Billowing Of Endologix Powerlink Stent Mimicking Endoleaks

INTRODUCTION: Endoleaks remains one of the primary concerns of endovascular aortic aneurysm repair (EVAR) and is routinely followed with CT angiography (CTA). However, certain imaging findings can mimic endoleaks. CASE PRESENTATION: A 65-year-old woman who had endovascular aortic repair (EVAR) of an...

Descripción completa

Detalles Bibliográficos
Autores principales: Wu, Alex, Karuppasamy, Karunakaravel, Wang, Weiping
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Kowsar 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4755061/
https://www.ncbi.nlm.nih.gov/pubmed/26949697
http://dx.doi.org/10.5812/cardiovascmed.32448
Descripción
Sumario:INTRODUCTION: Endoleaks remains one of the primary concerns of endovascular aortic aneurysm repair (EVAR) and is routinely followed with CT angiography (CTA). However, certain imaging findings can mimic endoleaks. CASE PRESENTATION: A 65-year-old woman who had endovascular aortic repair (EVAR) of an abdominal aortic aneurysm with Endologix Powerlink system developed marked new circumferential cauliflower-like bulging of contrast-filled sacs at mid-stent-graft with enlargement of the excluded aneurysm at 3-year follow-up. CONCLUSIONS: Considering the unique construct of the Powerlink stents, this is thought to represent aneurysmal degeneration of the outer fabric material from the metal struts and may potentially pressurize the excluded sac with risk for rupture.