Validation of the i-STAT system for the analysis of blood parameters in fish

Portable clinical analysers, such as the i-STAT system, are increasingly being used for blood analysis in animal ecology and physiology because of their portability and easy operation. Although originally conceived for clinical application and to replace robust but lengthy techniques, researchers ha...

Descripción completa

Detalles Bibliográficos
Autores principales: Harter, T. S., Shartau, R. B., Brauner, C. J., Farrell, A. P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4732487/
https://www.ncbi.nlm.nih.gov/pubmed/27293658
http://dx.doi.org/10.1093/conphys/cou037
Descripción
Sumario:Portable clinical analysers, such as the i-STAT system, are increasingly being used for blood analysis in animal ecology and physiology because of their portability and easy operation. Although originally conceived for clinical application and to replace robust but lengthy techniques, researchers have extended the use of the i-STAT system outside of humans and even to poikilothermic fish, with only limited validation. The present study analysed a range of blood parameters [pH, haematocrit (Hct), haemoglobin (Hb), HCO(3)(−), partial pressure of CO(2) (PCO(2)), partial pressure of O(2) (PO(2)), Hb saturation (sO(2)) and Na(+) concentration] in a model teleost fish (rainbow trout, Oncorhynchus mykiss) using the i-STAT system (CG8+ cartridges) and established laboratory techniques. This methodological comparison was performed at two temperatures (10 and 20°C), two haematocrits (low and high) and three PCO(2) levels (0.5, 1.0 and 1.5%). Our results indicate that pH was measured accurately with the i-STAT system over a physiological pH range and using the i-STAT temperature correction. Haematocrit was consistently underestimated by the i-STAT, while the measurements of Na(+), PCO(2), HCO(3)(−) and PO(2) were variably inaccurate over the range of values typically found in fish. The algorithm that the i-STAT uses to calculate sO(2) did not yield meaningful results on rainbow trout blood. Application of conversion factors to correct i-STAT measurements is not recommended, due to significant effects of temperature, Hct and PCO(2) on the measurement errors and complex interactions may exist. In conclusion, the i-STAT system can easily generate fast results from rainbow trout whole blood, but many are inaccurate values.