Cargando…

Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes

OBJECTIVE: We sought to investigate the incidence of complications associated with V2 Guideliner, understand the mechanisms and evaluate the impact of alterations made to the V3 Guideliner. METHODS: Retrospective analysis of consecutive cases employing V2 Guideliner from two university teaching hosp...

Descripción completa

Detalles Bibliográficos
Autores principales: Alkhalil, Mohammad, Smyth, Alison, Walsh, Simon J, McQuillan, Conor, Spence, Mark S, Owens, Colum G, Hanratty, Colm G
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4731838/
https://www.ncbi.nlm.nih.gov/pubmed/26848394
http://dx.doi.org/10.1136/openhrt-2015-000331
_version_ 1782412602411843584
author Alkhalil, Mohammad
Smyth, Alison
Walsh, Simon J
McQuillan, Conor
Spence, Mark S
Owens, Colum G
Hanratty, Colm G
author_facet Alkhalil, Mohammad
Smyth, Alison
Walsh, Simon J
McQuillan, Conor
Spence, Mark S
Owens, Colum G
Hanratty, Colm G
author_sort Alkhalil, Mohammad
collection PubMed
description OBJECTIVE: We sought to investigate the incidence of complications associated with V2 Guideliner, understand the mechanisms and evaluate the impact of alterations made to the V3 Guideliner. METHODS: Retrospective analysis of consecutive cases employing V2 Guideliner from two university teaching hospitals. Complications were identified, analysed and classified into major versus minor ones. To understand the potential anatomical mechanism of these complications, analysis of normal great vessel anatomy was undertaken in separate cohort of patients undergoing cardiac catheterisation via right radial approach. Further analysis of consecutive cases employing V3 Gudieliner took place and the incidence of complications were compared between V2 and V3 groups. RESULTS: Total of 188 cases of V2 Guideliner use were identified. One major complication was reported (coronary dissection). Proximal collar interaction and stent damage occurred in 19 cases (10%). Anatomical data suggest that extending the V2 Guideliner tubing sited the proximal collar of the device in the brachiocephalic/subclavian artery, a potential site of tortuosity and potential cause of the Guideliner proximal collar-stent interaction. Further analysis of 124 cases of V3 Guideliner use demonstrated no cases with proximal collar-stent interactions, one case of longitudinal stent deformation and two incidents of stent interaction with the distal edge of the V3 Guideliner. CONCLUSIONS: We have demonstrated a higher incidence of V2 Guideliner complications compared to previous series. The change in design of the V2 Guideliner was a likely contributor but the modifications with V3 Guideliner appear to have ameliorated this issue.
format Online
Article
Text
id pubmed-4731838
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-47318382016-02-04 Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes Alkhalil, Mohammad Smyth, Alison Walsh, Simon J McQuillan, Conor Spence, Mark S Owens, Colum G Hanratty, Colm G Open Heart Interventional Cardiology OBJECTIVE: We sought to investigate the incidence of complications associated with V2 Guideliner, understand the mechanisms and evaluate the impact of alterations made to the V3 Guideliner. METHODS: Retrospective analysis of consecutive cases employing V2 Guideliner from two university teaching hospitals. Complications were identified, analysed and classified into major versus minor ones. To understand the potential anatomical mechanism of these complications, analysis of normal great vessel anatomy was undertaken in separate cohort of patients undergoing cardiac catheterisation via right radial approach. Further analysis of consecutive cases employing V3 Gudieliner took place and the incidence of complications were compared between V2 and V3 groups. RESULTS: Total of 188 cases of V2 Guideliner use were identified. One major complication was reported (coronary dissection). Proximal collar interaction and stent damage occurred in 19 cases (10%). Anatomical data suggest that extending the V2 Guideliner tubing sited the proximal collar of the device in the brachiocephalic/subclavian artery, a potential site of tortuosity and potential cause of the Guideliner proximal collar-stent interaction. Further analysis of 124 cases of V3 Guideliner use demonstrated no cases with proximal collar-stent interactions, one case of longitudinal stent deformation and two incidents of stent interaction with the distal edge of the V3 Guideliner. CONCLUSIONS: We have demonstrated a higher incidence of V2 Guideliner complications compared to previous series. The change in design of the V2 Guideliner was a likely contributor but the modifications with V3 Guideliner appear to have ameliorated this issue. BMJ Publishing Group 2016-01-27 /pmc/articles/PMC4731838/ /pubmed/26848394 http://dx.doi.org/10.1136/openhrt-2015-000331 Text en Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/ This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
spellingShingle Interventional Cardiology
Alkhalil, Mohammad
Smyth, Alison
Walsh, Simon J
McQuillan, Conor
Spence, Mark S
Owens, Colum G
Hanratty, Colm G
Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes
title Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes
title_full Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes
title_fullStr Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes
title_full_unstemmed Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes
title_short Did the use of the Guideliner V2(TM) guide catheter extension increase complications? A review of the incidence of complications related to the use of the V2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the V3(TM) Guideliner and clinical outcomes
title_sort did the use of the guideliner v2(tm) guide catheter extension increase complications? a review of the incidence of complications related to the use of the v2 catheter, the influence of right brachiocephalic arterial anatomy and the redesign of the v3(tm) guideliner and clinical outcomes
topic Interventional Cardiology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4731838/
https://www.ncbi.nlm.nih.gov/pubmed/26848394
http://dx.doi.org/10.1136/openhrt-2015-000331
work_keys_str_mv AT alkhalilmohammad didtheuseoftheguidelinerv2tmguidecatheterextensionincreasecomplicationsareviewoftheincidenceofcomplicationsrelatedtotheuseofthev2cathetertheinfluenceofrightbrachiocephalicarterialanatomyandtheredesignofthev3tmguidelinerandclinicaloutcomes
AT smythalison didtheuseoftheguidelinerv2tmguidecatheterextensionincreasecomplicationsareviewoftheincidenceofcomplicationsrelatedtotheuseofthev2cathetertheinfluenceofrightbrachiocephalicarterialanatomyandtheredesignofthev3tmguidelinerandclinicaloutcomes
AT walshsimonj didtheuseoftheguidelinerv2tmguidecatheterextensionincreasecomplicationsareviewoftheincidenceofcomplicationsrelatedtotheuseofthev2cathetertheinfluenceofrightbrachiocephalicarterialanatomyandtheredesignofthev3tmguidelinerandclinicaloutcomes
AT mcquillanconor didtheuseoftheguidelinerv2tmguidecatheterextensionincreasecomplicationsareviewoftheincidenceofcomplicationsrelatedtotheuseofthev2cathetertheinfluenceofrightbrachiocephalicarterialanatomyandtheredesignofthev3tmguidelinerandclinicaloutcomes
AT spencemarks didtheuseoftheguidelinerv2tmguidecatheterextensionincreasecomplicationsareviewoftheincidenceofcomplicationsrelatedtotheuseofthev2cathetertheinfluenceofrightbrachiocephalicarterialanatomyandtheredesignofthev3tmguidelinerandclinicaloutcomes
AT owenscolumg didtheuseoftheguidelinerv2tmguidecatheterextensionincreasecomplicationsareviewoftheincidenceofcomplicationsrelatedtotheuseofthev2cathetertheinfluenceofrightbrachiocephalicarterialanatomyandtheredesignofthev3tmguidelinerandclinicaloutcomes
AT hanrattycolmg didtheuseoftheguidelinerv2tmguidecatheterextensionincreasecomplicationsareviewoftheincidenceofcomplicationsrelatedtotheuseofthev2cathetertheinfluenceofrightbrachiocephalicarterialanatomyandtheredesignofthev3tmguidelinerandclinicaloutcomes