Cargando…
Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet
A preterm with gestational age of 24 weeks was intubated at day of life 16. The intubation was done in a routine manner with the use of a stylet. It took a significant effort from the clinician to pull the stylet out after intubation. After intubation the respiratory status of the neonate deteriorat...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4771891/ https://www.ncbi.nlm.nih.gov/pubmed/26989545 http://dx.doi.org/10.1155/2016/4373207 |
_version_ | 1782418463865700352 |
---|---|
author | Das, Anirudha Chagalamarri, Shwetha Saridakis, Kim |
author_facet | Das, Anirudha Chagalamarri, Shwetha Saridakis, Kim |
author_sort | Das, Anirudha |
collection | PubMed |
description | A preterm with gestational age of 24 weeks was intubated at day of life 16. The intubation was done in a routine manner with the use of a stylet. It took a significant effort from the clinician to pull the stylet out after intubation. After intubation the respiratory status of the neonate deteriorated requiring significantly greater support. When ventilating and oxygenating the infant was getting progressively difficult, the decision was made to change the endotracheal tube (ETT). The cause for deterioration of respiratory status was then determined to be a sheared piece of plastic from the sheath of the stylet which was lodged in the lumen of the ETT. After removal of the plastic particle, the condition of the infant improved significantly. |
format | Online Article Text |
id | pubmed-4771891 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-47718912016-03-17 Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet Das, Anirudha Chagalamarri, Shwetha Saridakis, Kim Case Rep Pediatr Case Report A preterm with gestational age of 24 weeks was intubated at day of life 16. The intubation was done in a routine manner with the use of a stylet. It took a significant effort from the clinician to pull the stylet out after intubation. After intubation the respiratory status of the neonate deteriorated requiring significantly greater support. When ventilating and oxygenating the infant was getting progressively difficult, the decision was made to change the endotracheal tube (ETT). The cause for deterioration of respiratory status was then determined to be a sheared piece of plastic from the sheath of the stylet which was lodged in the lumen of the ETT. After removal of the plastic particle, the condition of the infant improved significantly. Hindawi Publishing Corporation 2016 2016-02-16 /pmc/articles/PMC4771891/ /pubmed/26989545 http://dx.doi.org/10.1155/2016/4373207 Text en Copyright © 2016 Anirudha Das et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Das, Anirudha Chagalamarri, Shwetha Saridakis, Kim Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet |
title | Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet |
title_full | Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet |
title_fullStr | Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet |
title_full_unstemmed | Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet |
title_short | Partial Obstruction of the Endotracheal Tube by the Plastic Coating Sheared from a Stylet |
title_sort | partial obstruction of the endotracheal tube by the plastic coating sheared from a stylet |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4771891/ https://www.ncbi.nlm.nih.gov/pubmed/26989545 http://dx.doi.org/10.1155/2016/4373207 |
work_keys_str_mv | AT dasanirudha partialobstructionoftheendotrachealtubebytheplasticcoatingshearedfromastylet AT chagalamarrishwetha partialobstructionoftheendotrachealtubebytheplasticcoatingshearedfromastylet AT saridakiskim partialobstructionoftheendotrachealtubebytheplasticcoatingshearedfromastylet |