Revision of the NIST Standard for (223)Ra: New Measurements and Review of 2008 Data

After discovering a discrepancy in the transfer standard currently being disseminated by the National Institute of Standards and Technology (NIST), we have performed a new primary standardization of the alpha-emitter (223)Ra using Live-timed Anticoincidence Counting (LTAC) and the Triple-to-Double C...

Descripción completa

Detalles Bibliográficos
Autores principales: Zimmerman, B. E., Bergeron, D. E., Cessna, J. T., Fitzgerald, R., Pibida, L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4730682/
https://www.ncbi.nlm.nih.gov/pubmed/26958437
http://dx.doi.org/10.6028/jres.120.004
Descripción
Sumario:After discovering a discrepancy in the transfer standard currently being disseminated by the National Institute of Standards and Technology (NIST), we have performed a new primary standardization of the alpha-emitter (223)Ra using Live-timed Anticoincidence Counting (LTAC) and the Triple-to-Double Coincidence Ratio Method (TDCR). Additional confirmatory measurements were made with the CIEMAT-NIST efficiency tracing method (CNET) of liquid scintillation counting, integral γ-ray counting using a NaI(Tl) well counter, and several High Purity Germanium (HPGe) detectors in an attempt to understand the origin of the discrepancy and to provide a correction. The results indicate that a −9.5 % difference exists between activity values obtained using the former transfer standard relative to the new primary standardization. During one of the experiments, a 2 % difference in activity was observed between dilutions of the (223)Ra master solution prepared using the composition used in the original standardization and those prepared using 1 mol·L(−1) HCl. This effect appeared to be dependent on the number of dilutions or the total dilution factor to the master solution, but the magnitude was not reproducible. A new calibration factor (“K-value”) has been determined for the NIST Secondary Standard Ionization Chamber (IC “A”), thereby correcting the discrepancy between the primary and secondary standards.