Acute transient non-physiological over-sensing in the ventricle with a DF4 lead

The DF-4 is a new defibrillator lead technology. We present two cases of non-physiological transient ventricular over-sensing in patients who underwent implantation of an ICD for secondary prevention. Case 1 had ventricular over-sensing during pacing threshold evaluation post defibrillation testing...

Descripción completa

Detalles Bibliográficos
Autores principales: Ng, Kevin Kit, Gould, Paul A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4750164/
https://www.ncbi.nlm.nih.gov/pubmed/26937124
http://dx.doi.org/10.1016/j.ipej.2015.10.008
_version_ 1782415387925676032
author Ng, Kevin Kit
Gould, Paul A.
author_facet Ng, Kevin Kit
Gould, Paul A.
author_sort Ng, Kevin Kit
collection PubMed
description The DF-4 is a new defibrillator lead technology. We present two cases of non-physiological transient ventricular over-sensing in patients who underwent implantation of an ICD for secondary prevention. Case 1 had ventricular over-sensing during pacing threshold evaluation post defibrillation testing while Case 2 had the lead integrity alert triggered immediately post discharge with transient over-sensing. No lead-connector issues were found. Case 1 was likely due to improper venting of the header and trapped air. Case 2 was hypothesized to be due to intermittent header pin non-contact secondary to blood in the header. These cases reveal that DF-4 leads are subject to both reported and potentially novel causes of transient acute ventricular over-sensing.
format Online
Article
Text
id pubmed-4750164
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-47501642016-03-02 Acute transient non-physiological over-sensing in the ventricle with a DF4 lead Ng, Kevin Kit Gould, Paul A. Indian Pacing Electrophysiol J Case Report The DF-4 is a new defibrillator lead technology. We present two cases of non-physiological transient ventricular over-sensing in patients who underwent implantation of an ICD for secondary prevention. Case 1 had ventricular over-sensing during pacing threshold evaluation post defibrillation testing while Case 2 had the lead integrity alert triggered immediately post discharge with transient over-sensing. No lead-connector issues were found. Case 1 was likely due to improper venting of the header and trapped air. Case 2 was hypothesized to be due to intermittent header pin non-contact secondary to blood in the header. These cases reveal that DF-4 leads are subject to both reported and potentially novel causes of transient acute ventricular over-sensing. Elsevier 2015-10-19 /pmc/articles/PMC4750164/ /pubmed/26937124 http://dx.doi.org/10.1016/j.ipej.2015.10.008 Text en Copyright © 2015, Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Ng, Kevin Kit
Gould, Paul A.
Acute transient non-physiological over-sensing in the ventricle with a DF4 lead
title Acute transient non-physiological over-sensing in the ventricle with a DF4 lead
title_full Acute transient non-physiological over-sensing in the ventricle with a DF4 lead
title_fullStr Acute transient non-physiological over-sensing in the ventricle with a DF4 lead
title_full_unstemmed Acute transient non-physiological over-sensing in the ventricle with a DF4 lead
title_short Acute transient non-physiological over-sensing in the ventricle with a DF4 lead
title_sort acute transient non-physiological over-sensing in the ventricle with a df4 lead
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4750164/
https://www.ncbi.nlm.nih.gov/pubmed/26937124
http://dx.doi.org/10.1016/j.ipej.2015.10.008
work_keys_str_mv AT ngkevinkit acutetransientnonphysiologicaloversensingintheventriclewithadf4lead
AT gouldpaula acutetransientnonphysiologicaloversensingintheventriclewithadf4lead