Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography
This study was undertaken to evaluate the effect of external metal filters on the image quality of computed tomography (CT) and single photon emission computed tomography (SPECT)/CT images. Images of Jaszack phantom filled with water and containing iodine contrast filled syringes were acquired using...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4728890/ https://www.ncbi.nlm.nih.gov/pubmed/26865755 http://dx.doi.org/10.4103/0971-6203.170790 |
_version_ | 1782412192984858624 |
---|---|
author | Rana, Nivedita Rawat, Dinesh Parmar, Madan Dhawan, Devinder Kumar Bhati, Ashok Kumar Mittal, Bhagwant Rai |
author_facet | Rana, Nivedita Rawat, Dinesh Parmar, Madan Dhawan, Devinder Kumar Bhati, Ashok Kumar Mittal, Bhagwant Rai |
author_sort | Rana, Nivedita |
collection | PubMed |
description | This study was undertaken to evaluate the effect of external metal filters on the image quality of computed tomography (CT) and single photon emission computed tomography (SPECT)/CT images. Images of Jaszack phantom filled with water and containing iodine contrast filled syringes were acquired using CT (120 kV, 2.5 mA) component of SPECT/CT system, ensuring fixation of filter on X-ray collimator. Different thickness of filters of Al and Cu (1 mm, 2 mm, 3 mm, and 4 mm) and filter combinations Cu 1 mm, Cu 2 mm, Cu 3 mm each in combination with Al (1 mm, 2 mm, 3 mm, and 4 mm), respectively, were used. All image sets were visually analyzed for streak artifacts and contrast to noise ratio (CNR) was derived. Similar acquisition was done using Philips CT quality control (QC) phantom and CNR were calculated for its lexan, perspex, and teflon inserts. Attenuation corrected SPECT/CT images of Jaszack phantom filled with 444–555 MBq (12–15 mCi) of (99m)Tc were obtained by applying attenuation correction map generated by hardened X-ray beam for different filter combination, on SPECT data. Uniformity, root mean square (rms) and contrast were calculated in all image sets. Less streak artifacts at iodine water interface were observed in images acquired using external filters as compared to those without a filter. CNR for syringes, spheres, and inserts of Philips CT QC phantom was almost similar to Al 2 mm, Al 3 mm, and without the use of filters. CNR decreased with increasing copper thickness and other filter combinations. Uniformity and rms were lower, and value of contrast was higher for SPECT/CT images when CT was acquired with Al 2 mm and 3 mm filter than for images acquired without a filter. The study suggests that for Infinia Hawkeye 4, SPECT/CT system, Al 2 mm, and 3 mm are the optimum filters for improving image quality of SPECT/CT images of Jaszack or Philips CT QC phantom keeping other parameters of CT constant. |
format | Online Article Text |
id | pubmed-4728890 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-47288902016-02-10 Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography Rana, Nivedita Rawat, Dinesh Parmar, Madan Dhawan, Devinder Kumar Bhati, Ashok Kumar Mittal, Bhagwant Rai J Med Phys Original Article This study was undertaken to evaluate the effect of external metal filters on the image quality of computed tomography (CT) and single photon emission computed tomography (SPECT)/CT images. Images of Jaszack phantom filled with water and containing iodine contrast filled syringes were acquired using CT (120 kV, 2.5 mA) component of SPECT/CT system, ensuring fixation of filter on X-ray collimator. Different thickness of filters of Al and Cu (1 mm, 2 mm, 3 mm, and 4 mm) and filter combinations Cu 1 mm, Cu 2 mm, Cu 3 mm each in combination with Al (1 mm, 2 mm, 3 mm, and 4 mm), respectively, were used. All image sets were visually analyzed for streak artifacts and contrast to noise ratio (CNR) was derived. Similar acquisition was done using Philips CT quality control (QC) phantom and CNR were calculated for its lexan, perspex, and teflon inserts. Attenuation corrected SPECT/CT images of Jaszack phantom filled with 444–555 MBq (12–15 mCi) of (99m)Tc were obtained by applying attenuation correction map generated by hardened X-ray beam for different filter combination, on SPECT data. Uniformity, root mean square (rms) and contrast were calculated in all image sets. Less streak artifacts at iodine water interface were observed in images acquired using external filters as compared to those without a filter. CNR for syringes, spheres, and inserts of Philips CT QC phantom was almost similar to Al 2 mm, Al 3 mm, and without the use of filters. CNR decreased with increasing copper thickness and other filter combinations. Uniformity and rms were lower, and value of contrast was higher for SPECT/CT images when CT was acquired with Al 2 mm and 3 mm filter than for images acquired without a filter. The study suggests that for Infinia Hawkeye 4, SPECT/CT system, Al 2 mm, and 3 mm are the optimum filters for improving image quality of SPECT/CT images of Jaszack or Philips CT QC phantom keeping other parameters of CT constant. Medknow Publications & Media Pvt Ltd 2015 /pmc/articles/PMC4728890/ /pubmed/26865755 http://dx.doi.org/10.4103/0971-6203.170790 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Rana, Nivedita Rawat, Dinesh Parmar, Madan Dhawan, Devinder Kumar Bhati, Ashok Kumar Mittal, Bhagwant Rai Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography |
title | Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography |
title_full | Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography |
title_fullStr | Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography |
title_full_unstemmed | Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography |
title_short | Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography |
title_sort | evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4728890/ https://www.ncbi.nlm.nih.gov/pubmed/26865755 http://dx.doi.org/10.4103/0971-6203.170790 |
work_keys_str_mv | AT rananivedita evaluationofexternalbeamhardeningfiltersonimagequalityofcomputedtomographyandsinglephotonemissioncomputedtomographycomputedtomography AT rawatdinesh evaluationofexternalbeamhardeningfiltersonimagequalityofcomputedtomographyandsinglephotonemissioncomputedtomographycomputedtomography AT parmarmadan evaluationofexternalbeamhardeningfiltersonimagequalityofcomputedtomographyandsinglephotonemissioncomputedtomographycomputedtomography AT dhawandevinderkumar evaluationofexternalbeamhardeningfiltersonimagequalityofcomputedtomographyandsinglephotonemissioncomputedtomographycomputedtomography AT bhatiashokkumar evaluationofexternalbeamhardeningfiltersonimagequalityofcomputedtomographyandsinglephotonemissioncomputedtomographycomputedtomography AT mittalbhagwantrai evaluationofexternalbeamhardeningfiltersonimagequalityofcomputedtomographyandsinglephotonemissioncomputedtomographycomputedtomography |