.nlm.nih.gov/pubmed/26782369 http://dx.doi.org/10.12659/MSM.895573 |
_version_ | 1782411655138770944 |
---|---|
author | Lee, Kyeong-Bong Kim, Min-Kyu Jeong, Ju-Ri Lee, Wan-Hee |
author_facet | Lee, Kyeong-Bong Kim, Min-Kyu Jeong, Ju-Ri Lee, Wan-Hee |
author_sort | Lee, Kyeong-Bong |
collection | PubMed |
description | BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity. MATERIAL/METHODS: Subjects were 50 patient volunteers in a rehabilitation hospital who had experienced their first episode of unilateral stroke with hemiparesis during the previous 6 months (26 men, 24 women; mean age [±SD], 55.96 [±12.81] years), with no use of medications that could induce drowsiness, evidence of restrictive lung disease, history of asthma, use of psychotropic drugs, or alcohol consumption habit. Maximum inspiratory pressure, peak inspiratory flow, and vital capacity for pulmonary functions were assessed using an electronic inspiratory loading device (PowerBreathe, K5, 2010) by 2 examiners, with patients in an unassisted sitting position, and 1 examiner re-assessed with same patients at the same time of a day after 1 week. Intra-class correlation coefficients were used to assess reliability. RESULTS: Intra-rater reliability ranged from intra-class correlation coefficients (ICCs)=0.959 to 0.986 in variables. For the inter-rater reliability between 2 examiners, the ICCs ranged from 0.933 to 0.985. Intra-rater and inter-rater reliability were good in variables (maximal inspiratory pressure, peak inspiratory flow, and vital capacity). CONCLUSIONS: The intra- and inter-examiner reliability of the pulmonary function measurements, maximum inspiratory pressure, peak inspiratory flow, and vital capacity, for the post-stroke patients was very high. The results suggest that the electronic inspiratory loading device would be useful for clinical rehabilitative assessment of pulmonary function. |
format | Online Article Text |
id | pubmed-4725617 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | International Scientific Literature, Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-47256172016-01-31 Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients Lee, Kyeong-Bong Kim, Min-Kyu Jeong, Ju-Ri Lee, Wan-Hee Med Sci Monit Clinical Research BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity. MATERIAL/METHODS: Subjects were 50 patient volunteers in a rehabilitation hospital who had experienced their first episode of unilateral stroke with hemiparesis during the previous 6 months (26 men, 24 women; mean age [±SD], 55.96 [±12.81] years), with no use of medications that could induce drowsiness, evidence of restrictive lung disease, history of asthma, use of psychotropic drugs, or alcohol consumption habit. Maximum inspiratory pressure, peak inspiratory flow, and vital capacity for pulmonary functions were assessed using an electronic inspiratory loading device (PowerBreathe, K5, 2010) by 2 examiners, with patients in an unassisted sitting position, and 1 examiner re-assessed with same patients at the same time of a day after 1 week. Intra-class correlation coefficients were used to assess reliability. RESULTS: Intra-rater reliability ranged from intra-class correlation coefficients (ICCs)=0.959 to 0.986 in variables. For the inter-rater reliability between 2 examiners, the ICCs ranged from 0.933 to 0.985. Intra-rater and inter-rater reliability were good in variables (maximal inspiratory pressure, peak inspiratory flow, and vital capacity). CONCLUSIONS: The intra- and inter-examiner reliability of the pulmonary function measurements, maximum inspiratory pressure, peak inspiratory flow, and vital capacity, for the post-stroke patients was very high. The results suggest that the electronic inspiratory loading device would be useful for clinical rehabilitative assessment of pulmonary function. International Scientific Literature, Inc. 2016-01-19 /pmc/articles/PMC4725617/ /pubmed/26782369 http://dx.doi.org/10.12659/MSM.895573 Text en © Med Sci Monit, 2016 This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivs 3.0 Unported License |
spellingShingle | Clinical Research Lee, Kyeong-Bong Kim, Min-Kyu Jeong, Ju-Ri Lee, Wan-Hee Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients |
title | Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients |
title_full | Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients |
title_fullStr | Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients |
title_full_unstemmed | Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients |
title_short | Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients |
title_sort | reliability of an electronic inspiratory loading device for assessing pulmonary function in post-stroke patients |
topic | Clinical Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4725617/ https://www.ncbi.nlm.nih.gov/pubmed/26782369 http://dx.doi.org/10.12659/MSM.895573 |
work_keys_str_mv | AT leekyeongbong reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients AT kimminkyu reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients AT jeongjuri reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients AT leewanhee reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients |