Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis

INTRODUCTION: The purpose of the present study was to determine whether the abatacept autoinjector can be used by the intended population without patterns of preventable use errors, and is acceptable when assessed against key user needs. METHODS: Two independently conducted simulated-use studies, wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Schiff, Michael, Koo, Joe, Jin, Erik, Schiller, Eric, Day, Ashley, Stevens, Rebecca, Laskar, Christina
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Healthcare 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4769728/
https://www.ncbi.nlm.nih.gov/pubmed/26833303
http://dx.doi.org/10.1007/s12325-016-0286-9
_version_ 1782418143288754176
author Schiff, Michael
Koo, Joe
Jin, Erik
Schiller, Eric
Day, Ashley
Stevens, Rebecca
Laskar, Christina
author_facet Schiff, Michael
Koo, Joe
Jin, Erik
Schiller, Eric
Day, Ashley
Stevens, Rebecca
Laskar, Christina
author_sort Schiff, Michael
collection PubMed
description INTRODUCTION: The purpose of the present study was to determine whether the abatacept autoinjector can be used by the intended population without patterns of preventable use errors, and is acceptable when assessed against key user needs. METHODS: Two independently conducted simulated-use studies, with no active drug administered, quantified use errors and evaluated the abatacept autoinjector and competitor devices on key attributes (comfort, control, ease of use, confidence of dose) and overall acceptability. Autoinjector preference was also assessed. Participants were patients with rheumatoid arthritis, caregivers, and healthcare professionals (HCPs). Participants were informed that a new rheumatoid arthritis autoinjector was being tested but were blinded to the intended drug and sponsor identity. RESULTS: In the formative (pre-validation) study (n = 54), two high-priority use errors occurred, both of which resulted from protocol non-compliance rather than mental confusion or physical limitations. In the summative (validation) study (n = 99), one high-priority use error occurred; this was deemed a simulated-use study artifact as participant behavior was guided by actual experience associated with the feel of drug delivery into the skin rather than by protocol, so no mitigation steps were considered necessary. Across user groups, average scores were consistently high for the pre-defined key attributes. Overall acceptability scores (7-point scale) were significantly higher for the abatacept versus competitor autoinjectors—formative study: patients 6.7 vs 5.2 (P = 0.0001), caregivers 7.0 vs 4.6 (P = 0.0093), HCPs 6.8 vs 5.1 (P = 0.0020); summative study: patients 6.5 vs 5.9 (P = 0.0404), caregivers 6.8 vs 5.8 (P = 0.0047), HCPs 6.8 vs 5.1 (P = 0.0002). The abatacept autoinjector was preferred to competitor devices: patients 85.7% vs 14.3% (P = 0.00002), caregivers 84.2% vs 15.8% (P = 0.00443), HCPs 95.0% vs 5.0% (P = 0.00004). Positive experiences with the abatacept autoinjector were attributed to the rubberized grip, device size, visualization of dose progression, button ergonomics, and ease of use. CONCLUSION: The abatacept autoinjector demonstrated usability without patterns of preventable use errors, and with high acceptability ratings across all key attributes assessed. Preference over competitor autoinjectors was due to device ergonomics, visualization of dose progression, confidence of dose delivery, and overall ease of use. FUNDING: Bristol-Myers Squibb. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1007/s12325-016-0286-9) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-4769728
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer Healthcare
record_format MEDLINE/PubMed
spelling pubmed-47697282016-03-29 Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis Schiff, Michael Koo, Joe Jin, Erik Schiller, Eric Day, Ashley Stevens, Rebecca Laskar, Christina Adv Ther Original Research INTRODUCTION: The purpose of the present study was to determine whether the abatacept autoinjector can be used by the intended population without patterns of preventable use errors, and is acceptable when assessed against key user needs. METHODS: Two independently conducted simulated-use studies, with no active drug administered, quantified use errors and evaluated the abatacept autoinjector and competitor devices on key attributes (comfort, control, ease of use, confidence of dose) and overall acceptability. Autoinjector preference was also assessed. Participants were patients with rheumatoid arthritis, caregivers, and healthcare professionals (HCPs). Participants were informed that a new rheumatoid arthritis autoinjector was being tested but were blinded to the intended drug and sponsor identity. RESULTS: In the formative (pre-validation) study (n = 54), two high-priority use errors occurred, both of which resulted from protocol non-compliance rather than mental confusion or physical limitations. In the summative (validation) study (n = 99), one high-priority use error occurred; this was deemed a simulated-use study artifact as participant behavior was guided by actual experience associated with the feel of drug delivery into the skin rather than by protocol, so no mitigation steps were considered necessary. Across user groups, average scores were consistently high for the pre-defined key attributes. Overall acceptability scores (7-point scale) were significantly higher for the abatacept versus competitor autoinjectors—formative study: patients 6.7 vs 5.2 (P = 0.0001), caregivers 7.0 vs 4.6 (P = 0.0093), HCPs 6.8 vs 5.1 (P = 0.0020); summative study: patients 6.5 vs 5.9 (P = 0.0404), caregivers 6.8 vs 5.8 (P = 0.0047), HCPs 6.8 vs 5.1 (P = 0.0002). The abatacept autoinjector was preferred to competitor devices: patients 85.7% vs 14.3% (P = 0.00002), caregivers 84.2% vs 15.8% (P = 0.00443), HCPs 95.0% vs 5.0% (P = 0.00004). Positive experiences with the abatacept autoinjector were attributed to the rubberized grip, device size, visualization of dose progression, button ergonomics, and ease of use. CONCLUSION: The abatacept autoinjector demonstrated usability without patterns of preventable use errors, and with high acceptability ratings across all key attributes assessed. Preference over competitor autoinjectors was due to device ergonomics, visualization of dose progression, confidence of dose delivery, and overall ease of use. FUNDING: Bristol-Myers Squibb. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1007/s12325-016-0286-9) contains supplementary material, which is available to authorized users. Springer Healthcare 2016-01-30 2016 /pmc/articles/PMC4769728/ /pubmed/26833303 http://dx.doi.org/10.1007/s12325-016-0286-9 Text en © The Author(s) 2016 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Schiff, Michael
Koo, Joe
Jin, Erik
Schiller, Eric
Day, Ashley
Stevens, Rebecca
Laskar, Christina
Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis
title Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis
title_full Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis
title_fullStr Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis
title_full_unstemmed Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis
title_short Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis
title_sort usability and acceptability of the abatacept pre-filled autoinjector for the subcutaneous treatment of rheumatoid arthritis
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4769728/
https://www.ncbi.nlm.nih.gov/pubmed/26833303
http://dx.doi.org/10.1007/s12325-016-0286-9
work_keys_str_mv AT schiffmichael usabilityandacceptabilityoftheabataceptprefilledautoinjectorforthesubcutaneoustreatmentofrheumatoidarthritis
AT koojoe usabilityandacceptabilityoftheabataceptprefilledautoinjectorforthesubcutaneoustreatmentofrheumatoidarthritis
AT jinerik usabilityandacceptabilityoftheabataceptprefilledautoinjectorforthesubcutaneoustreatmentofrheumatoidarthritis
AT schillereric usabilityandacceptabilityoftheabataceptprefilledautoinjectorforthesubcutaneoustreatmentofrheumatoidarthritis
AT dayashley usabilityandacceptabilityoftheabataceptprefilledautoinjectorforthesubcutaneoustreatmentofrheumatoidarthritis
AT stevensrebecca usabilityandacceptabilityoftheabataceptprefilledautoinjectorforthesubcutaneoustreatmentofrheumatoidarthritis
AT laskarchristina usabilityandacceptabilityoftheabataceptprefilledautoinjectorforthesubcutaneoustreatmentofrheumatoidarthritis