Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels

The aim of this study was to examine the impact of dose rate on accuracy of intensity modulated radiation therapy (IMRT) plan delivery by comparing the gamma agreement between the calculated and measured portal doses by pretreatment quality assurance (QA) using electronic portal imaging device dosim...

Descripción completa

Detalles Bibliográficos
Autores principales: Kaviarasu, Karunakaran, Raj, N. Arunai Nambi, Murthy, K. Krishna, Babu, A. Ananda Giri, Prasad, Bhaskar Laxman Durga
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4728894/
https://www.ncbi.nlm.nih.gov/pubmed/26865759
http://dx.doi.org/10.4103/0971-6203.170786
_version_ 1782412193867759616
author Kaviarasu, Karunakaran
Raj, N. Arunai Nambi
Murthy, K. Krishna
Babu, A. Ananda Giri
Prasad, Bhaskar Laxman Durga
author_facet Kaviarasu, Karunakaran
Raj, N. Arunai Nambi
Murthy, K. Krishna
Babu, A. Ananda Giri
Prasad, Bhaskar Laxman Durga
author_sort Kaviarasu, Karunakaran
collection PubMed
description The aim of this study was to examine the impact of dose rate on accuracy of intensity modulated radiation therapy (IMRT) plan delivery by comparing the gamma agreement between the calculated and measured portal doses by pretreatment quality assurance (QA) using electronic portal imaging device dosimetry and creating a workflow for the pretreatment IMRT QA at hospital levels. As the improvement in gamma agreement leads to increase in the quality of IMRT treatment delivery, gamma evaluation was carried out for the calculated and the measured portal images for the criteria of 3% dose difference and 3 mm distance-to-agreement (DTA). Three gamma parameters: Maximum gamma, average gamma, and percentage of the field area with a gamma value>1.0 were analyzed. Three gamma index parameters were evaluated for 40 IMRT plans (315 IMRT fields) which were calculated for 400 monitor units (MU)/min dose rate and maximum multileaf collimator (MLC) speed of 2.5 cm/s. Gamma parameters for all 315 fields are within acceptable limits set at our center. Further, to improve the gamma results, we set an action level for this study using the mean and standard deviation (SD) values from the 315 fields studied. Forty out of 315 IMRT fields showed low gamma agreement (gamma parameters>2 SD as per action level of the study). The parameters were recalculated and reanalyzed for the dose rates of 300, 400 and 500 MU/min. Lowering the dose rate helped in getting an enhanced gamma agreement between the calculated and measured portal doses of complicated fields. This may be attributed to the less complex motion of MLC over time and the MU of the field/segment. An IMRT QA work flow was prepared which will help in improving the quality of IMRT delivery.
format Online
Article
Text
id pubmed-4728894
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-47288942016-02-10 Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels Kaviarasu, Karunakaran Raj, N. Arunai Nambi Murthy, K. Krishna Babu, A. Ananda Giri Prasad, Bhaskar Laxman Durga J Med Phys Technical Note The aim of this study was to examine the impact of dose rate on accuracy of intensity modulated radiation therapy (IMRT) plan delivery by comparing the gamma agreement between the calculated and measured portal doses by pretreatment quality assurance (QA) using electronic portal imaging device dosimetry and creating a workflow for the pretreatment IMRT QA at hospital levels. As the improvement in gamma agreement leads to increase in the quality of IMRT treatment delivery, gamma evaluation was carried out for the calculated and the measured portal images for the criteria of 3% dose difference and 3 mm distance-to-agreement (DTA). Three gamma parameters: Maximum gamma, average gamma, and percentage of the field area with a gamma value>1.0 were analyzed. Three gamma index parameters were evaluated for 40 IMRT plans (315 IMRT fields) which were calculated for 400 monitor units (MU)/min dose rate and maximum multileaf collimator (MLC) speed of 2.5 cm/s. Gamma parameters for all 315 fields are within acceptable limits set at our center. Further, to improve the gamma results, we set an action level for this study using the mean and standard deviation (SD) values from the 315 fields studied. Forty out of 315 IMRT fields showed low gamma agreement (gamma parameters>2 SD as per action level of the study). The parameters were recalculated and reanalyzed for the dose rates of 300, 400 and 500 MU/min. Lowering the dose rate helped in getting an enhanced gamma agreement between the calculated and measured portal doses of complicated fields. This may be attributed to the less complex motion of MLC over time and the MU of the field/segment. An IMRT QA work flow was prepared which will help in improving the quality of IMRT delivery. Medknow Publications & Media Pvt Ltd 2015 /pmc/articles/PMC4728894/ /pubmed/26865759 http://dx.doi.org/10.4103/0971-6203.170786 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Technical Note
Kaviarasu, Karunakaran
Raj, N. Arunai Nambi
Murthy, K. Krishna
Babu, A. Ananda Giri
Prasad, Bhaskar Laxman Durga
Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels
title Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels
title_full Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels
title_fullStr Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels
title_full_unstemmed Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels
title_short Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels
title_sort impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4728894/
https://www.ncbi.nlm.nih.gov/pubmed/26865759
http://dx.doi.org/10.4103/0971-6203.170786
work_keys_str_mv AT kaviarasukarunakaran impactofdoserateonaccuracyofintensitymodulatedradiationtherapyplandeliveryusingthepretreatmentportaldosimetryqualityassuranceandsettinguptheworkflowathospitallevels
AT rajnarunainambi impactofdoserateonaccuracyofintensitymodulatedradiationtherapyplandeliveryusingthepretreatmentportaldosimetryqualityassuranceandsettinguptheworkflowathospitallevels
AT murthykkrishna impactofdoserateonaccuracyofintensitymodulatedradiationtherapyplandeliveryusingthepretreatmentportaldosimetryqualityassuranceandsettinguptheworkflowathospitallevels
AT babuaanandagiri impactofdoserateonaccuracyofintensitymodulatedradiationtherapyplandeliveryusingthepretreatmentportaldosimetryqualityassuranceandsettinguptheworkflowathospitallevels
AT prasadbhaskarlaxmandurga impactofdoserateonaccuracyofintensitymodulatedradiationtherapyplandeliveryusingthepretreatmentportaldosimetryqualityassuranceandsettinguptheworkflowathospitallevels