Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry

BACKGROUND: The study aims to investigate different ground plane segmentation designs of an ultrasound transducer to reduce gradient field induced eddy currents and the associated geometric distortion and temperature map errors in echo-planar imaging (EPI)-based MR thermometry in transcranial magnet...

Descripción completa

Detalles Bibliográficos
Autores principales: Lechner-Greite, Silke M., Hehn, Nicolas, Werner, Beat, Zadicario, Eyal, Tarasek, Matthew, Yeo, Desmond
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4741031/
https://www.ncbi.nlm.nih.gov/pubmed/26848391
http://dx.doi.org/10.1186/s40349-016-0047-x
_version_ 1782413937615044608
author Lechner-Greite, Silke M.
Hehn, Nicolas
Werner, Beat
Zadicario, Eyal
Tarasek, Matthew
Yeo, Desmond
author_facet Lechner-Greite, Silke M.
Hehn, Nicolas
Werner, Beat
Zadicario, Eyal
Tarasek, Matthew
Yeo, Desmond
author_sort Lechner-Greite, Silke M.
collection PubMed
description BACKGROUND: The study aims to investigate different ground plane segmentation designs of an ultrasound transducer to reduce gradient field induced eddy currents and the associated geometric distortion and temperature map errors in echo-planar imaging (EPI)-based MR thermometry in transcranial magnetic resonance (MR)-guided focused ultrasound (tcMRgFUS). METHODS: Six different ground plane segmentations were considered and the efficacy of each in suppressing eddy currents was investigated in silico and in operando. For the latter case, the segmented ground planes were implemented in a transducer mockup model for validation. Robust spoiled gradient (SPGR) echo sequences and multi-shot EPI sequences were acquired. For each sequence and pattern, geometric distortions were quantified in the magnitude images and expressed in millimeters. Phase images were used for extracting the temperature maps on the basis of the temperature-dependent proton resonance frequency shift phenomenon. The means, standard deviations, and signal-to-noise ratios (SNRs) were extracted and contrasted with the geometric distortions of all patterns. RESULTS: The geometric distortion analysis and temperature map evaluations showed that more than one pattern could be considered the best-performing transducer. In the sagittal plane, the star (d) (3.46 ± 2.33 mm) and star-ring patterns (f) (2.72 ± 2.8 mm) showed smaller geometric distortions than the currently available seven-segment sheet (c) (5.54 ± 4.21 mm) and were both comparable to the reference scenario (a) (2.77 ± 2.24 mm). Contrasting these results with the temperature maps revealed that (d) performs as well as (a) in SPGR and EPI. CONCLUSIONS: We demonstrated that segmenting the transducer ground plane into a star pattern reduces eddy currents to a level wherein multi-plane EPI for accurate MR thermometry in tcMRgFUS is feasible.
format Online
Article
Text
id pubmed-4741031
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-47410312016-02-05 Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry Lechner-Greite, Silke M. Hehn, Nicolas Werner, Beat Zadicario, Eyal Tarasek, Matthew Yeo, Desmond J Ther Ultrasound Research BACKGROUND: The study aims to investigate different ground plane segmentation designs of an ultrasound transducer to reduce gradient field induced eddy currents and the associated geometric distortion and temperature map errors in echo-planar imaging (EPI)-based MR thermometry in transcranial magnetic resonance (MR)-guided focused ultrasound (tcMRgFUS). METHODS: Six different ground plane segmentations were considered and the efficacy of each in suppressing eddy currents was investigated in silico and in operando. For the latter case, the segmented ground planes were implemented in a transducer mockup model for validation. Robust spoiled gradient (SPGR) echo sequences and multi-shot EPI sequences were acquired. For each sequence and pattern, geometric distortions were quantified in the magnitude images and expressed in millimeters. Phase images were used for extracting the temperature maps on the basis of the temperature-dependent proton resonance frequency shift phenomenon. The means, standard deviations, and signal-to-noise ratios (SNRs) were extracted and contrasted with the geometric distortions of all patterns. RESULTS: The geometric distortion analysis and temperature map evaluations showed that more than one pattern could be considered the best-performing transducer. In the sagittal plane, the star (d) (3.46 ± 2.33 mm) and star-ring patterns (f) (2.72 ± 2.8 mm) showed smaller geometric distortions than the currently available seven-segment sheet (c) (5.54 ± 4.21 mm) and were both comparable to the reference scenario (a) (2.77 ± 2.24 mm). Contrasting these results with the temperature maps revealed that (d) performs as well as (a) in SPGR and EPI. CONCLUSIONS: We demonstrated that segmenting the transducer ground plane into a star pattern reduces eddy currents to a level wherein multi-plane EPI for accurate MR thermometry in tcMRgFUS is feasible. BioMed Central 2016-02-03 /pmc/articles/PMC4741031/ /pubmed/26848391 http://dx.doi.org/10.1186/s40349-016-0047-x Text en © Lechner-Greite et al. 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Lechner-Greite, Silke M.
Hehn, Nicolas
Werner, Beat
Zadicario, Eyal
Tarasek, Matthew
Yeo, Desmond
Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry
title Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry
title_full Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry
title_fullStr Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry
title_full_unstemmed Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry
title_short Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry
title_sort minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based mr thermometry
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4741031/
https://www.ncbi.nlm.nih.gov/pubmed/26848391
http://dx.doi.org/10.1186/s40349-016-0047-x
work_keys_str_mv AT lechnergreitesilkem minimizingeddycurrentsinducedinthegroundplaneofalargephasedarrayultrasoundapplicatorforechoplanarimagingbasedmrthermometry
AT hehnnicolas minimizingeddycurrentsinducedinthegroundplaneofalargephasedarrayultrasoundapplicatorforechoplanarimagingbasedmrthermometry
AT wernerbeat minimizingeddycurrentsinducedinthegroundplaneofalargephasedarrayultrasoundapplicatorforechoplanarimagingbasedmrthermometry
AT zadicarioeyal minimizingeddycurrentsinducedinthegroundplaneofalargephasedarrayultrasoundapplicatorforechoplanarimagingbasedmrthermometry
AT tarasekmatthew minimizingeddycurrentsinducedinthegroundplaneofalargephasedarrayultrasoundapplicatorforechoplanarimagingbasedmrthermometry
AT yeodesmond minimizingeddycurrentsinducedinthegroundplaneofalargephasedarrayultrasoundapplicatorforechoplanarimagingbasedmrthermometry