Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients

BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity. MATERIAL/METHODS: Subjects were 50 patient volun...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Kyeong-Bong, Kim, Min-Kyu, Jeong, Ju-Ri, Lee, Wan-Hee
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Scientific Literature, Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4725617/
https://www.ncbi.nlm.nih.gov/pubmed/26782369
http://dx.doi.org/10.12659/MSM.895573
_version_ 1782411655138770944
author Lee, Kyeong-Bong
Kim, Min-Kyu
Jeong, Ju-Ri
Lee, Wan-Hee
author_facet Lee, Kyeong-Bong
Kim, Min-Kyu
Jeong, Ju-Ri
Lee, Wan-Hee
author_sort Lee, Kyeong-Bong
collection PubMed
description BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity. MATERIAL/METHODS: Subjects were 50 patient volunteers in a rehabilitation hospital who had experienced their first episode of unilateral stroke with hemiparesis during the previous 6 months (26 men, 24 women; mean age [±SD], 55.96 [±12.81] years), with no use of medications that could induce drowsiness, evidence of restrictive lung disease, history of asthma, use of psychotropic drugs, or alcohol consumption habit. Maximum inspiratory pressure, peak inspiratory flow, and vital capacity for pulmonary functions were assessed using an electronic inspiratory loading device (PowerBreathe, K5, 2010) by 2 examiners, with patients in an unassisted sitting position, and 1 examiner re-assessed with same patients at the same time of a day after 1 week. Intra-class correlation coefficients were used to assess reliability. RESULTS: Intra-rater reliability ranged from intra-class correlation coefficients (ICCs)=0.959 to 0.986 in variables. For the inter-rater reliability between 2 examiners, the ICCs ranged from 0.933 to 0.985. Intra-rater and inter-rater reliability were good in variables (maximal inspiratory pressure, peak inspiratory flow, and vital capacity). CONCLUSIONS: The intra- and inter-examiner reliability of the pulmonary function measurements, maximum inspiratory pressure, peak inspiratory flow, and vital capacity, for the post-stroke patients was very high. The results suggest that the electronic inspiratory loading device would be useful for clinical rehabilitative assessment of pulmonary function.
format Online
Article
Text
id pubmed-4725617
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher International Scientific Literature, Inc.
record_format MEDLINE/PubMed
spelling pubmed-47256172016-01-31 Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients Lee, Kyeong-Bong Kim, Min-Kyu Jeong, Ju-Ri Lee, Wan-Hee Med Sci Monit Clinical Research BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity. MATERIAL/METHODS: Subjects were 50 patient volunteers in a rehabilitation hospital who had experienced their first episode of unilateral stroke with hemiparesis during the previous 6 months (26 men, 24 women; mean age [±SD], 55.96 [±12.81] years), with no use of medications that could induce drowsiness, evidence of restrictive lung disease, history of asthma, use of psychotropic drugs, or alcohol consumption habit. Maximum inspiratory pressure, peak inspiratory flow, and vital capacity for pulmonary functions were assessed using an electronic inspiratory loading device (PowerBreathe, K5, 2010) by 2 examiners, with patients in an unassisted sitting position, and 1 examiner re-assessed with same patients at the same time of a day after 1 week. Intra-class correlation coefficients were used to assess reliability. RESULTS: Intra-rater reliability ranged from intra-class correlation coefficients (ICCs)=0.959 to 0.986 in variables. For the inter-rater reliability between 2 examiners, the ICCs ranged from 0.933 to 0.985. Intra-rater and inter-rater reliability were good in variables (maximal inspiratory pressure, peak inspiratory flow, and vital capacity). CONCLUSIONS: The intra- and inter-examiner reliability of the pulmonary function measurements, maximum inspiratory pressure, peak inspiratory flow, and vital capacity, for the post-stroke patients was very high. The results suggest that the electronic inspiratory loading device would be useful for clinical rehabilitative assessment of pulmonary function. International Scientific Literature, Inc. 2016-01-19 /pmc/articles/PMC4725617/ /pubmed/26782369 http://dx.doi.org/10.12659/MSM.895573 Text en © Med Sci Monit, 2016 This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivs 3.0 Unported License
spellingShingle Clinical Research
Lee, Kyeong-Bong
Kim, Min-Kyu
Jeong, Ju-Ri
Lee, Wan-Hee
Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
title Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
title_full Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
title_fullStr Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
title_full_unstemmed Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
title_short Reliability of an Electronic Inspiratory Loading Device for Assessing Pulmonary Function in Post-Stroke Patients
title_sort reliability of an electronic inspiratory loading device for assessing pulmonary function in post-stroke patients
topic Clinical Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4725617/
https://www.ncbi.nlm.nih.gov/pubmed/26782369
http://dx.doi.org/10.12659/MSM.895573
work_keys_str_mv AT leekyeongbong reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients
AT kimminkyu reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients
AT jeongjuri reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients
AT leewanhee reliabilityofanelectronicinspiratoryloadingdeviceforassessingpulmonaryfunctioninpoststrokepatients