Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution
The quick-EXAFS (QEXAFS) method adds time resolution to X-ray absorption spectroscopy (XAS) and allows dynamic structural changes to be followed. A completely new QEXAFS setup consisting of monochromator, detectors and data acquisition system is presented, as installed at the SuperXAS bending-magnet...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
International Union of Crystallography
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4733929/ https://www.ncbi.nlm.nih.gov/pubmed/26698072 http://dx.doi.org/10.1107/S1600577515018007 |
_version_ | 1782412872458960896 |
---|---|
author | Müller, Oliver Nachtegaal, Maarten Just, Justus Lützenkirchen-Hecht, Dirk Frahm, Ronald |
author_facet | Müller, Oliver Nachtegaal, Maarten Just, Justus Lützenkirchen-Hecht, Dirk Frahm, Ronald |
author_sort | Müller, Oliver |
collection | PubMed |
description | The quick-EXAFS (QEXAFS) method adds time resolution to X-ray absorption spectroscopy (XAS) and allows dynamic structural changes to be followed. A completely new QEXAFS setup consisting of monochromator, detectors and data acquisition system is presented, as installed at the SuperXAS bending-magnet beamline at the Swiss Light Source (Paul Scherrer Institute, Switzerland). The monochromator uses Si(111) and Si(311) channel-cut crystals mounted on one crystal stage, and remote exchange allows an energy range from 4.0 keV to 32 keV to be covered. The spectral scan range can be electronically adjusted up to several keV to cover multiple absorption edges in one scan. The determination of the Bragg angle close to the position of the crystals allows high-accuracy measurements. Absorption spectra can be acquired with fast gridded ionization chambers at oscillation frequencies of up to 50 Hz resulting in a time resolution of 10 ms, using both scan directions of each oscillation period. The carefully developed low-noise detector system yields high-quality absorption data. The unique setup allows both state-of-the-art QEXAFS and stable step-scan operation without the need to exchange whole monochromators. The long-term stability of the Bragg angle was investigated and absorption spectra of reference materials as well as of a fast chemical reaction demonstrate the overall capabilities of the new setup. |
format | Online Article Text |
id | pubmed-4733929 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | International Union of Crystallography |
record_format | MEDLINE/PubMed |
spelling | pubmed-47339292016-02-02 Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution Müller, Oliver Nachtegaal, Maarten Just, Justus Lützenkirchen-Hecht, Dirk Frahm, Ronald J Synchrotron Radiat Research Papers The quick-EXAFS (QEXAFS) method adds time resolution to X-ray absorption spectroscopy (XAS) and allows dynamic structural changes to be followed. A completely new QEXAFS setup consisting of monochromator, detectors and data acquisition system is presented, as installed at the SuperXAS bending-magnet beamline at the Swiss Light Source (Paul Scherrer Institute, Switzerland). The monochromator uses Si(111) and Si(311) channel-cut crystals mounted on one crystal stage, and remote exchange allows an energy range from 4.0 keV to 32 keV to be covered. The spectral scan range can be electronically adjusted up to several keV to cover multiple absorption edges in one scan. The determination of the Bragg angle close to the position of the crystals allows high-accuracy measurements. Absorption spectra can be acquired with fast gridded ionization chambers at oscillation frequencies of up to 50 Hz resulting in a time resolution of 10 ms, using both scan directions of each oscillation period. The carefully developed low-noise detector system yields high-quality absorption data. The unique setup allows both state-of-the-art QEXAFS and stable step-scan operation without the need to exchange whole monochromators. The long-term stability of the Bragg angle was investigated and absorption spectra of reference materials as well as of a fast chemical reaction demonstrate the overall capabilities of the new setup. International Union of Crystallography 2016-01-01 /pmc/articles/PMC4733929/ /pubmed/26698072 http://dx.doi.org/10.1107/S1600577515018007 Text en © Oliver Müller et al. 2016 http://creativecommons.org/licenses/by/2.0/uk/ This is an open-access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited. |
spellingShingle | Research Papers Müller, Oliver Nachtegaal, Maarten Just, Justus Lützenkirchen-Hecht, Dirk Frahm, Ronald Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution |
title | Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution |
title_full | Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution |
title_fullStr | Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution |
title_full_unstemmed | Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution |
title_short | Quick-EXAFS setup at the SuperXAS beamline for in situ X-ray absorption spectroscopy with 10 ms time resolution |
title_sort | quick-exafs setup at the superxas beamline for in situ x-ray absorption spectroscopy with 10 ms time resolution |
topic | Research Papers |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4733929/ https://www.ncbi.nlm.nih.gov/pubmed/26698072 http://dx.doi.org/10.1107/S1600577515018007 |
work_keys_str_mv | AT mulleroliver quickexafssetupatthesuperxasbeamlineforinsituxrayabsorptionspectroscopywith10mstimeresolution AT nachtegaalmaarten quickexafssetupatthesuperxasbeamlineforinsituxrayabsorptionspectroscopywith10mstimeresolution AT justjustus quickexafssetupatthesuperxasbeamlineforinsituxrayabsorptionspectroscopywith10mstimeresolution AT lutzenkirchenhechtdirk quickexafssetupatthesuperxasbeamlineforinsituxrayabsorptionspectroscopywith10mstimeresolution AT frahmronald quickexafssetupatthesuperxasbeamlineforinsituxrayabsorptionspectroscopywith10mstimeresolution |