An inappropriate defibrillator shock during ventricular tachycardia

Ventricular oversensing in patients with defibrillators is an infrequent but deleterious condition. We report a patient with a cardiac resynchronization–defibrillation device that presented with hyperkalemia and syncope. Device interrogation revealed ventricular double-counting within the QRS of a s...

Descripción completa

Detalles Bibliográficos
Autores principales: Faustino, Mariana, Madeira, Francisco, Morais, Carlos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4750157/
https://www.ncbi.nlm.nih.gov/pubmed/26937116
http://dx.doi.org/10.1016/j.ipej.2015.10.002
Descripción
Sumario:Ventricular oversensing in patients with defibrillators is an infrequent but deleterious condition. We report a patient with a cardiac resynchronization–defibrillation device that presented with hyperkalemia and syncope. Device interrogation revealed ventricular double-counting within the QRS of a slow ventricular tachycardia, resulting detection of the slow ventricular tachycardia in the ventricular fibrillation zone, and delivery of an effective therapy, below device programmed detection rate. This case of defibrillator inappropriate detection emphasizes the relevance of device electrogram interrogation in order to minimize inappropriate therapies.